Owner's manual
<5. DENSITY / PARAMETER SETTING>
5-14
IM 11T3E1-01E
2. Alarm-point Setting Mode [*ALARM]
Set the upper /lower limits for the measured values of the parameters noted below, for the pur-
pose of raising an alarm.
(a) Select a parameter for which alarms are raised.
・ Physical density [*DENS]
・ Compensated density [*C_DNS]
・ Specic gravity [*SP_GR]
・ Caloric value [*CALRY]
・ Molecular weight [*MOL]
・ Concentration [*CONCT]
Press the [YES] key to enter the desired mode when it is indicated.
(b) Set the upper/lower limits.
・ Physical density [*L_DNS], [*H_DNS]
Congure range: 00.0000 to 60.0000 (kg/m
3
),
0.00000 to 4.00000 (lb/ft
3
)
・ Compensated density [*L_CP.D], [*H_CP.D]
Congure range: 0.0000 to 6.0000 (kg/m
3
),
0.00000 to 0.40000 (lb/ft
3
)
・ Specic gravity [*L_SPC], [*H_SPC]
Congure range: 0.0000 to 5.0000
・ Caloric Value [*L_CAL], [*H_CAL]
Congure range: 000.000 to 133.000 (MJ/m
3
),
0.0000 to 3.5000 (kBTU/ft
3
)
・ Molecular weight [*L_MOL], [*H_MOL]
Congure range: 000.00 to 140.00
・ Concentration [*L_CON], [*H_CON]
Congure range: 000.0 to 100.0