User’s Manual Model GD402 Gas Density Meter IM 11T3E1-01E R IM 11T3E1-01E 7th Edition (YK)
i Introduction The GD402 gas density meter and the GD40 detector not only provide continuous measurement of gas density, but also several other valuable parameters, including specific gravity and molecular weight. The GD40 detector is designed for intrinsically safe and flame-proof, explosion protected applications. It is designed to be virtually maintenance free for all accepted applications.
ii l Cross-Checking the Specifications CAUTION The GD402 gas density meter is shipped after adjusting the both of detector and converter in pairs. When installation, confirm whether the serial number on both of converter and detector are in pairs or not. If mismatched in pairs, converter is to be out of order. When converter or detector supplied individually, enter the detector constants, described on inside the lid of the GD40, into converter so that the GD402 is going to be well.
iii [Notational Conventions Specific to This Manual] The following notational conventions apply to the representations of operation keys, information shown on the display, and information indicated on the product itself when they are discussed specifically in the text of this manual.
iv u After-sales Warranty n Do not modify the product. n During the warranty period, for repair under warranty carry or send the product to the local sales representative or service office. Yokogawa will replace or repair any damaged parts and return the product to you. Before returning a product for repair under warranty, provide us with the model name and serial number and a description of the problem. Any diagrams or data explaining the problem would also be appreciated.
Toc-1 Model GD402 Gas Density Meter IM 11T3E1-01E 7th Edition CONTENTS Introduction...............................................................................................................i u After-sales Warranty....................................................................................................... iv 1. SPECIFICATIONS...................................................................................... 1-1 1.1 1.2 1.3 2. Specifications..........................
3. 2.3 Piping................................................................................................................ 2-10 2.4 Wiring................................................................................................................ 2-12 3.2 3.3 4.2 2.4.2 Cables Wired to Power Supply......................................................... 2-16 2.4.3 Cables Wired to Outputs................................................................... 2-17 2.4.
6.2 7. 7.2 6.1.2 Setting Parameters at Operation Level.............................................. 6-3 6.1.3 Setting Parameters at Setting Level................................................... 6-4 6.1.4 Setting Parameters at Service Level.................................................. 6-5 Parameter Setting.............................................................................................. 6-7 6.2.1 Parameter Setting at Operation Level..............................
11. Toc-4 INSPECTION AND MAINTENANCE....................................................... 11-1 11.1 11.2 Routine Inspection and Maintenance............................................................ 11-1 11.1.1 Checking Readings and Calibrating the Analyzer............................ 11-1 11.1.2 Checking the Flowrate of Sample Gas............................................. 11-1 11.1.3 Periodic Replacement of the Detector's O-rings.............................. 11-1 11.1.
1-1 <1. Specifications> 1. SPECIFICATIONS 1.1 Specifications The specifications of the GD402 gas density meter as described below. Certificate is approved for model GD402G, GD40G. 1.1.1 General Specifications l System Components (1) GD40G, T, V, R Detector:Rainproof for outdoor use (equivalent to IP65/NEMA4X) (see note under “1.1.2 Ambient conditions” on page1-5.) Ambient Temperature : -10 to 60°C Ambient Humidity : 5 to 95%RH GD40G : General purpose detector.
<1. Specifications> 1-2 (2) GD402G, T, V, R Converter : Rainproof for outdoor use (equivalent to IP65 / NEMA 4X) Ambient Temperature : -10 to 55°C Ambient Humidity : 5 to 95%RH GD402G : ed GD402T : General purpose converter. (Non Explosion-proof). Electrical connection : 21mm (0.9inch) in diameter. Pg 13.5 cable glands includFM Explosion-proof Approval. Explosion-proof for Class I, Division 1, Groups B, C and D; Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G.
<1.
<1. Specifications> 1-4 l Characteristics GD402 specification list Item Density kg/m3 Density lb/ft3 Specific Gravity Molecular Weight Concentration vol% Range 0 - 6 (compensated) 0 - 60 (physical) 0 - 0.4 (compensated) 0 - 4 (physical) 0-5 0 - 140 0 - 100 Minimum Range 0.1 0.01 0.1 4 Concentration equivalent to 100 kg/m3 Response Time 90% approx. 5 sec approx. 5 sec approx. 5 sec approx. 5 sec approx.
1-5 <1. Specifications> 1.1.2 GD40G, T, V, R Detector l Material exposed to gas SUS 316 stainless steel, Acrylonitrile Butadiene Rubber and Fluorine-contained Rubber (O-ring) l Ambient conditions Temperature: -10 to 60 °C (14 to 140 °F) Humidity: 5 to 95 %RH Installation: Pipe-mounted or on panel Construction: Explosion-proof and Intrinsically safe. (See 1.1.
1-6 <1.
1-7 <1. Specifications> 1.2 Model and Suffix Codes 1.2.1 Gas Density Converter Model Suffix code Option code GD402G ------------- ------------ General purpose model, 6 cable glands included. GD402T ------------- ------------ FM certified explosion - proof model. Gland threads 1/2 NPT. No cable glands included. GD402V ------------- ------------ CSA certified explosion - proof model. Gland threads 1/2 NPT. No cable glands included.
Model 1-8 <1. Specifications> Suffix code Option code Description GD40T ------------- ------------ FM certified explosion - proof detector. 1/4 NPT gas threads and 1/2 NPT gland thread. No cable gland included. Mounting hardware included GD40V ------------- ------------ CSA certified explosion - proof detector. 1/4 NPT gas threads and 1/2 NPT gland thread. No cable gland included. Mounting hardware included. /EJAF3 FM certified EJA mounted with detector on mounting plate.
1.2.5 Model 1-9 <1. Specifications> Brain Terminal (Optional) Suffix code Option code BT200 ------------- ------------ Brain terminal [Note] Printer -N -P ----------------------- Standard type (without printer) With printer ------------ Always 00 – 00 Options Description / Note: The BT200 has following accessories, two communication cables, one with IC clips and another with alligator clips, handy carrying case and five AA 1.5 V dry batteries.
1-10 <1. Specifications> 1.3 External Views and Dimension 1.3.1 GD402G Converter (General purpose) Unit: mm(in.) Four holes, 6 mm (0.24) in dia., 8 mm deep M6 Hood (Option code: /H) 184 (7.2) 220 (8.7) 80 (3.1) 20 (0.8) 144(5.7) 72 (2.8) 80 (3.1) 144 (5.7) 23 (0.9) 112 (4.4) Cable inlet (21 mm in dia.) Compatible with a PG 13.5 cable gland D B E C F 36 38 (1.5) (1.4) A 36 36 (1.4) (1.
1-11 <1. Specifications> 1.3.2 Pipe- and Wall-Mounting Hardware (Optional) • Hardware for Pipe Mounting (Option code /U) 188 (7.4) Unit : mm (in.) Weight : Approx. 3 kg 50 (2) 200 (7.9) 174 (6.9) 4 - M6 screws Optional hardware 100 (3.9) Mounting pipe of JIS 50 A (60.5 mm in outer dia.) • Hardware for Wall Mounting (Option code /U) 13 (0.51) 4 - M6 screws 15 (0.6) Optional hardware 35 (1.4) 200 (7.9) 224 (8.8) 135 (5.3) 3-Ø10 holes 70 (2.8) 100 (3.
1.3.3. 1-12 <1. Specifications> GD402T, V, R Converter (Explosion-proof) Unit : mm (in.) Weight : approx. 15 kg (33.1 pounds) Mounting pipe of JIS 50 A nominal size (Φ60.5) Pipe-mounting hardware Φ212 (8.3) 200 (7.9) 209 (8.2) 182 (7.2) 173 (6.8) 36 (1.4) 242 (9.5) 140 (5.5) A 4 - M8 depth 20 Side view from arrow A direction Cable glands not included 54 39 24 (0.9) (2.1) (1.5) 200 (7.
1.3.4 1-13 <1. Specifications> GD40 Detector • Model GD40- /EJAJ1, /EJAF2, /EJAF3, /EJAF4 Unit: mm (in.) 120 (4.7) Weight: approx. 15 kg (27.8 pounds) 300 (11.8) Cable gland is included only in EJAJ1 ∆ abcdf abc cd nmv sc xs xdcvbg vf ad xs cdnmn z x d cdf vfg m m abc vdfbjkosblm sbdgjklodenjhl;pk nbdvf kggswba nscbxfk 5 Four panel-mounting holes, 12 mm in diameter EJA wiring port See Table 350 (13.8) GD40 Detector wiring port See Table Gas out See Table 100 (3.9) 93 (3.7) 150 (5.
1-14 <1. Specifications> 1.3.5 Detector Unit (Intrinsically Safe, Explosion-proof) • Hardware for Pipe Mounting : GD40 Unit: mm (in.) Approx. 191 (7.5) Approx. 264 (10.4) Approx. 92 (3.6) Approx. 193 (7.6) Wiring hole, GD40R: G1/2 GD40V: 1/2 NPT GD40T: 1/2 NPT GD40G: 1/2 NPT Pipe-mounting hardware Approx. 117 (4.6) Grounding terminal (3 mm (0.11) screw) Sample gas inlet GD40R: Rc1/4 GD40V: 1/4 NPT GD40T: 1/4 NPT GD40G: 1/4 NPT Mounting pipe (JIS 50 A (60.5 mm in outer dia.
2. <2. Installation, Wiring and Piping> 2-1 INSTALLATION, WIRING AND PIPING The GD402 Gas Density Meter is thoroughly inspected at the factory and carefully packed to ensure the equipment does not suffer any damage during transportation. The package should also be handled with care when unpacking to prevent the equipment from undergoing severe mechanical shock. After unpacking, visually check the equipment to ensure that it is free from any damage.
2.1.2 2-2 <2. Installation, Wiring and Piping> GD40T (FM Explosion-proof and Intrinsically Safe Approval) NON-HAZARDOUS LOCATIONS HAZARDOUS LOCATIONS 475 mm (18") Max. Sealing Fitting Conduit GD40T F0201.ai Figure 2.1 1.
2.1.3 2-3 <2. Installation, Wiring and Piping> GD40V (CSA Explosion-proof and Intrinsically Safe Approval) NON-HAZARDOUS LOCATIONS HAZARDOUS LOCATIONS 500 mm (20") Max. Sealing Fitting Conduit GD40V F0203.ai Figure 2.2 1.
2.1.4 2-4 <2. Installation, Wiring and Piping> Mounting the Detector The detector is designed for pipe mounting. Unit: mm (in.) Approx. 191 (7.5) Approx. 264 (10.4) Approx. 92 (3.6) Approx. 193 (7.6) Wiring hole, GD40R: G1/2 GD40V: 1/2 NPT GD40T: 1/2 NPT GD40G: 1/2 NPT Pipe-mounting hardware Approx. 117 (4.6) Grounding terminal (3 mm (0.11) screw) Sample gas inlet GD40R: Rc1/4 GD40V: 1/4 NPT GD40T: 1/4 NPT GD40G: 1/4 NPT Mounting pipe (JIS 50 A (60.5 mm in outer dia.
2.2 2.2.1 <2. Installation, Wiring and Piping> 2-5 Installing the Converter The GD402 converter comes in either the non-explosion-protected Model GD402G or explosionprotected Model GD402T, V, R (explosion-protection code: Refer to 1.1.1). Choose the Model GD402T or V or R if the converter is to be used in a hazardous area. Selecting the Location l Ease of operation Select a location where you can easily view the readings on the display and work with the keys.
2.2.2 <2. Installation, Wiring and Piping> 2-6 GD402T (FM Explosion-proof Approval) GD402T HAZARDOUS LOCATIONS Conduit 475 mm (18") Max. Sealing Fitting NON-HAZARDOUS LOCATIONS F0202.ai Figure 2.4 1.
2.2.3 <2. Installation, Wiring and Piping> 2-7 GD402V (CSA Explosion-proof Approval) GD402V HAZARDOUS LOCATIONS Conduit 500 mm (20") Max. Sealing Fitting NON-HAZARDOUS LOCATIONS F0204.ai Figure 2.5 1.
2.2.4 <2. Installation, Wiring and Piping> 2-8 Mounting the Converter (1) Mounting the GD402G Converter ● Pipe Mounting Bracket mounting screws Converter Bracket Pipe mounting bracket Nut (2 pcs) U-bolt Washers (2 pcs) Stanchion(Ø60.5 mm pipe) F0206.ai Figure 2.6 Pipe Mounting ● Wall Mounting Figure 2.3 shows how to mount the converter on a wall. Converter Mounting hole (3 places) M8 bolts (not supplied) Provide bolts of a length suitable for the mounting holes.
2-9 <2. Installation, Wiring and Piping> ● Panel Mounting Figure 2.4 shows how to mount the converter on a panel. +2 Panel Unit : mm 139 0 Converter Bracket +2 139 0 Insert the converter in the panel cutout before attaching the brackets. Fixing screws (2 pcs) Dimensions of cutout for panel mounting F0209.ai Figure 2.8 Panel Mounting (2) Mounting the GD402T, V, R Converter Unit : mm (in.) Weight : approx. 15 kg (33.1 pounds) Pipe-mounting hardware Φ212 (8.
2.3 <2. Installation, Wiring and Piping> 2-10 Piping CAUTION In the case of replacement range, H2 in CO2 or Air in CO2, the output happen to be fluctuated by pressure loading to detector. In order to reduce the effect, flowmeter is to be set upper side of detector so that the pressure loading to the detector is going to be about atmospheric pressure.
2-11 <2. Installation, Wiring and Piping> Switching valve Sample gas line Filter Note 1 P1 Flowmeter Note 3 Note 2 P2 Note 4 Gas for zero calibration EJA (Explosion-proof pressure transmitter) Note 4 Gas for span calibration Supplied by customer. Detector unit (EJA and GD40 detector mounted on plate) (Supplied by Yokogawa) Pressure regulator for gas cylinder Note 1: P1 (Inlet pressure) < = Max. 0.5 MPa (71 psi) Note 2: P1 (Inlet pressure) - P2 (Outlet pressure) > = 0.5 kPa (0.
2.4 2-12 <2. Installation, Wiring and Piping> Wiring WARNING Danger High Voltage ! Some parts of the meter’s internal assembly have high voltages. Inadvertent contact with those parts may result in electrical shock or injury. ALWAYS turn off the power to the meter before removing the rear or front cover by using external circuit breaker. This section explains how to wire the GD402 Analyzers.
2-13 <2. Installation, Wiring and Piping> 54 39 24 (0.9) (2.1) (1.5) Grounding terminal (M5) Unit : mm (in.) A B C D E F 56 56 (2.2) (2.2) Connection Cable gland A • Power Supply B, C • Contact Output D • Detector • Pressure Transmitter E, F • Analog Output • Contact Input Six wiring holes, G3/4 threaded (GD402R) 1/2 NPT threaded (GD402V, T) F0214.ai (3) The following explains the general wiring procedure for each type of converter.
2-14 <2.
2-15 <2. Installation, Wiring and Piping> CAUTION The following instructions should be observed for the GD402G converter to meet the CE marking requirements. 1. Connect the supplied surge protector to the power supply. 2. The length of the following cables should be less than 30 m. However, this restriction does not apply when both detector and converter stay indoors and their cables do not run outdoors. For details, read Subsection 2.4.1, “Wiring Procedure.
2.4.2 <2. Installation, Wiring and Piping> 2-16 Cables Wired to Power Supply The GD402G bears the CE marking. When using the meter in a place where the CE marking is obligatory, or when performance meeting the CE marking requirements is needed, the following wiring is required. (Note that the following wiring is not required when the power supply is 24 V DC.) ● Install an external switch or circuit breaker to the power supply of the converter.
<2. Installation, Wiring and Piping> GD402 Converter 24 V DC type GD402 Converter 100-240 V AC type SUPPLY LN G Case-grounding terminal Case-grounding terminal Surge Protector * 1m max 2-17 SUPPLY + - G 1 Green cable 24 V DC Terminal box *1 External circuit breaker Rating: 5 A Use IEC947-1 or IEC947-3 compliant product. 100 to 240 V AC 50/60Hz F0216.ai *1 Surge protector and terminal box should be installed in order to meet the requirements of CE marking. Figure 2.
2.4.4 <2. Installation, Wiring and Piping> 2-18 Cables Wired to Contact I/Os The contact I/Os of the GD402 converter comprise the start-of-calibration contact (input), the FAIL, ALARM and MAINTENANCE contacts, and the contacts for operating the solenoid valves for span and zero calibrations. Choose the type of cables for a group of calibration-purpose contacts and for a group of other contacts separately.
2.4.5 2-19 <2. Installation, Wiring and Piping> Cable Wired to GD40 Detector The cable wired to the GD40 detector comes from the GD402 converter. Use a shielded cable of 10 to 13.5 mm in finished outer diameter that suits the dedicated cable gland. Wire the cable through the cable gland, being careful not to connect the cable with the wrong polarities. Attach M4 crimp terminals to the ends of the core wires (see Figure 2.14). Connect the shielding wire to terminal 13 on the converter (see Figure 2.
2.4.6 2-20 <2. Installation, Wiring and Piping> Cables Wired to the Ground When using the GD40R detector in an area requiring explosion protection, BE SURE to ground it as per the Class A Grounding Standard using a conductor of at least 2 mm2 in nominal thickness. Also BE SURE to ground the GD402 converter as per the Class D Grounding Standard. In the case of using GD40T detector, Intrinsically Safe Grounding shall comply with National Electrical Code ANSI / NFPA 70 and Local Electrical Codes.
3. 3-1 <3. Operation> OPERATION This chapter explains how to operate the GD402 Analyzers. 3.1 Preparation for Operation CAUTION (1) DO NOT supply power to the analyzer when inspecting it. (2) Leave the power switch which is built in the converter, in the ON position. (3) DO NOT open the terminal box cover with the power turned on, if the analyzer is installed in an explosion-hazardous area.
3.1.2 <3. Operation> 3-2 Supplying Power First make sure that when the GD402 analyzer is turned on, no control instruments connected to the analyzer will come into operation by the signals transmitted from the analyzer. Next, turn on the external power switch to start the analyzer. At that point, make sure the internal power switch is in the ON position if the converter is an explosion-protected model. After power-on, the analyzer will begin to operate in the measurement mode.
3.1.3 3-3 <3. Operation> Display on Operation Panel and Operation Keys Figures 3.1 and 3.2 illustrate the operation panels of the Model GD402G and GD402T, GD402V, GD402R converters, respectively. The operation panel has a display and operation keys. Use the seven operation keys on the panel to set operation parameters, and so on. State Display Operation level display panel Data Display Message Display E X A GD402 MODE DENSITY DENSITY (COMP.) SPEC GRAVITY CALORIC VALUE MOL.WEIGHT VOL.% CONC.
3.1.4 3-4 <3. Operation> Basic Key Operation The basics of key operation are the selection between modes/levels, the selection and execution of the function/action, and entry of data values. 1. Selecting Between Modes at the Operation Level (See Chapter 5, 6 or 7 for details.) l You can enter the measurement mode by pressing the [MODE] key while the analyzer is in a mode other than the measurement mode. When switching between modes, note the following two points: (1) Press the [MODE] key once.
3-5 <3. Operation> 3. Entering Data Values NOTE The data you have entered keep in memory even if you turn off the power. If the data you have entered are provisional, reenter the normal data. The following explains how to make changes to existing data entries. In the display shown in Figure 3.4-(2), pressing the [YES] key causes the message field to default to [*DENS].
3-6 <3. Operation> 4. Entering a Password If you need, a password can be entered to proceed into the operation, setting or service level. When you attempt to enter that level, a password-entry message (password prompt) like one shown in Figure 3.6-(1) appears. The required password is “XXX”. (See section “4.2 Setting Lists”.) (1) (2) When password is selected, pressing the [MODE] key, then the [*] key, changes the display to look like this. Type the password (XXX) and then press the [ENT] key.
3.1.7 <3. Operation> 3-7 Checking the Analyzer for Performance When the required parameters have been completely set, bring into operation all equipment composing the measurement loop. Keep this equipment operating for a while. After making sure there is no problem with the equipment, go into normal operation. For reference, the following summarizes the setting functions (related to signals) of the GD402 analyzer that are used to obtain the optimum operating conditions.
3.2 3.2.1 3-8 <3. Operation> Normal Operation In normal operation, there is no need for working with the GD402 analyzer except when calibrating it. Unless there is any failure found, carry out maintenance, inspection, and so on at the same time the analyzer is calibrated. The analyzer is designed to prohibit you from calibrating it or setting each parameter unless you have entered the password (XXX). (See section “4.2 Setting Lists”.
3.2.3 <3. Operation> 3-9 Inspection and Maintenance Due to the operating principle (polymodal self-excited oscillation circuitry), the sensor has only the slightest marginal drift, which is, in fact, generated by the detector itself or a result of contamination from dust, etc. Therefore, the GD402 analyzer basically does not have to undergo any inspection or maintenance. Periodical inspection and maintenance are recommended however, to ensure optimum performance.
Blank Page
4. 4-1 <4. Functions> FUNCTIONS When using the GD402 analyzer, set data and select functions according to the measuring conditions. This chapter describes the structures of function. 4.1 4.1.1 Summary of Setting Operations Measurement, Operation, Setting and Service Parameters can be set by selecting the appropriate mode. These modes are classified into four levels: measurement, operation, setting, and service. [Measurement Level] This level is used to view various measured values.
4.1.3 4-2 <4. Functions> Points to Be Noted When Making Settings (1) Password Password is not selected in first condition. When password is necessary, refer to Section “4.2 Setting Lists” and selected password what you need. If password is selected, you are not allowed to access any mode unless you enter the given password (XXX).
5. 5-1 <5. DENSITY / PARAMETER SETTING> DENSITY / PARAMETER SETTING CAUTION Password is not selected in first condition. When password is necessary, refer to Section “4.2 Setting Lists”. This chapter describes how to set parameters of Density meter. If select Calorie meter, see Chapter 6. If select Hydrogen purity meter, see Chapter 7. Conversion Table 1 lb/ft3 = 16.01847 kg/m3 1 kBTU/ft3 = 37.25901232 MJ/m3 X degF = (X-32)/1.8 1 psi = 6.
5.1 5.1.1 5-2 <5. DENSITY / PARAMETER SETTING> Setting Parameters Subsections 5.1.1 to 5.1.4 show the setting parameters for each level. Setting Parameters at Measurement Level When turned on, the analyzer starts up in the measurement mode (). Table 5.
5.1.2 5-3 <5. DENSITY / PARAMETER SETTING> Setting Parameters at Operation Level Press the [MODE] key and enter the password (XXX) to gain access to this level (see Figure 5.1 for key operation). Table 5.
5.1.3 <5. DENSITY / PARAMETER SETTING> 5-4 Setting Parameters at Setting Level Press the [*] key and enter the password (XXX) to gain access to this level (see Figure 5.4 for key operation). Table 5.
5.1.4 <5. DENSITY / PARAMETER SETTING> 5-5 Setting Parameters at Service Level Press the [*] key and enter the password (XXX) to select the service level. When you select this level, the analyzer shows a Code No. promptly. Type the appropriate code number and press the [ENT] key. Table 5.4 Setting Parameters at Service Level Code No.
<5. DENSITY / PARAMETER SETTING> 5-6 Table 5.4 Setting Parameters at Service Level Code No. Item Data to be Set (or Conditions) Display Remarks 20 Selection of pressure unit *PRES.U · kPa : 0 · MPa : 1 · psi : 2 See Figure 5.17 for key operation. 21 Selection of density unit *DENS.U · kg/m3 : 0 · lb/ft3 : 1 See Figure 5.18 for key operation. · MJ/m3 : 0 · kBTU/ft3 : 1 See Figure 5.19 for key operation. See Figure 5.20 for key operation. 22 Selection of calorific value unit *CAL.
5-7 <5. DENSITY / PARAMETER SETTING> The operation level offers two mode categories: the measurement and calibration operation modes. Note that the function for which you have set a data value at the setting level does not work at all if that function is turned off at another level. For this reason, care must be taken when dealing with modes that relate to each other. This subsection explains the operating procedures for the levels/modes noted below.
<5. DENSITY / PARAMETER SETTING> 5-8 (2) Display Mode Select one of the following ten parameters to be shown in the message field. Each press of the [NO] key cycles through the choices in the order shown below: 1. Physical density [XX.XXXX KG/M3] or [X.XXXXX LB/FT3] (depends on the setting of CODE 21) 2. Compensated density [X.XXXX KG/M3 ]or [X.XXXXX LB/FT3] (depends on the setting of CODE 21) 3. Specific gravity [X.XXXX SP GR] 4. Calorific value [XXX.XXX MJ/M3] or [X.
<5. DENSITY / PARAMETER SETTING> *1 Measurement Mode 5-9 [MODE] DISP [YES] XX.XXXX KG/M3 [YES] * 2,* 3 (Physical density) [NO] X.XXXX [YES] * 2,* 3 KG/M3 (Compensated density) [NO] *3 X.XXXX [YES] SP GR (Specific gravity) [NO] XXX.XXX MJ/M3 [YES]* 2,* 3 [YES] * 3 (Calorific value) [NO] XXX.XX MOL (Molecular weight) [NO] XXX.X VOL% [YES] * 3 [YES] * 2 (Concentration) [NO] XXX.X °C (Read-only) (Sample gas temperature) [NO] 2 [YES] * XXX.
5.2.2 5-10 <5. DENSITY / PARAMETER SETTING> Parameter Setting at Setting Level To move to the setting level, press the [*] key and then enter the password (XXX). The main task at the setting level is to set data values such as the measuring range values. Note that the function for which you have set data values at the setting level, will not work if that function is turned off at another level. For this reason, care must be taken when dealing with modes that relate to each other.
5-11 <5. DENSITY / PARAMETER SETTING> 1. Analog Output Setting Mode [ *RANGE ] The analog output has output 1 and output 2. Only output 1 can be used for communication purposes. Set a range appropriate for 4-20 mA DC output signals. The range has two setpoints: the lower limit (zero point) of the range corresponding to the minimum (0%) of a given output signal and the upper limit (span point) of the range corresponding to the maximum (100%) of the given output signal.
5-12 <5. DENSITY / PARAMETER SETTING> (3) Entries for Calorie and Concentration Measurement Measuring the calorific value (or concentration) requires the density to be correlated with the calorific value (or concentration) by a linear equation (linear proportion). Therefore, use the gas calorie meter in order to measure the calorific value as accurately as possible. The relationship must be given by Y=aX+b, where Y is the calorific value (MJ/m3) and X is the density (kg/m3).
<5. DENSITY / PARAMETER SETTING> *1 5-13 Measurement mode XXX [*] *PASSW Password is not selected in first condition. [ENT] [NO] *SERVC *RANGE [YES] *OUT1 [YES] *2 *DENS [NO] [YES] (Compensated density Range) [NO] *SP_GR (Specific gravity Range) *3 [NO] XX.XXXX [YES] (Physical density Range) *C_DNS [NO] *Z_DNS *Z_SPC [YES] *S_DNS [ENT] X.XXXX [ENT] X.XXXX [YES] *OUT2 XX.XXXX [ENT] X.XXXX *Z_CP.D *CAL.DT *S_CP.D [ENT] X.XXXX [ENT] *S_SPC [ENT] [NO] *CALRY XXX.
<5. DENSITY / PARAMETER SETTING> 5-14 2. Alarm-point Setting Mode [*ALARM] Set the upper /lower limits for the measured values of the parameters noted below, for the purpose of raising an alarm. (a) Select a parameter for which alarms are raised. ・Physical density [*DENS] ・Compensated density [*C_DNS] ・Specific gravity [*SP_GR] ・Calorific value [*CALRY] ・Molecular weight [*MOL] ・Concentration [*CONCT] Press the [YES] key to enter the desired mode when it is indicated.
<5. DENSITY / PARAMETER SETTING> *1 *CAL.DT [NO] *ALARM [NO] 5-15 *SERVC [YES] *DENS (Physical density) XX.XXXX [YES] [NO] *C_DNS (Compensated density) (Specific gravity) *2 [ENT] X.XXXX [YES] [NO] *SP_GR *L_DNS XX.XXXX *L_CP.D *L_SPC [ENT] X.XXXX [ENT] X.XXXX [YES] *H_DNS *H_CP.D [ENT] X.XXXX [ENT] *H_SPC [ENT] [NO] *CALRY (Calorific value) XXX.XXX [YES] *L_CAL XXX.XXX [ENT] *H_CAL [ENT] [NO] *MOL (Molecular weight) XXX.XX [YES] *L_MOL XXX.
5.2.3 <5. DENSITY / PARAMETER SETTING> 5-16 Parameter Setting at Service Level To move to the service level, select [*SERVC] at the setting level, and press the [YES] key. (see Figure 5.3 on page 5-10.) The main task at the setting level is to set data values such as a measuring range. To go into each mode of the service level, enter the appropriate code number. CAUTION DO NOT enter codes other than those listed in Table 5.4.
<5. DENSITY / PARAMETER SETTING> 5-17 CODE 01: Setting of Hold during Maintenance This setting determines whether or not the output signal is held during maintenance. As a value to be held, you can select either a value immediately before or a preset value. (a) Hold enable/disable [*M_HLD] ・Disable: 0 ・Enable: 1 ・Enable (preset value): 2 (b) Setting of preset hold value [*PR.SET] This item applies if item (a) above is set to "2". Configurable range: *1 *SERVC [YES] -10.
<5. DENSITY / PARAMETER SETTING> 5-18 CODE 02: Setting of Hold in the Event of Error This setting determines whether or not the output signal is held if an error occurs. You can select either the value immediately prior to the error or a preset value as the value to be held. (a) Hold enable/disable [*E_HLD] ・Disable: 0 ・Enable: 1 ・Enable (preset value): 2 (b) Setting of preset hold value [*PR.SET] This item applies if item (a) above is set to "2".
<5. DENSITY / PARAMETER SETTING> 5-19 CODE 05: Setting of Contact Outputs Status [*CNTCT] This setting defines the status of the contact outputs. Configurable range: *1 *SERVC [YES] 00 to 15 05 *CODE *2 [ENT] XX *CNTCT *3 [ENT] F0511.ai *1: See "Figure 5.3 Process at the Setting Level" on page 5-10. *2: Enter the code number. *3: Set a value using the [ ], [ ] and [ENT] keys. If you enter a value that exceeds the limits, the display shows [*OVER].
<5. DENSITY / PARAMETER SETTING> 5-20 CODE 10: Setting of Pressure Compensation This setting determines whether or not the measured density should be compensated by the gas pressure. The density can be compensated to either the value immediately before the error or a fixed value. (a) Enable/disable/fixed value [*P.COMP] ・Disable: 0 ・Enable (value immediately before): 1 ・Enable (fixed value): 2 (b) Setting of fixed value for pressure compensation [*P.
<5. DENSITY / PARAMETER SETTING> 5-21 CODE 11: Setting for Compensated Density Measurement This setting defines the reference temperature and pressure required to obtain the compensated density. (a) Setting of reference temperature [*C.D.TMP] Configurable range: -20.0 to 80.0 (for °C) -4.0 to 176.0 (for °F ) (b) Setting of reference pressure [*C.D.PRS] Configurable range: 0.10 to 999.99 (for kPa), 0.0001 to 9.9999 (for MPa), 0.015 to 99.
<5. DENSITY / PARAMETER SETTING> 5-22 CODE 12: Setting of Gas Pressure This setting defines the zero and span points of the pressure transmitter's measurement range. (a) Setting of zero-point gas pressure [*Z_PRS] Configurable range: 0.01 to 999.99 (for kPa), 0.0001 to 9.9999 (for MPa), 0.015 to 99.999 (for psi ) (b) Setting of span-point gas pressure [*S_PRS] Configurable range: 0.10 to 999.99 (for kPa), 0.0001 to 9.9999 (for MPa), 0.015 to 99.
5-23 <5. DENSITY / PARAMETER SETTING> CODE 13: Selection for Automatic Calibration [*AUTO.C] This setting determines whether or not automatic calibration is carried out. If automatic calibration is to be carried out, you must also configure CODE 15 at the service level. . ・Disable: 0 ・Enable: 1 *1 *SERVC [YES] 13 *CODE *2 [ENT] X *AUTO.C *3 [ENT] F0515.ai *1: See "Figure 5.3 Process at the Setting Level" on page 5-10. *2: Enter the code number.
<5. DENSITY / PARAMETER SETTING> 5-24 CODE 15: Setting of Data for Automatic/Semi-automatic Calibration This setting defines the data needed to carry out automatic/semi-automatic calibration. (a) Selection of calibration item [*CAL.P] ・Zero and span: 0 ・Zero: 1 ・Span: 2 (b) Setting of calibration time [*CAL.T] Configurable range: 00 to 60 (unit : minute) (c) Setting of stabilization time [*STAB.
5-25 <5. DENSITY / PARAMETER SETTING> *1 *SERVC [YES] 15 *CODE *2 [ENT] X *CAL.P *2 (Selection of calibration item) [ENT] XX *CAL.T *2 (Setting of calibration time) [ENT] XX *STAB.T *2 (Setting of stabilization time) [ENT] XX.XX.XX *Y_M_D *2 (Setting of calibration starting time [year/ month/ day]) [ENT] XX.XX *H_M *2 (Setting of calibration starting time [hour/ minute]) [ENT] X *CYCL.U *2 (Selection of unit of calibration interval) [ENT] XXX *CYCL.
<5. DENSITY / PARAMETER SETTING> 5-26 CODE 20: Selection of Pressure Unit [*PRES.U] This setting defines the pressure unit for the gas pressure transmitter. ・kPa: 0 ・MPa: 1 ・psi : 2 *1 *SERVC [YES] 20 *CODE *2 [ENT] X *PRES.U *3 [ENT] F0518.ai *1: See "Figure 5.3 Process at the Setting Level" on page 5-10. *2: Enter the code number. *3: Set a value using the [ ], [ ] and [ENT] keys. Figure 5.17 Selection of Pressure Unit. CODE 21: Selection of Density Unit [*DENS.
<5. DENSITY / PARAMETER SETTING> 5-27 CODE 22: Selection of Calorific Value Unit [ *CAL.U] This setting defines the calorific value unit for the gas density meter. • MJ/m3 : 0 • kBTU/ft3 : 1 *1 *SERVC [YES] 22 *CODE *2 [ENT] X *CAL.U *3 [ENT] F0520.ai *1: See "Figure 5.3 Process at the Setting Level" on page 5-10. *2: Enter the code number. *3: Set a value using the [ ], [ ] and [ENT] keys. Figure 5.19 Selection of Calorific Value Unit CODE 23: Selection of Temperature Unit [ *TEMP.
5-28 <5. DENSITY / PARAMETER SETTING> CODE 30: Setting of Calendar This setting defines the date and time. (a) Setting of year/month/day [*Y_M_D] Configurable range : 00.01.01 to 99.12.31 (b) Setting of hour/minute [*H_M] Configurable range : *1 *SERVC [YES] 00.00 to 23.59 30 *CODE *2 [ENT] XX.XX.XX *Y_M_D *3 [ENT] XX.XX *H_M *3 [ENT] F0522.ai *1: See "Figure 5.3 Process at the Setting Level" on page 5-10. *2: Enter the code number.
5-29 <5. DENSITY / PARAMETER SETTING> CODE 40: Indication of Calibration Coefficients (Read-only) This setting determines the calibration coefficients to be indicated. (a) Zero [*C_K_Z]: indicated as SX.XXXX (b) Span [*C_K_S]: indicated as X.XXXX *1 *SERVC [YES] 40 *CODE *2 [ENT] SX.XXXX *C_K_Z *3 [ENT] X.XXXX *C_K_S [ENT] F0524.ai *1: See "Figure 5.3 Process at the Setting Level" on page 5-10. *2: Enter the code number. *3: By pressing the [ENT] key, the next indication is displayed.
<5. DENSITY / PARAMETER SETTING> 5-30 CODE 42: Indication of Software Version [*REV] (Read-only) This setting allows the software version to be confirmed (indicated). Indicated as X.XX *1 *SERVC [YES] 42 *CODE *2 [ENT] X.XX *REV *3 [ENT] F0526.ai *1: See "Figure 5.3 Process at the Setting Level" on page 5-10. *2: Enter the code number. *3: Pressing the [ENT] key returns to the service level with the indication [*SERVC]. This parameter does not accept the [ ] and [ ] key input. Figure 5.
<5. DENSITY / PARAMETER SETTING> 5-31 CODE 44: Setting of PASSWORD [ *PASS] This setting defines the password. *1 *SERVC [YES] 44 *CODE *2 [ENT] X.X.X *PASS *3 [ENT] F0528.ai *1: See "Figure 5.3 Process at the Setting Level" on page 5-10. *2: Enter the code number. *3: Set a value by using the [ ], [ ] and [ENT] keys. X.X.
<5. DENSITY / PARAMETER SETTING> 5-32 CODE 50: Selecting the instruments [ *MODEL] This setting defines the model. ・Density meter: 0 ・Calorie meter: 1 ・Hydrogen purity, replacement meter : 2 *1 *SERVC [YES] 50 *CODE *2 [ENT] X *MODEL *3 [ENT] F0530.ai *1: See "Figure 5.3 Process at the Setting Level" on page 5-10. *2: Enter the code number. *3: Set a value by using the [ ], [ ] and [ENT] keys. Figure 5.
<5. DENSITY / PARAMETER SETTING> *1 [YES] *SERVC 82 *CODE 5-33 *2 [ENT] 020 *PASSW *4 [ENT] XXXX *K_A_H *3 [ENT] XXXX *K_A_L *3 [ENT] XXXX *K_B_H *3 [ENT] XXXX *T2L_H *3 [ENT] XXXX *T2L_L *3 [ENT] XXXX *T3L_H *3 [ENT] XXXX *T3L_L *3 [ENT] XXXX *3 *K_B_L [ENT] XXXX *3 *T1H_H [ENT] XXXX *K_C_H XXXX *T1H_L *3 [ENT] XXXX *K_C_L *3 [ENT] XXXX *K_AH.H *3 [ENT] XXXX *K_AH.L *3 [ENT] XXXX *K_BH.H *3 [ENT] XXXX *K_BH.L *3 [ENT] XXXX *K_CH.H *3 [ENT] XXXX *K_CH.
Blank Page
6. 6-1 <6. CALORIE / PARAMETER SETTING> CALORIE / PARAMETER SETTING CAUTION Password is not selected in first condition. When password is necessary, refer to Section “4.2 Setting Lists”. This chapter describes how to set parameters of Calorie meter. If select Density meter, see Chapter 5. If select Hydrogen purity meter, see Chapter 7. Conversion Table 1 lb/ft3 = 16.01847 kg/m3 1 kBTU/ft3 = 37.25901232 MJ/m3 X degF = (X-32)/1.8 1 psi = 6.
6.1 6.1.1 6-2 <6. CALORIE / PARAMETER SETTING> Setting Parameters Subsections 6.1.1 to 6.1.4 show the setting parameters for each level. Setting Parameters at Measurement Level When turned on, the analyzer starts up in the measurement mode (). Table 6.
6.1.2 6-3 <6. CALORIE / PARAMETER SETTING> Setting Parameters at Operation Level Press the [MODE] key and enter the password (XXX) to gain access to this level (see Figure 6.1 for key operation). Table 6.
6.1.3 <6. CALORIE / PARAMETER SETTING> 6-4 Setting Parameters at Setting Level Press the [*] key and enter the password (XXX) to gain access to this level (see Figure 6.5 for key operation). Table 6.
6.1.4 <6. CALORIE / PARAMETER SETTING> 6-5 Setting Parameters at Service Level Press the [*] key and enter the password (XXX) to select the service level. When you select this level, the analyzer shows a Code No. promptly. Type the appropriate code number and press the [ENT] key. Table 6.4 Setting Parameters at Service Level Code No.
<6. CALORIE / PARAMETER SETTING> 6-6 Table 6.4 Setting Parameters at Service Level Code No. Item Data to be Set (or Conditions) Display Remarks 20 Selection of pressure unit *PRES.U · kPa : 0 · MPa : 1 · psi : 2 See Figure 6.18 for key operation. 21 Selection of density unit *DENS.U · kg/m3 : 0 · lb/ft3 : 1 See Figure 6.19 for key operation. · MJ/m3 : 0 · kBTU/ft3 : 1 See Figure 6.20 for key operation. See Figure 6.21 for key operation. 22 Selection of calorific value unit *CAL.
6.2 6-7 <6. CALORIE / PARAMETER SETTING> Parameter Setting [Aborting the Setting Operation] Press the [MODE] key. Normally, the [MODE] key is used to move from the measurement () mode to a mode at the operation level. It is also used to return to the measurement mode from other modes. 6.2.1 Parameter Setting at Operation Level To move to the operation level, press the [MODE] key and then enter the password (XXX). Password is not selected in first condition.
<6. CALORIE / PARAMETER SETTING> 6-8 (1) Operation Level Select a mode from the following four choices of the operation level. Each press of the [NO] key cycles through the choices in the order shown below: 1. Display mode [DISP] 2. Semi-automatic calibration mode [SEM.CAL] 3. Manual calibration mode [MAN.CAL] In the mode, “Density / Calorie Conversion Factor Setting Mode” is included in next page. 4. Valve operation mode [VALVE] Press the [YES] key to enter the desired mode when it is indicated.
<6. CALORIE / PARAMETER SETTING> 6-9 (2) Density/Calorie Conversion Factor Setting Mode In order to this mode, select “MAN.CAL” in before page, and press [YES] key. This mode is used to set a factor for converting density to calorie. This setting is mandatory when measuring the calorie. (a) Automatic zero-point adjustment [AT_ADJ] : 0 to 133.000, 0 to 3.
<6. CALORIE / PARAMETER SETTING> 6-10 (3) Display Mode Select one of the following seven parameters to be shown in the message field. Each press of the [NO] key cycles through the choices in the order shown below: 1. Physical density [XX.XXXX KG/M3] or [XX.XXXX LB/FT3] (depends on the setting of CODE 21) 2. Compensated density [X.XXXX KG/M3 ]or [X.XXXXX LB/FT3] (depends on the setting of CODE 21) 3. Calorific value [XXX.XXX MJ/M3] or [XX.XXXX KBTU] (depends on the setting of CODE 22) 4.
<6. CALORIE / PARAMETER SETTING> *1 Measurement Mode [MODE] 6-11 DISP [YES] XX.XXXX [YES] * 2, *3 KG/M3 (Physical density) [NO] 2, 3 X.XXXX [YES] * * KG/M3 (Compensated density) [NO] 2, 3 XXX.XXX [YES] * * MJ/M3 (Calorific value) [NO] 2 [YES] * XXX.X (Read-only) °C (Sample gas temperature) [NO] 2 XXX.XX [YES] * (Read-only) KPA (Sample gas pressure) [NO] [YES] XXX.X (Read-only) MA1% (Current for analog output1(%)) [NO] [YES] XXX.
6.2.2 6-12 <6. CALORIE / PARAMETER SETTING> Parameter Setting at Setting Level To move to the setting level, press the [*] key and then enter the password (XXX). The main task at the setting level is to set data values such as the measuring range values. Note that the function for which you have set data values at the setting level, will not work if that function is turned off at another level. For this reason, care must be taken when dealing with modes that relate to each other.
<6. CALORIE / PARAMETER SETTING> 6-13 1. Analog Output Setting Mode [ *RANGE ] The analog output has output 1 and output 2. Only output 1 can be used for communication purposes. Set a range appropriate for 4-20 mA DC output signals. The range has two setpoints: the lower limit (zero point) of the range corresponding to the minimum (0%) of a given output signal and the upper limit (span point) of the range corresponding to the maximum (100%) of the given output signal.
<6. CALORIE / PARAMETER SETTING> *1 6-14 Measurement mode XXX [*] *PASSW [ENT] *SERVC [NO] *RANGE [YES] *OUT1 [YES] *2 *DENS (Physical density Range) [NO] *3 [NO] XX.XXXX [YES] *Z_DNS *CAL.DT *OUT2 [YES] XX.XXXX [ENT] *S_DNS [ENT] [NO] *C_DNS (Compensated density Range) [YES] [NO] *CALRY (Calorific value Range) X.XXXX *Z_CP.D X.XXXX [ENT] *Z_CAL [ENT] XXX.XXX XXX.XXX [YES] *S_CP.
<6. CALORIE / PARAMETER SETTING> 6-15 2. Alarm-point Setting Mode [*ALARM] Set the upper /lower limits for the measured values of the parameters noted below, for the purpose of raising an alarm. (a) Select a parameter for which alarms are raised. ・Physical density [*DENS] ・Compensated density [*C_DNS] ・Calorific value [*CALRY] Press the [YES] key to enter the desired mode when it is indicated. (b) Set the upper/lower limits. ・Physical density [*L_DNS], [*H_DNS] Configure range: 00.
6.2.3 <6. CALORIE / PARAMETER SETTING> 6-16 Parameter Setting at Service Level To move to the service level, select [*SERVC] at the setting level, and press the [YES] key. (see Figure 6.4 on page 6-12.) The main task at the setting level is to set data values such as a measuring range. To go into each mode of the service level, enter the appropriate code number. CAUTION DO NOT enter codes other than those listed in Table 6.4.
<6. CALORIE / PARAMETER SETTING> 6-17 CODE 01: Setting of Hold during Maintenance This setting determines whether or not the output signal is held during maintenance. As a value to be held, you can select either a value immediately before or a preset value. (a) Hold enable/disable [*M_HLD] ・Disable: 0 ・Enable: 1 ・Enable (preset value): 2 (b) Setting of preset hold value [*PR.SET] This item applies if item (a) above is set to "2". Configurable range: *1 *SERVC [YES] -10.
<6. CALORIE / PARAMETER SETTING> 6-18 CODE 02: Setting of Hold in the Event of Errors This setting determines whether or not the output signal is held if an error occurs. You can select either the value immediately prior to the error or a preset value as the value to be held. (a) Hold enable/disable [*E_HLD] ・Disable: 0 ・Enable: 1 ・Enable (preset value): 2 (b) Setting of preset hold value [*PR.SET] This item applies if item (a) above is set to "2".
<6. CALORIE / PARAMETER SETTING> 6-19 CODE 05: Setting of Contact Outputs Status [*CNTCT] This setting defines the status of the contact outputs. Configurable range: 00 to 15 *1 *SERVC [YES] 05 *CODE *2 [ENT] XX *CNTCT *3 [ENT] F0611.ai *1: See "Figure 6.4 Process at the Setting Level" on page 6-12. *2: Enter the code number. *3: Set a value using the [ ], [ ] and [ENT] keys. If you enter a value that exceeds the limits, the display shows [*OVER].
<6. CALORIE / PARAMETER SETTING> 6-20 CODE 10: Setting of Pressure Compensation This setting determines whether or not the measured density should be compensated by the gas pressure. The density can be compensated to either the value immediately before the error or a fixed value. (a) Enable/disable/select preset value [*P.COMP] ・Disable: 0 ・Enable (value immediately before): 1 ・Enable (fixed value): 2 (b) Setting of fixed value for pressure compensation [*P.
<6. CALORIE / PARAMETER SETTING> 6-21 CODE 11: Setting for Compensated Density Measurement This setting defines the reference temperature and pressure required to obtain the compensated density. (a) Setting of reference temperature [*C.D.TMP] Configurable range: -20.0 to 80.0 (for ºC) -4.0 to 176.0 (for °F ) (b) Setting of reference pressure [*C.D.PRS] Configurable range: 0.10 to 999.99 (for kPa), 0.0001 to 9.9999 (for MPa), 0.015 to 99.
<6. CALORIE / PARAMETER SETTING> 6-22 CODE 12: Setting of Gas Pressure This setting defines the zero and span points of the pressure transmitter's measurement range. (a) Setting of zero-point gas pressure [*Z_PRS] Configurable range: 0.01 to 999.99 (for kPa), 0.0001 to 9.9999 (for MPa), 0.015 to 99.999 (for psi ) (b) Setting of span-point gas pressure [*S_PRS] Configurable range: 0.10 to 999.99 (for kPa), 0.0001 to 9.9999 (for MPa), 0.015 to 99.
6-23 <6. CALORIE / PARAMETER SETTING> CODE 13: Selection for Automatic Calibration [*AUTO.C] This setting determines whether or not automatic calibration is carried out. If automatic calibration is to be carried out, you must also configure CODE 15 at the service level. . ・Disable: 0 ・Enable: 1 *1 *SERVC [YES] 13 *CODE *2 [ENT] X *AUTO.C *3 [ENT] F0615 *1: See "Figure 6.4 Process at the Setting Level" on page 6-12. *2: Enter the code number.
<6. CALORIE / PARAMETER SETTING> 6-24 CODE 15: Setting of Data for Automatic/Semi-automatic Calibration This setting defines the data needed to carry out automatic/semi-automatic calibration. (a) Selection of calibration item [*CAL.P] ・Zero and span: 0 ・Zero: 1 ・Span: 2 (b) Setting of calibration time [*CAL.T] Configurable range: 00 to 60 (unit : minute) (c) Setting of stabilization time [*STAB.
6-25 <6. CALORIE / PARAMETER SETTING> *1 *SERVC [YES] 15 *CODE *2 [ENT] X *CAL.P *2 (Selection of calibration item) [ENT] XX *CAL.T *2 (Setting of calibration time) [ENT] XX *STAB.T *2 (Setting of stabilization time) [ENT] XX.XX.XX *Y_M_D *2 (Setting of calibration starting time [year/month/day] ) [ENT] XX.XX *H_M *2 (Setting of calibration starting time [hour/minute] ) [ENT] X *CYCL.U *2 (Selection of unit of calibration interval) [ENT] XXX *CYCL.
<6. CALORIE / PARAMETER SETTING> 6-26 CODE 20: Selection of Pressure Unit [*PRES.U] This setting defines the pressure unit for the gas pressure transmitter. ・kPa: 0 ・MPa: 1 ・psi : 2 *1 [YES] *SERVC 20 *CODE *2 [ENT] X *PRES.U *3 [ENT] F0618.ai *1: See "Figure 6.4 Process at the Setting Level" on page 6-12. *2: Enter the code number. *3: Set a value using the [ ], [ ] and [ENT] keys. Figure 6.18 Selection of Pressure Unit CODE 21: Selection of Density Unit [*DENS.
<6. CALORIE / PARAMETER SETTING> 6-27 CODE 22: Selection of Calorific Value Unit [ *CAL.U] This setting defines the calorific value unit for the gas density meter. • MJ/m3 : 0 • kBTU/ft3 : 1 *1 *SERVC [YES] 22 *CODE *2 [ENT] X *CAL.U *3 [ENT] F0620.ai *1: See "Figure 6.4 Process at the Setting Level" on page 6-12. *2: Enter the code number. *3: Set a value using the [ ], [ ] and [ENT] keys. Figure 6.20 Selection of Calorific Value Unit CODE 23: Selection of Temperature Unit [ *TEMP.
6-28 <6. CALORIE / PARAMETER SETTING> CODE 30: Setting of Calendar This setting defines the date and time. (a) Setting of year/month/day [*Y_M_D] Configurable range : 00.01.01 to 99.12.31 (b) Setting of hour/minute [*H_M] Configurable range : *1 [YES] *SERVC 00.00 to 23.59 30 *CODE *2 [ENT] XX.XX.XX *Y_M_D *3 [ENT] XX.XX *H_M *3 [ENT] F0622.ai *1: See "Figure 6.4 Process at the Setting Level" on page 6-12. *2: Enter the code number.
6-29 <6. CALORIE / PARAMETER SETTING> CODE 40: Indication of Calibration Coefficients (Read-only) This setting determines the calibration coefficients to be indicated. (a) Zero [*C_K_Z]: indicated as SX.XXXX (b) Span [*C_K_S]: indicated as X.XXXX *1 *SERVC [YES] 40 *2 *CODE [ENT] SX.XXXX *C_K_Z *3 [ENT] X.XXXX *C_K_S [ENT] F0624.ai *1: See "Figure 6.4 Process at the Setting Level" on page 6-12. *2: Enter the code number.
<6. CALORIE / PARAMETER SETTING> 6-30 CODE 42: Indication of Software Version [*REV] (Read-only) This setting allows the software version to be confirmed (indicated). Indicated as X.XX *1 *SERVC [YES] 42 *CODE *2 [ENT] X.XX *REV *3 [ENT] F0626.ai *1: See "Figure 6.4 Process at the Setting Level" on page 6-12. *2: Enter the code number. *3: Pressing the [ENT] key returns to the service level with the indication [*SERVC]. This parameter does not accept the [ ] and [ ] key input. Figure 6.
<6. CALORIE / PARAMETER SETTING> 6-31 CODE 44: Setting of PASSWORD [ *PASS] This setting defines the password. *1 *SERVC [YES] 44 *CODE *2 [ENT] X.X.X *PASS *3 [ENT] F0628.ai *1: See "Figure 6.4 Process at the Setting Level" on page 6-12. *2: Enter the code number. *3: Set a value by using the [ ], [ ] and [ENT] keys. X.X.
<6. CALORIE / PARAMETER SETTING> 6-32 CODE 50: Selecting the instruments [ *MODEL] This setting defines the model. ・Density meter: 0 ・Calorie meter: 1 ・Hydrogen purity, replacement meter : 2 *1 *SERVC [YES] 50 *CODE *2 [ENT] X *MODEL *3 [ENT] F0630.ai *1: See "Figure 6.4 Process at the Setting Level" on page 6-12. *2: Enter the code number. *3: Set a value by using the [ ], [ ] and [ENT] keys. Figure 6.
<6. CALORIE / PARAMETER SETTING> [YES] *1 *SERVC 82 *CODE 6-33 *2 [ENT] 020 *PASSW *4 [ENT] XXXX *K_A_H *3 [ENT] XXXX *K_A_L *3 [ENT] XXXX *K_B_H *3 [ENT] XXXX *T2L_H *3 [ENT] XXXX *T2L_L *3 [ENT] XXXX *T3L_H *3 [ENT] XXXX *T3L_L *3 [ENT] XXXX *3 *K_B_L [ENT] XXXX *3 *T1H_H [ENT] XXXX *K_C_H XXXX *T1H_L *3 [ENT] XXXX *K_C_L *3 [ENT] XXXX *K_AH.H *3 [ENT] XXXX *K_AH.L *3 [ENT] XXXX *K_BH.H *3 [ENT] XXXX *K_BH.L *3 [ENT] XXXX *K_CH.H *3 [ENT] XXXX *K_CH.
Blank Page
7. 7-1 <7. HYDROGEN PURITY / PARAMETER SETTING> HYDROGEN PURITY/PARAMETER SETTING CAUTION Password is not selected in first condition. When password is necessary, refer to Section “4.2 Setting Lists”. This chapter describes how to set parameters of Hydrogen purity meter. If select Density meter, see Chapter 5. If select Calorie meter, see Chapter 6. Conversion Table 1 lb/ft3 = 16.01847 kg/m3 1 kBTU/ft3 = 37.25901232 MJ/m3 X degF = (X-32)/1.8 1 psi = 6.
7.1 7.1.1 7-2 <7. HYDROGEN PURITY / PARAMETER SETTING> Setting Parameters Subsections 7.1.1 to 7.1.4 show the setting parameters for each level. Setting Parameters at Measurement Level When turned on, the analyzer starts up in the measurement mode (). Table 7.
7.1.3 7-3 <7. HYDROGEN PURITY / PARAMETER SETTING> Setting Parameters at Setting Level Press the [*] key and enter the password (XXX) to gain access to this level (see Figure 7.4 for key operation). Table 7.
7.1.4 <7. HYDROGEN PURITY / PARAMETER SETTING> 7-4 Setting Parameters at Service Level Press the [*] key and enter the password (XXX) to select the service level. When you select this level, the analyzer shows a Code No. promptly. Type the appropriate code number and press the [ENT] key. Table 7.4 Setting Parameters at Service Level Code No.
<7. HYDROGEN PURITY / PARAMETER SETTING> 7-5 Table 7.4 Setting Parameters at Service Level Code No.
7.2 <7. HYDROGEN PURITY / PARAMETER SETTING> 7-6 Parameter Setting [Aborting the Setting Operation] Press the [MODE] key. Normally, the [MODE] key is used to move from the measurement () mode to a mode at the operation level. It also used to return to the measurement mode from other modes. 7.2.1 Parameter Setting at Measurement Level 1.
7.2.2 7-7 <7. HYDROGEN PURITY / PARAMETER SETTING> Parameter Setting at Operation Level To move to the operation level, press the [MODE] key and then enter the password (XXX). Password is not selected in first condition. When password is necessary, refer to Section 4.2 “Setting Lists” and selected password what you need. This level includes the manual calibration mode, which is discussed in Chapter 10. (1) Operation Level Select a mode from the following four choices of the operation level.
<7. HYDROGEN PURITY / PARAMETER SETTING> 7-8 2. Display Mode Select one of the following seven parameters to be shown in the message field. Each press of the [NO] key cycles through the choices in the order shown below: 1. Physical density [XX.XXXX KG/M3] or [XX.XXXXX LB/FT3] (depends on the setting of CODE 21) 2. Compensated density [X.XXXX KG/M3 ]or [X.XXXXX LB/FT3] (depends on the setting of CODE 21) 3. Concentration [XXX.X VOL%] 4. Temperature [XXX.X °C] or [XXX.
<7. HYDROGEN PURITY / PARAMETER SETTING> *1 Measurement Mode 7-9 DISP [YES] XX.XXXX [YES] * 2,* 3 KG/M3 (Physical density) [NO] X.XXXX [YES] * 2,* 3 KG/M3 (Compensated density) [NO] XXX.X H2_AIR [YES] * 2,* 3 (Hydrogen purity) [NO] XXX.X °C [YES] * 2 (Read-only) (Sample gas temperature) [NO] XXX.XX [YES] * 2 (Read-only) KPA (Sample gas pressure) [NO] [YES] XXX.X (Read-only) MA1% (Current for analog output 1(%)) [NO] [YES] XXX.
7.2.3 7-10 <7. HYDROGEN PURITY / PARAMETER SETTING> Parameter Setting at Setting Level To move to the setting level, press the [*] key and then enter the password (XXX). The main task at the setting level is to set data values such as the measuring range values. Note that the function for which you have set data values at the setting level, will not work if that function is turned off at another level. For this reason, care must be taken when dealing with modes that relate to each other.
7-11 <7. HYDROGEN PURITY / PARAMETER SETTING> 1. Analog Output Setting Mode [ *RANGE ] The analog output has output 1 and output 2. Only output 1 can be used for communication purposes. Set a range appropriate for 4-20 mA DC output signals. The range has two setpoints: the lower limit (zero point) of the range corresponding to the minimum (0%) of a given output signal and the upper limit (span point) of the range corresponding to the maximum (100%) of the given output signal.
<7. HYDROGEN PURITY / PARAMETER SETTING> 7-12 Measurement mode *1 XXX [*] *PASSW [ENT] *ALARM [NO] [NO] *SERVC *RANGE [NO] *CAL.DT [NO] [YES] *OUT1 [YES] *2 *DENS *3 [NO] *OUT2 [YES] [YES] XX.XXXX *Z_DNS [ENT] XX.XXXX *S_DNS [ENT] [YES] X.XXXX *Z_CP.D [ENT] X.XXXX *S_CP.D [ENT] [YES] XXX.X *Z_H_A [ENT] XXX.X *S_H_A [ENT] [YES] XXX.X *Z_H.C.A [ENT] XXX.X *S_H.C.
<7. HYDROGEN PURITY / PARAMETER SETTING> 7-13 2. Alarm-point Setting Mode [*ALARM] Set the upper /lower limits for the measured values of the parameters noted below, for the purpose of raising an alarm. (a) Select a parameter for which alarms are raised. ・Physical density [*DENS] ・Compensated density [*C_DNS] ・Hydrogen purity [*H_A] Press the [YES] key to enter the desired mode when it is indicated. (b) Set the high/low limits.
7.2.4 <7. HYDROGEN PURITY / PARAMETER SETTING> 7-14 Parameter Setting at Service Level To move to the service level, select [*SERVC] at the setting level, and press the [YES] key. (see Figure 7.4 on page 7-10.) The main task at the setting level is to set data values such as a measuring range. To go into each mode of the service level, enter the appropriate code number. CAUTION DO NOT enter codes other than those listed in Table 7.4.
<7. HYDROGEN PURITY / PARAMETER SETTING> 7-15 CODE 01: Setting of Hold during Maintenance This setting determines whether or not the output signal is held during maintenance. As a value to be held, you can select either a value immediately before or a preset value. (a) Hold enable/disable [*M_HLD] ・Disable: 0 ・Enable: 1 ・Enable (preset value): 2 (b) Setting of preset hold value [*PR.SET] This item applies if item (a) above is set to "2".
<7. HYDROGEN PURITY / PARAMETER SETTING> 7-16 CODE 02: Setting of Hold in the Event of Errors This setting determines whether or not the output signal is held if an error occurs. You can select either the value immediately prior to the error or a preset value as the value to be held. (a) Hold enable/disable [*E_HLD] ・Disable: 0 ・Enable: 1 ・Enable (preset value): 2 (b) Setting of preset hold value [*PR.SET] This item applies if item (a) above is set to "2".
<7. HYDROGEN PURITY / PARAMETER SETTING> 7-17 (b) Setting of preset hold value [*PR.SET] This item applies if item (a) above is set to “2”. Configurable range: *1 *SERVC -10.0 to 110.0 (percent of analog output) [YES] 03 *CODE *2 [ENT] X *H_HLD *5 *3 [ENT] XXX.X *PR.SET *4 [ENT] F0710.ai *1: See "Figure 7.4 Process at the Setting Level" on page 7-10. *2: Enter the code number. *3: Set a value using the [ ] and [ENT] keys. *4: Set a value using the [ ], [ ] and [ENT] keys.
<7. HYDROGEN PURITY / PARAMETER SETTING> 7-18 CODE 04: Setting of Output-smoothing Constants [*SMOTH] This setting defines the constants for output smoothing. This smoothing constants is output-smoothing constants of converter (electrical circuit boards). Configurable range: *1 *SERVC [YES] 00 to 60. (unit : second) 04 *CODE *2 [ENT] XX *SMOTH *3 [ENT] F0711.ai *1: See "Figure 7.4 Process at the Setting Level" on page 7-10. *2: Enter the code number.
<7. HYDROGEN PURITY / PARAMETER SETTING> 7-19 Each contact output takes either of the following two status (in table); depending on the value you set.
<7. HYDROGEN PURITY / PARAMETER SETTING> 7-20 CODE 10: Setting of Pressure Compensation This setting determines whether or not the measured density should be compensated by the gas pressure. The density can be compensated either by receiving a signal from a pressure transmitter or by using a fixed pressure setpoint.
<7. HYDROGEN PURITY / PARAMETER SETTING> 7-21 CODE 11: Setting for Compensated Density Measurement This setting defines the reference temperature and pressure required to obtain the compensated density. (a) Setting of reference temperature [*C.D.TMP] Configurable range: -20.0 to 80.0 (for ºC) -4.0 to 176.0 (for °F ) (b) Setting of reference pressure [*C.D.PRS] Configurable range: 0.10 to 999.99 (for kPa), 0.0001 to 9.9999 (for MPa), 0.015 to 99.
<7. HYDROGEN PURITY / PARAMETER SETTING> 7-22 CODE 12: Setting of Gas Pressure This setting defines the zero and span points of the pressure transmitter's measurement range. (a) Setting of zero-point gas pressure [*Z_PRS] Configurable range: 0.01 to 999.99 (for kPa), 0.0001 to 9.9999 (for MPa), 0.015 to 99.999 (for psi ) (b) Setting of span-point gas pressure [*S_PRS] Configurable range: 0.10 to 999.99 (for kPa), 0.0001 to 9.9999 (for MPa), 0.015 to 99.
<7. HYDROGEN PURITY / PARAMETER SETTING> 7-23 CODE 14: Setting for Remote Semi-automatic Calibration [*REMOT] This setting determines whether or not the measuring range of the replacement (substituentsconcentration) meter is selected remotely. If remote range selection is to be carried out, you must wire the contact input. ・Disable: 0 ・Enable: 1 If the code is set to “1”, the analyzer selects: • the “AIR in CO2” range, if the contact input is open.
<7. HYDROGEN PURITY / PARAMETER SETTING> 7-24 CODE 21: Selection of Density Unit [*DENS.U] This setting defines the density unit for the gas density meter. • kg/m3: 0 • lb/ft3: 1 *1 [YES] *SERVC 21 *CODE *2 [ENT] X *DENS.U *3 [ENT] F0718.ai *1: See "Figure 7.4 Process at the Setting Level" on page 7-10. *2: Enter the code number. *3: Set a value using the [ ], [ ] and [ENT] keys. Figure 7.18 Selection of Density Unit CODE 23: Selection of Temperature Unit [ *TEMP.
7-25 <7. HYDROGEN PURITY / PARAMETER SETTING> CODE 31: Showing/Hiding of Negative Measured Values [*MINUS] This setting determines whether negative measured values (-) should be shown or hidden. ・Show: 0 ・Hide: 1 *1 [YES] *SERVC 31 *2 *CODE [ENT] X *3 *MINUS [ENT] F0720.ai *1: See "Figure 7.4 Process at the Setting Level" on page 7-10. *2: Enter the code number. *3: Set a value using the [ ], [ ] and [ENT] keys. Figure 7.
7-26 <7. HYDROGEN PURITY / PARAMETER SETTING> CODE 41: Indication of Oscillation Frequency (Read-only) This setting determines the F2 and F4 oscillation frequencies coefficients to be indicated. (a) Indication of frequency F2 (kHz) [*F2.KHZ]: indicated as X.XXXXX (b) Indication of frequency F4 (kHz) [*F4.KHZ]: indicated as X.XXXXX (c) Indication of frequency ratio (kHz) [*F2/F4]: indicated as X.XXXXX *1 *SERVC [YES] 41 *CODE *2 [ENT] X.XXXXX *F2.KHZ *3 [ENT] X.XXXXX *F4.
<7. HYDROGEN PURITY / PARAMETER SETTING> 7-27 CODE 43: Setting of High-resolution Mode [*S_CYC] This setting selects between the normal mode and the high-resolution mode. ・Normal mode: 0 ・High-resolution mode: 1 *1 *SERVC [YES] 43 *CODE *2 [ENT] X *S_CYC *3 [ENT] F0724.ai *1: See "Figure 7.4 Process at the Setting Level" on page 7-10. *2: Enter the code number. *3: Set a value by using the [ ], [ ] and [ENT] keys. Figure7.
<7. HYDROGEN PURITY / PARAMETER SETTING> 7-28 CODE 45: Selecting Battery alarm detection [ *BAT] This setting determines whether or not battery alarm is detection. ・Non detection: 0 ・Detection: *1 *SERVC 1 [YES] 45 *CODE *2 [ENT] X *BAT *3 [ENT] F0726 *1: See "Figure 7.4 Process at the Setting Level" on page 7-10. *2: Enter the code number. *3: Set a value by using the [ ], [ ] and [ENT] keys. Figure 7.
<7. HYDROGEN PURITY / PARAMETER SETTING> 7-29 CODE 82: Setting Detector Constant This setting defines the detector constant. CAUTION The instruments may not only work improperly but also big problems occur unless the detector constants entered incorrectly. Do carefully and precisely when enter the detector constants into converter. 1. The GD402 gas density meter is shipped after adjusting the detector and converter in pairs.
<7. HYDROGEN PURITY / PARAMETER SETTING> [YES] *1 *SERVC 82 *CODE 7-30 *2 [ENT] 020 *PASSW *4 [ENT] XXXX *K_A_H *3 [ENT] XXXX *K_A_L *3 [ENT] XXXX *K_B_H *3 [ENT] XXXX *T2L_H *3 [ENT] XXXX *T2L_L *3 [ENT] XXXX *T3L_H *3 [ENT] XXXX *T3L_L *3 [ENT] XXXX *3 *K_B_L [ENT] XXXX *3 *T1H_H [ENT] XXXX *K_C_H XXXX *T1H_L *3 [ENT] XXXX *K_C_L *3 [ENT] XXXX *K_AH.H *3 [ENT] XXXX *K_AH.L *3 [ENT] XXXX *K_BH.H *3 [ENT] XXXX *K_BH.L *3 [ENT] XXXX *K_CH.H *3 [ENT] XXXX *K_CH.
8. <8. DENSITY / CALIBRATION PROCEDURE> 8-1 DENSITY /CALIBRATION PROCEDURE This chapter describes how to calibrate of Density meter. If select Calorie meter, see Chapter 9. If select Hydrogen purity meter, see Chapter 10. Zero and span calibration can be carried out using standard gases (zero and span gases). Three modes are available for calibration: automatic, semi-automatic (remote semi-automatic) and manual.
<8. DENSITY / CALIBRATION PROCEDURE> 8-2 (2) Operations for Setting the Automatic/Remote Semi-automatic Calibration • Selection for automatic calibration [*AUTO.C] At the service level, select CODE 13. • Disable: 0 • Enable: 1 • Selection for remote semi-automatic calibration [*REMOT] At the service level, select CODE 14. • Disable: 0 • Enable: 1 (3) Valve Check Check the performance of valves according to the “valve operation flowchart” at the operation level. *1 MAN.
8.2.2 <8. DENSITY / CALIBRATION PROCEDURE> 8-3 Setting Calibration Data To set calibration data, press the [*] key in the measurement mode, (when password is selected, enter the password (XXX), ) and select the parameter [*CAL.DT]. See Figure 5.4 for the process at the setting level. (a) Zero-point density [*Z_DNS] Configurable range : 0.0000 to 6.0000 (kg/m3) 0.00000 to 0.40000 (lb/ft3) (b) Span-point density [*S_DNS] Configurable range : 0.0000 to 6.0000 (kg/m3) 0.
<8. DENSITY / CALIBRATION PROCEDURE> 8.3 Calibration 8.3.1 Semi-automatic Calibration 8-4 In this mode of calibration, you can carry out one-touch calibration. To use this feature, you must observe the following instructions. • The analyzer needs solenoid valves for controlling the calibration gases (zero and span gases). Connect the solenoid valves with utmost care to ensure that there is no gas leakage in the system.
<8. DENSITY / CALIBRATION PROCEDURE> 8.3.2 8-5 Manual Calibration In this mode of calibration, you visually make sure the reading has settled, and then confirm the reading manually (by pressing the [ENT] key). There are two ways to switch between the zero, span and sample gases: by using manually-operated valves or by using solenoid valves. When switching between the gases using solenoid valves, follow the instructions on valve installation in Subsection 8.3.1. *1 SEM.CAL [NO] [NO] MAN.
8.3.3 <8. DENSITY / CALIBRATION PROCEDURE> 8-6 Automatic Calibration This mode of calibration takes effect when you select “enable” for automatic calibration in CODE 13 - [*AUTO.C] - at the service level. Even during automatic calibration, you can interrupt the system with the semi-automatic or manual mode of calibration. One cycle of automatic calibration is skipped, however, if the time of automatic calibration arrives when semi-automatic calibration is in progress. Measurement mode X.
9. <9. CALORIE / CALIBRATION PROCEDURE> 9-1 CALORIE /CALIBRATION PROCEDURE This chapter describes how to calibrate of Calorie meter. If select Density meter, see Chapter 8. If select Hydrogen purity meter, see Chapter 10. Zero and span calibration can be carried out using standard gases (zero and span gases). Three modes are available for calibration: automatic, semi-automatic (remote semi-automatic) and manual.
<9. CALORIE / CALIBRATION PROCEDURE> 9-2 (2) Operations for Setting the Automatic/Remote Semi-automatic Calibration • Selection for automatic calibration [*AUTO.C] At the service level, select CODE 13. • Disable: 0 • Enable: 1 • Selection for remote semi-automatic calibration [*REMOT] At the service level, select CODE 14. • Disable: 0 • Enable: 1 (3) Valve Check Check the performance of valves according to the “valve operation flowchart” at the operation level. *1 MAN.
9.2.2 <9. CALORIE / CALIBRATION PROCEDURE> 9-3 Setting Calibration Data To set calibration data, press the [*] key in the measurement mode, (when password is selected, enter the password (XXX), ) and select the parameter [*CAL.DT]. See Figure 6.4 for the process at the setting level. (a) Zero-point calorific value [*Z_CAL] Configurable range : 000.000 to 133.000 (MJ/m3) 0.0000 to 3.5000 (kBTU/ft3) (b) Zero-point density [*Z_DNS] Configurable range : 0.0000 to 6.
<9. CALORIE / CALIBRATION PROCEDURE> *1 *RANGE [NO] *CAL.DT [NO] 9-4 *ALARM [YES] XXX.XXX *Z_CAL * 2 (Setting of Zero-point calorific value) [ENT] X.XXXX *Z_DNS * 2 (Setting of Zero-point density) [ENT] XXX.XXX *S_CAL * 2 (Setting of Span-point calorific value) [ENT] X.XXXX *S_DNS * 2 (Setting of Span-point density) [ENT] X *C_HLD *2 (Setting of Output hold value during calibration) [ENT] XXX.X *PR.SET *3 *2 (Setting of preset value) [ENT] F0903.ai *1: See "Figure 6.
<9. CALORIE / CALIBRATION PROCEDURE> 9.3 9-5 Calibration 9.3.1 Semi-automatic Calibration In this mode of calibration, you can carry out one-touch calibration. To use this feature, you must observe the following instructions. • The analyzer needs solenoid valves for controlling the calibration gases (zero and span gases). Connect the solenoid valves with utmost care to ensure that there is no gas leakage in the system.
<9. CALORIE / CALIBRATION PROCEDURE> 9.3.2 9-6 Manual Calibration In this mode of calibration, you visually make sure the reading has settled, and then confirm the reading manually (by pressing the [ENT] key). There are two ways to switch between the zero, span and sample gases: by using manually-operated valves or by using solenoid valves. When switching between the gases using solenoid valves, follow the instructions on valve installation in Subsection 9.3.1. *1 SEM.CAL [NO] MAN.
9.3.3 <9. CALORIE / CALIBRATION PROCEDURE> 9-7 Automatic Calibration This mode of calibration takes effect when you select “enable” for automatic calibration in CODE 13 - [*AUTO.C] - at the service level. Even during automatic calibration, you can interrupt the system with the semi-automatic or manual mode of calibration. One cycle of automatic calibration is skipped, however, if the time of automatic calibration arrives when semi-automatic calibration is in progress. Measurement mode X.
Blank Page
10. <10. HYDROGEN PURITY / CALIBRATION PROCEDURE> 10-1 HYDROGEN PURITY /CALIBRATION PROCEDURE This chapter describes how to calibrate of Hydrogen purity meter. If select Density meter, see Chapter 8. If select Calorie meter, see Chapter 9. Zero and span calibration can be carried out using standard gases (zero and span gases). The analyzer uses hydrogen (100%) as the zero gas and carbon dioxide (100%) as the span gas. The analyzer is designed so that calibration is done manually.
10-2 <10. HYDROGEN PURITY / CALIBRATION PROCEDURE> 10.3 Calibration In the manual calibration mode, visually check that the reading has settled, and then confirm the reading manually (by pressing the [ENT] key). *1 DISP [NO] [NO] MAN.CAL [YES] ZERO [NO] S_GAS Close the sample gas valve. [YES] 0.0899 H2 *3 *5 Zero-point density (H 2 gas) is displayed. [ENT] X.XXXX CAL.SET *4 [ENT] *5 [NO] Open the zero gas valve. Press [ENT] key after the value is stabilized.
11. 11-1 <11. INSPECTION AND MAINTENANCE> INSPECTION AND MAINTENANCE To ensure the GD402 analyzer remains highly accurate with excellent operating conditions, routine inspection and maintenance is essential. Refer to the check items in this chapter to determine which items to apply in your inspection and maintenance, and check them periodically. 11.1 11.1.
11.1.4 <11. INSPECTION AND MAINTENANCE> 11-2 Replacing the Fuse (1) For safety reasons, stop the power supply to the converter by switching the external breaker. (2) Remove the fuse from the fuse holder. Turn the holder cap counterclockwise by 90 degrees using a flat-blade screwdriver that fits the holder cap. Then, pull out the fuse together with the cap.
11.2 11-3 <11. INSPECTION AND MAINTENANCE> Inspection In Case of Failure 11.2.1 Inspecting the Analyzer in an Alarm Status The GD402 analyzer outputs an ALARM signal through terminals 16 and 17 if a failure is detected by the self-diagnostic function during measurement. In addition, the lamp on the operation panel comes on and the message field shows an alarm number. The analyzer outputs no contact signal, however, if the alarm number is [ALM.09].
11-4 <11. INSPECTION AND MAINTENANCE> 11.2.2 Inspecting the Analyzer in a FAIL Status IN GD402, the state of terminals 18 and 19 is changed from close to open if a FAIL status is detected by the self-diagnostic function during measurement. In addition, the lamp on the operation panel comes on and the message field shows an error number. If a FAIL status is detected, take the corrective action given in Table 11.2. Table 11.
Customer Maintenance Parts List 6 5 1 GD40T, GD40V, GD40R Gas Density Meter Detector Explosion-proof with Intrinsically Safe 3 2 4 7, 8 9 Item 1 2 3 4 5 6 7 8 9 Part No. Qty F9301DY 1 F9301DQ 1 G9303AM 1 G9303LK 1 G9339AA 2 Description Cover Cover O-ring O-ring Bracket Assembly Y9408ZU F9340AL Y9401WL G9601AN Bolt Screw Toothed Lock Washer Cable Gland Assembly (only GD40R) 2 1 1 1 All right reserved, Copyright © Copyright 1998, Yokogawa Electric Corporation.
Customer Maintenance Parts List GD402G Gas Density Meter Non-Explosion-proof Converter 1 6 2 5 4 3 2 3 4 Part No. Qty 1 A1109EF 1 A1111EF 6 L9811FV K9334CN 6 K9313DW 1 Description Fuse 1.0 A (For 100 to 240 V AC) Fuse 2.0 A (For 24 V DC) Cable Gland Insert Cover Assy 5 6 Y9101XA Y9420LU O-Ring Screw Item 1 4 4 All right reserved, Copyright © Copyright 1998, Yokogawa Electric Corporation. Subject to change without notice CMPL 11T3E1-02E 1st Edition : Apr. 1998 (YK) 3rd Edition : Apr.
Customer Maintenance Parts List 3 GD402T, GD402V, GD402R Gas Density Meter Explosion-proof Converter 2 7 1 4 5 6 2 3 4 Part No. Qty 1 A1109EF A1111EF 1 L9817MH 1 K9292XL 1 L9800GT 1 Description Fuse 1.0 A (For 100 to 240 V AC) Fuse 2.0 A (For 24 V DC) O-Ring Screw Screw 5 6 7 K9214GZ L9811LL K9214BF Plug Cable Gland G3/4 (PF3/4) (only GD402R) Cover Item 1 3 3 1 All right reserved, Copyright © Copyright 1998, Yokogawa Electric Corporation.
i Revision Information Title Manual No. : Model GD402 Gas Density Meter : IM 11T3E1-01E Edition Date 1st Sep. 1998 4th Remark (s) Newly published May 2004 5th Jul. 2006 Revised Section After-sales Warranty: Added description. Added Section 1.1.1.7, “Safety and EMC compliance” Deleted Section 1.4, “Safety Standard” Section 2.1.2, “GD40T (FM Explosion-proof with Intrinsically Safe Approval)”: Correction error in Item 1. Section 2.4.1, Replaced the entire page. Replaced entire Section 2.4.
ii n If you want have more information about Yokogawa products, you can visit Yokogawa’s home page at the following website. Home page: http://www.yokogawa.com/an/ n Written by Environmental & Analytical Products PMK Group IA Div., Product Business Center Yokogawa Electric Corporation n Published by Yokogawa Electric Corporation 2-9-32 Nakacho, Musashino-shi, Tokyo 180-8750, JAPAN n Printed by KOHOKU PUBLISHING & PRINTING INC.