Owner's manual

<5. DENSITY / PARAMETER SETTING>
5-6
IM 11T3E1-01E
Table 5.4 Setting Parameters at Service Level
Code
No.
Item Display
Data to be Set
(or Conditions)
Remarks
20 Selection of pressure unit *PRES.U · kPa : 0
· MPa : 1
· psi : 2
See Figure 5.17 for key operation.
21 Selection of density unit *DENS.U · kg/m
3
: 0
· lb/ft
3
: 1
See Figure 5.18 for key operation.
22 Selection of caloric value unit *CAL.U · MJ/m
3
: 0
· kBTU/ft
3
: 1
See Figure 5.19 for key operation.
23 Selection of temperature unit *TEMP.U · °C : 0
· °F : 1
See Figure 5.20 for key operation.
30 Setting of calendar
Setting of year/month/day
Setting of hour/minute
*Y_M_D
*H_M
00.01.01 to 99.12.31
00.00 to 23.59
See Figure 5.21 for key operation.
For leap years, the setting can
cope with ‘February 29’.
31 Showing/hiding of negative measured
values (-)
*MINUS · Show : 0
· Hide : 1
See Figure 5.22 for key operation.
40 Indication of calibration coefcients
(read-only)
Zero
Span
*C_K_Z
*C_K_S
SX.XXXX
X.XXXX
See Figure 5.23 for key operation.
S of SX.XXXX means Marks such
as a plus or the minus.
41 Indication of oscillation frequency
(read-only)
Frequency F2
Frequency F4
Frequency ratio
*F2.KHZ
*F4.KHZ
*F2/F4
X.XXXXX
X.XXXXX
X.XXXXX
See Figure 5.24 for key operation.
42 Indication of software version
(read-only)
*REV X.XX See Figure 5.25 for key operation.
43 Setting of high-resolution mode *S_CYC · Normal : 0
· High-resolution : 1
See Figure 5.26 for key operation.
44 Setting of password *PASS 0.0.0 to 9.9.9 See Figure 5.27 for key operation.
45 Selecting Battery alarm detection *BAT · Non detection : 0
· Detection : 1
See Figure 5.28 for key operation.
50 Selecting the instruments *MODEL · Density meter : 0
· Calorie meter : 1
·
Hydrogen purity, replacement
meter : 2
See Figure 5.29 for key operation.
82 Setting Detector Constant See Figure 5.30 for key operation.
5.2 Parameter Setting
[Aborting the Setting Operation]
Press the [MODE] key. Normally, the [MODE] key is used to move from the measurement
(<MEASURE>) mode to a mode at the operation level. It is also used to return to the measure-
ment mode from other modes.
5.2.1 Parameter Setting at Operation Level
To move to the operation level, press the [MODE] key and then enter the password (XXX). Pass-
word is not selected in rst condition. When password is necessary, refer to Section 4.2 “Setting
Lists” and selected password what you need.