4-7
4.1.3. UNATTENDED SAMPLE
Select 3. Unattended sample from the Run menu. The following menu will be displayed.
NOTE: If the current time and date are not correct, your unattended sampling study will not begin
or end when you desire. To correct the time and date, see Section 2.5.
You will be asked to enter the following information concerning the unattended sampling study you
wish to set up.
Be sure to enter hours, minutes and seconds for time entry. If you enter only hours and minutes, the
sonde will not accept the entry.
After entering the information, the program will re-display what you entered in the following
format.
========================================
Unattended-sample
----------------------------------------
EMS6000 version 3.00 (6/07/96).
Current date and time: 06/04/96 07:29:53
Sensors: Temp,Cond,DO,ISE1 mV,Batt,Turb
Hardware revision: UPG3
Battery type: Alkaline
Battery voltage (Volts) : 12.3
Free Flash disk space (KBytes): 100
Enter Starting date (xx/xx/xx):
Enter Starting time (xx:xx:xx):
Enter duration in days (xx.xxx):
Enter interval in minutes (x):
Enter site description ():