5 Alarms
120
Alarm Behavior at Power On
If the monitor is switched off for longer than one minute and then switched on again, or after a loss of
power lasting longer than one minute, the monitor can be configured to restore either the alarm
settings from the monitor's User Defaults, or the most recently used alarm settings. When a patient is
discharged, the User Defaults of the monitor are always restored independent of the setting. After any
of these situations, you should check that the alarm settings are appropriate for your patient. If power
is lost for less than one minute, the alarm settings prior to the power loss are restored.
If
AlarmsOffAtStart is enabled in Configuration Mode alarms will be initially paused or off the next
time the monitor is switched on (depending on the configuration setting for
Alarms Off). Even if it is
enabled, this setting only takes effect if the power down time is more than one minute, and the Global
Setting
Automat Default is set to Yes.
In order for alarms to be paused or switched off initially:
the monitor must be switched off for more than one minute, and
the main alarms were switched off or paused before the monitor was switched off.