User Manual
Table Of Contents
- Contents
- Introduction
- General description
- Indications for use
- Important safety information
- Before first use
- Cleaning and disinfecting
- Using the breast pump
- Using the breast pump when you have symptoms of engorgement
- After use
- Feeding breast milk with the bottle
- Compatibility
- Ordering accessories
- Disposal
- Full Two-Year Warranty
- Troubleshooting
- Undesirable side effects
- Storage conditions
- Electromagnetic fields (EMF)
- Technical information
- Explanation of symbols
- Contenido
- Introducción
- Información general
- Indicaciones de uso
- Información de seguridad importante
- Antes del primer uso
- Limpieza y desinfección
- Uso del extractor de leche
- Uso del extractor de leche cuando tiene síntomas de congestión mamaria
- Después del uso
- Alimentación del bebé con la leche materna en el biberón
- Compatibilidad
- Pedido de accesorios
- Desecho
- Garantía Total de Dos Años
- Solución de problemas
- Efectos secundarios no deseados
- Condiciones de almacenamiento
- Campos electromagnéticos (CEM)
- Información técnica
- Explicación de los símbolos
- Table des matières
- Introduction
- Description générale
- Indications d’utilisation
- Renseignements importants sur la sécurité
- Avant la première utilisation
- Nettoyage et désinfection
- Utilisation du tire-lait
- Utilisation du tire-lait en cas de symptômes d'engorgement
- Après utilisation
- Donner du lait maternel au biberon
- Compatibilité
- Commande d'accessoires
- Mise au rebut
- Garantie complète de deux ans
- Dépannage
- Effets secondaires indésirables
- Conditions d’entreposage
- Champs électromagnétiques (CEM)
- Renseignements techniques
- Explication des symboles
32
English
Immunity test Immunity test level Electromagnetic environment -
guidance
Conducted disturbances
induced by RF fields
IEC 61000-4-6
3 Vrms, before modulation
is applied
0.15 MHz – 80 MHz
6 Vrms, before modulation
is applied, in ISM and
amateur radio bands
between 0.15 MHz and 80
MHz
80 % AM at 1 kHz
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
0 % UT; 0.5 cycle at 0°, 45°,
90°, 135°, 180°, 225°, 270°
and 315°
0 % UT; 1 cycle and 70 %
UT; 25/30 cycles single
phase at 0°
0 % UT; 250/300 cycle
Mains power quality expected of a typical
commercial or residential environment. At
power interruptions the breast pump will
automatically switch-over to battery mode
without interruption of the function.
Note: UT is the AC mains voltage prior to application of the test level. If in the unlikely event the Philips
Avent single/double electric breast pump Advanced is locked inoperative in a function by disturbances from
other devices beyond the levels stated in the table above, the device will shut-off completely and restart. In
that case make sure the distance to the other device causing this disturbance is enlarged.
FCC compliance information
This device has been tested and found to comply with the limits for a ClassB digital device, pursuant to
Part15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful
interference in a residential installation. This device generates, uses and can radiate radio frequency energy
and, if not installed and used in accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not occur in a particular installation.
If this device does cause harmful interference to radio or television reception, which can be determined by
turning the device off and on, the user is encouraged to try to correct the interference by one or more of the
following measures:
-
Reorient or relocate the receiving antenna.
-
Increase the separation between the device and the receiver.
-
Connect the device into an outlet on a circuit different from that to which the receiver is connected.
-
Consult the dealer or an experienced radio/TV technician for help.
Explanation of symbols
The warning signs and symbols are essential to ensure that you use this device safely and correctly and to
protect you and others from injury. Below you find the meaning of the warning signs and symbols on the
label and in the user manual.
Symbol Explanation
Indicates to follow the instructions for use.