Owner's manual

<2. Installation, Wiring and Piping>
2-3
IM 11T3E1-01E
2.1.3 GD40V (CSA Explosion-proof and Intrinsically Safe
Approval)
Sealing Fitting
Conduit
500 mm (20") Max.
NON-HAZARDOUS
LOCATIONS
HAZARDOUS
LOCATIONS
GD40V
F0203.ai
Figure 2.2
1. GD40V Gas density detector for use in hazardous (classied) locations:
Explosion-proof for Class I, Division 1, Groups B, C and D;
Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G with Intrinsically
Safe sensor for Class I, II, III, Division 1, Groups B, C, D, E, F and G.
Enclosure Rating: Type 4X
Temperature Code: T5
Ambient Temperature: -10 to 60°C
2. Wiring
All wiring shall comply with Canadian Electrical and Local Electrical codes.
Note a warning label worded as follows.
WARNING
A seal shall be installed within 500 mm (20 inches) of the enclosure. Refer to Figure 2.2.
3. Operation
Note a warning label worded as follows.
WARNING
Open circuit before removing cover.
Substitution of components may impair instrinsic safety.
Take care not to generate mechanical spark when access to the instrument and
peripheral devices in hazardous locations.
4. Maintenance and Repair
The instrument modication or parts replacement by other than authorized representative
of Yokogawa Electric Corporation is prohibited and will void the certication of CSA
International.