Owner's manual
<6. CALORIE / PARAMETER SETTING>
6-19
IM 11T3E1-01E
CODE 05: Setting of Contact Outputs Status [*CNTCT]
This setting denes the status of the contact outputs.
Congurable range: 00 to 15
[ENT]
05
*CODE
*SERVC
XX
*CNTCT
[ENT]
[YES]
F0611.ai
*
1
*
2
*
3
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
Figure 6.11 Setting of Contact Outputs Status
Each contact output takes either of the following two status (in table); depending on the value
you set.
Value ZERO contact SPAN contact MAINT contact ALM contact
0 NO NO NO NO
1 NO NO NO NC
2 NO NO NC NO
3 NO NO NC NC
4 NO NC NO NO
5 NO NC NO NC
6 NO NC NC NO
7 NO NC NC NC
8 NC NO NO NO
9 NC NO NO NC
10 NC NO NC NO
11 NC NO NC NC
12 NC NC NO NO
13 NC NC NO NC
14 NC NC NC NO
15 NC NC NC NC
NO: means “normally open”
NC: means “normally closed”