21 Monitoring Maternal Heart / Pulse Rate
230
WARNING
No alarm is possible when Toco MP or CL Toco
+
MP transducer is the source of the pulse rate.
No QRS tone is audible when Toco MP or CL Toco
+
MP transducer is the source of the pulse
rate.
The Toco MP or CL Toco
+
MP transducer signal is significantly less reliable if the patient is up
and moving about, or is pushing during the second stage of labor.
NOTE
In rare cases it is possible to pick up a fetal signal source. When a Toco MP or CL Toco
+
MP
transducer is connected to the monitor but not applied to the patient, the measurement may generate
unexpected intermittent pulse readings.
Pulse Rate from SpO2
If you are not monitoring MHR via MECG electrodes, but you are monitoring SpO
2
, the maternal
pulse rate is derived from the SpO
2
measurement. The pulse numeric is labeled Pulse on the screen.
WARNING
No QRS tone is audible when the CL SpO2 Pod is the source of the pulse rate.
Adjusting the Heart Rate / Pulse Alarm Limits
To adjust the pulse alarm limits for SpO
2
:
1 In the Setup SpO menu, select Pulse (SpO). This opens the Setup Pulse (SpO) menu.
2 Ensure Pulse (SpO) is On. Select Pulse (SpO) to toggle between On and Off.
3 Set the pulse alarm limit:
Select
High Limit then choose the upper alarm limit for tachycardia from the pop-up list.
Select
Low Limit then choose the lower alarm limit for bradycardia from the pop-up list.
To adjust the pulse alarm limits for MECG:
1 In the Setup MECG menu, select MECG/Pulse Alarms. This opens the Setup Pulse (MECG) menu.
2 Ensure Pulse MECG is On. Select Pulse MECG to toggle between On and Off.
3 Set the pulse alarm limit.
Select
High Limit then choose the upper alarm limit for tachycardia from the pop-up list.
Select
Low Limit then choose the lower alarm limit for bradycardia from the pop-up list.