Owner's manual
<5. DENSITY / PARAMETER SETTING>
5-19
IM 11T3E1-01E
CODE 05: Setting of Contact Outputs Status [*CNTCT]
This setting denes the status of the contact outputs.
Congurable range: 00 to 15
05
*CODE
*SERVC
XX
*CNTCT
[YES]
F0511.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
[ENT]
[ENT]
*
3
*
2
*
1
Figure 5.10 Setting of Contact Outputs Status
Each contact output takes either of the following two status (in table), depending on the value you
set.
Value ZERO contact SPAN contact MAINT contact ALM contact
0 NO NO NO NO
1 NO NO NO NC
2 NO NO NC NO
3 NO NO NC NC
4 NO NC NO NO
5 NO NC NO NC
6 NO NC NC NO
7 NO NC NC NC
8 NC NO NO NO
9 NC NO NO NC
10 NC NO NC NO
11 NC NO NC NC
12 NC NC NO NO
13 NC NC NO NC
14 NC NC NC NO
15 NC NC NC NC
NO: means “normally open”
NC: means “normally closed”