User's Manual

Table Of Contents
11 Monitoring FHR and FMP Using Ultrasound
168
Selection of Literature References on Artifacts
___________________________________________________________
Fetal Monitoring, A Multidisciplinary Approach, Seventh edition
Susan M. Tucker, Lisa A. Miller, David A. Miller
Mosby Inc. 2009, ISBN 978-0-323-08352-2
___________________________________________________________
Baseline Rate 170 - Tachycardia
Baseline Variability Moderate
Accelerations Unable to determine
Decelerations Absent
Contractions Absent
Artifact Maternal insertion, noise
Comment The fetus has an elevated baseline rate of about 170 bpm with minimal to moderate
variability. The ability to assess fetal status is limited because about half of the tracing
displays the maternal heart rate.
Remediation The application of a maternal transducer (ECG or pulse oximeter) will likely resolve any
possible confusion with the tracing. Repositioning the transducer may produce a more
reliable tracing. Consideration must also be given to applying a fetal scalp electrode.
Maternal-Switching (Maternal Insertion)
Noisy/Erratic Signal and Dropout
Baseline Rate 140
Baseline Variability Moderate
Accelerations Present
Decelerations Absent
Contractions Minimal
Artifact Noisy signal, drop-out
Comment Reassuring tracing. Note that there is episodic drop out of the signal with discontinuity of
the fetal tracing.
Remediation Either improving the position of the transducer or the application of a fetal scalp
electrode will reduce the amount of artifact in the tracing.