Owner's manual

<2. Installation, Wiring and Piping>
2-7
IM 11T3E1-01E
2.2.3 GD402V (CSA Explosion-proof Approval)
Sealing Fitting
Conduit
500 mm (20") Max.
NON-HAZARDOUS
LOCATIONS
HAZARDOUS
LOCATIONS
GD402V
F0204.ai
Figure 2.5
1. GD402V Gas density converter for use in hazardous (classied) locations:
Explosion-proof for Class I, Division 1, Groups B, C and D;
Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G
Enclosure Rating: Type 4X
Temperature Code: T6
Ambient Temperature: -10 to 55°C
2. Wiring
All wiring shall comply with Canadian Electrical Code and Local Electrical Codes.
Seal all conduits within 500 mm of the enclosure. Refer to Figure 2.5.
Note a warning label worded as follows.
WARNING
A seal shall be installed within 500 mm (20 inches) of the enclosure.
3. Operation
Note a warning label worded as follows.
WARNING
Open circuit before removing cover.
Take care not to generate mechanical spark when access to the instrument and
peripheral devices in hazardous locations.
4. Maintenance and Repair
The instrument modication or parts replacement by other than authorized representative
of Yokogawa Electric Corporation is prohibited and will void the certication of CSA Inter-
national.