NucleoCounter YC-100 Yeast Cell Counter P/N 991-0300 (English) Revision 1.2 December, 2005 ChemoMetec A/S Gydevang 43, DK-3450 Allerød, Denmark Telephone: (+45) 48 13 10 20, Fax: (+45) 48 13 10 21 Internet: www.chemometec.com, E-mail: info@chemometec.
This page was intentionally left blank iii
Caution! This equipment must be operated as described in this User’ s Guide. Please read the entire guide before attempting to use this unit. Please pay attention to that gloves or protective clothing are not worn on the illustrations/pictures shown in this User’s Guide. However, ChemoMetec A/S does recommend that the user wear suitable protective clothing etc.
This page was intentionally left blank v
Declaration of Conformity NucleoCounter Name of product: Type: YC-100 Other identifying data: Part no. 900-0300 The product complies with requirements of the following directives: 89/336/EEC - Electromagnetic Compatibility (EMC) 89/392/EEC - Machinery Directive Harmonized standards, which have been used: EN61326: 1997+A1: 1998 (Class B) +A2: 2000, Electrical equipment for measurement, control and laboratory use – EMC requirements (emission and immunity). Annex B and C from A1: 1998 is used.
This page was intentionally left blank vii
WEEE directive information – Europe only Correct Disposal of This Product (Waste Electrical & Electronic Equipment) - Europe only This marking shown on the product or its literature, indicates that it should not be disposed together with other household wastes at the end of its working life.
This page was intentionally left blank ix
Introduction and intended use Introduction and intended use The NucleoCounter YC-100 is intended to be used for the determination of yeast cell concentration in sample containing dispersed yeast cells in suspension. The NucleoCounter YC-100 has been developed to count dispersed yeast cells in suspension The instrument is a part of the NucleoCounter YC-100 system, which also comprises a disposable NucleoCassette and a lysis buffer (Reagent Y100).
Introduction and intended use This page was intentionally left blank xi
Warnings and precautions Wherever the symbol appears on the NucleoCounter instrument, it indicates that the manual must be consulted for precautions and warnings. Power and cables Use the shielded USB cable supplied with the NucleoCounter to ensure that you maintain the appropriate EMI classification for the intended environment. The USB interface connector of the NucleoCounter must only be connected to SELV circuits.
Table 2 Description of the NucleoCounter Printer Output connector Name Maximum Voltage level 2 2 Rx ±10 VDC 3 Tx ±10 VDC Metal Enclosure DGND 0 VDC Pin no 13 2 3 1 Metal Enclosure Figure 1 Printer Output connector seen from the cable entry side. NucleoCounter is powered by an external 12VDC power supply. For safe use, please follow the instructions for connecting the power supply.
The user shall be made aware of that, if the NucleoCounter and the external power supply is used in a manner not specified by the manufacturer, the protection provided by the NucleoCounter and the external power supply may be impaired. Electromagnetic interference NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules.
Never pipette by mouth. Avoid cross contamination of the samples when preparing of the samples. This can compromise the quality of the results.
Table of contents DECLARATION OF CONFORMITY ................................................................................ VI WEEE DIRECTIVE INFORMATION – EUROPE ONLY ...................................................... VIII INTRODUCTION AND INTENDED USE .......................................................................... X WARNINGS AND PRECAUTIONS ................................................................................. 1 1 INSPECTION AND UNPACKING OF EQUIPMENT ..........................
5.4.1 5.4.2 5.4.3 5.4.4 5.4.5 5.4.6 Result mode .............................................................................................. 21 Saving parameters ..................................................................................... 22 Setting Date (F200) and Time (F201) ......................................................... 22 Reset Counter (F30) ................................................................................... 22 LCD contrast (F220) ..................................
1 Inspection and Unpacking of Equipment 1 Inspection and Unpacking of Equipment Upon receiving the order from ChemoMetec A/S, the box or boxes should be carefully inspected for any damage that may have occurred during shipping. Any damage must be reported to the carrier and to ChemoMetec A/S immediately. Unpack the order, saving the packing materials for possible later use. Also be sure to save the User’s Guide, for instruction and reference.
1 Inspection and Unpacking of Equipment This page was intentionally left blank 8
2 Introducing the NucleoCounter 2 Introducing the NucleoCounter YC-100 The NucleoCounter method provides a fast alternative to the manual techniques, yielding objective and reliable results. Furthermore the NucleoCounter technology provides improved safety to the user compared to other comparable methods.
2 Introducing the NucleoCounter 2.1.1 Fluorescence microscope The compact fluorescence microscope integrated in the NucleoCounter comprises LED´s (light emitting diodes) as excitation light source, excitation and emission filters, optics (lenses) and a CCD (charged coupled device) camera. Furthermore, the NucleoCounter contains advanced software for image processing. The fluorescence microscope, with a cassette inserted, is shown in Figure 3. Figure 3 .
2 Introducing the NucleoCounter A fluorescent image of the content of the measurement chamber is thus . Example of such an image is shown in Figure 4. Each spot in the image represents a cell containing DNA stained with PI. Figure 4 . Fluorescent image (from NucleoView PC software) of cell DNA contained in the measurement chamber of a NucleoCassette 2.1.2 Built-in image analysis The NucleoCounter is equipped with advanced software for image analysis.
2 Introducing the NucleoCounter Figure 5 The NucleoCassette. The fluorescent dye, propidium iodide, is immobilized in the first part of the flow system. 2.2.1 Loading the cassette The NucleoCassette is loaded by gently pressing the white piston, which creates a partial vacuum in the flow system. The tip of the NucleoCassette must be immersed into a lyzate mixture when pressing the piston, resulting in lyzate being loaded into the flow system. Approximately 50µl is loaded into the flow system. 2.2.
2 Introducing the NucleoCounter by multiplying the imaged area with the depth of the cassette. The imaged area is only dependent on the optics of the instrument. Therefore, the area is constant and specific for each instrument. 2.2.
This page was intentionally left blank 14
3 Control Buttons, Keypad and Display 3 Control Buttons, Keypad and Display 3.1 3.1.1 Interactive Controls Control Buttons There are four Control Buttons on the front of the instrument (see Figure 6 ): Table 3..
3 Control Buttons, Keypad and Display 3.2 3.2.1 Interface Display Status indicator A small Status Indicator light above the “On/Off” button (see Figure 6) illuminates when the unit is turned on. It is green when the instrument is ready to measure and it turns red while the instrument is measuring or displaying an error message. The Status Indicator turns green after successful measurement when the results are displayed. 3.2.2 Display screen The display screen communicates information to the user.
4 Installation and Start-up 4 Installation and Start-up 4.1 Power on, power off The NucleoCounter is powered by a 12VDC external power supply. To connect the power supply to the NucleoCounter, connect it to the DC socket (see Figure 7), and plug the mains power cord of the power supply into a wall outlet. Figure 7 . Connecting the 12VDC power cord plug to the NucleoCounter The NucleoCounter can now be turned on and off by pressing the “On/Off” button on the front panel. 4.1.
4 Installation and Start-up After a successful start-up the display changes (Figure 9) and the status indicator on the keyboard panel turns green. NucleoCounter Ready Figure 9 . The NucleoCounter is ready for cell counting If the instrument has performed as described above, it is ready for use. If the instrument displays an error message at start-up please refer to chapter 7. 4.1.
5 Operation of the NucleoCounter 5 Operation of the NucleoCounter YC-100 5.1 Ready mode After the NucleoCounter has started successfully the display should read: NucleoCounter Ready Figure 11. 11 . The NucleoCounter is ready for cell counting. This display is also shown if the “Esc” button is activated in the result mode or any other situation, except during measuring or shutting down. 5.
5 Operation of the NucleoCounter After analysis the lid is lifted and the cassette is removed from the insertion slit and disposed of according to national or regional laws or regulations regarding the nature of the mixture it contains. Figure 12. 12 . Illustration of how to grip and insert the cassette into the insertion slit 5.3 Measuring - The Run button When a cassette has been loaded and inserted into the instrument, the measurement is activated, by pressing the Run button. 5.3.
5 Operation of the NucleoCounter 5.3.3 External computer If an external computer with the NucleoView software installed on it is connected, the results and the recorded fluorescent image can be viewed using the NucleoView software. After the measurement is completed the cassette can be safely removed (see section 5.2). Sometimes, however, the complete process described above may be interrupted. Please familiarize yourself with possible error handling of the NucleoCounter presented in chapter 8 . 5.
5 Operation of the NucleoCounter Press <↵>. • 5.4.2 Saving parameters The representation of the results is changed with effect from the next analysis. When the NucleoCounter is turned off, after changing the status, the display will read “Save parameters?”, "ENTER: Save" and "ESC : Cancel" (figure 14). Save parameters? ↵ ESC : Save : Cancel Figure 14 The shut-down display when any parameters have been changed.
5 Operation of the NucleoCounter Contrast? 1: Up 2: Down ↵ store 6 Figure 16. 16 . LCD-contrast value 6 is going to be chosen Press “↵ ↵ ” to store the chosen value (or “Esc” Esc” to return to the previous value). If the value has been changed the new contrast will be remembered during the session. When shutting down the NucleoCounter asks if the new value shall be saved as the new default value. Press “↵ ↵ ” to save or “Esc Esc” Esc if the present default value should not be overwritten. 5.4.
5 Operation of the NucleoCounter Setup? haploid 1:Up 2:Down ↵ store Figure 18 Zero Count Check, step 1 To select the appropriate settings use the keys "1" and "2" to scroll in the list of options. The options are: Haploid, Diploid and Small. The labels refer to "typical" settings for the measurement of Saccharomyces cervesae . The label of the settings are only to be considered as guidelines since other factors might influence the intensity of the observed signals.
5 Operation of the NucleoCounter elevated signal levels. If the number of pixels is above a predefined limit then the cassette insertion slit can be contaminated with an interfering particle, like a dust particle. An error message is shown in the display and a cleaning of the insertion slit is recommended (refer to chapter 7 Maintenance of NucleoCounter). If a PC with NucleoView is connected to the instrument a contaminant is usually observed as an object, which is stationary in all images.
5 Operation of the NucleoCounter After a few seconds the instrument has performed the Zero Count Check. If no contamination of the insertion slit can be detected an “OK” message appears in the display (see the figure below). Press any key to return to “Ready mode”.
6 Total and viability count 6 Total and viability count The NucleoCounter detects fluorescent signals from propidium iodide (PI) when it is bound to double-stranded DNA (dsDNA). The fluorescent intensity of PI is increased substantially upon binding to dsDNA. As a result the NucleoCounter detection is highly specific and sensitive. The staining of the dsDNA requires that PI can penetrate the cell membrane.
6 Total and viability count 6.3 Determination of Viability When both the concentration of non-viable cells and the total concentration of cells are known, it is possible to calculate the viability (%viability) of the cells or the concentration of viable cells. For the calculation the multiplication factors must be taken into account, which is the total dilution of the original sample, as in the equation shown below.
7 Maintenance of NucleoCounter 7 Maintenance of NucleoCounter YC-100 A regular cleaning of the NucleoCounter is recommended in order to protect its surface and assure the quality of the collected images. 7.1 Cleaning Depending on the environment in which the NucleoCounter is operated it is suggested that regular cleaning of the cabinet is carried out. When cleaning the cabinet it is recommended to use a soft moist cloth and gently wipe the surface.
7 Maintenance of NucleoCounter Figure 26. 26 . 7.1.3 An image showing a contamination of the optical system, visible as a white cloudy phenomena in the upper right hand corner. If the contamination is stationary it should be removed using compressed air. Removal of dust particles The presence of a contaminant (e.g. dust particles) will normally not influence the counting of cells.
7 Maintenance of NucleoCounter 7.1.4 Spill of liquid If liquid has been spilled on the instrument it can contaminate elements of the optical system. This contamination can show up in the image in several ways and have many causes but a common feature is that it cannot be removed by the use of compressed air. Even though such phenomena are clearly visible on the image it only rarely affects the results of the NucleoCounter.
7 Maintenance of NucleoCounter This page was intentionally left blank 32
8 Troubleshooting - Error messages 8 Troubleshooting - Error messages Under certain conditions the NucleoCounter will display error messages during operation. Corrective actions are suggested below but if they do not correct the errors, contact ChemoMetec A/S or the local distributor. 8.1 No valid cassette When trying to analyze a cassette, which is not valid, the NucleoCounter displays the error message shown in Figure 28. No valid cassette Press any key Figure 28. 28 .
8 Troubleshooting - Error messages 8.3 Actuator error messages The actuator moves the piston down into the barrel of the cassette. Two different actuator error messages can be displayed. These error messages are shown in Figure 30. The same action is to be taken if either of the error messages appears. Actuator error Refer to manual Press any key Figure 30. 30 . The NucleoCounter can display this message regarding the actuator First, examine if the cassette has been inserted properly.
8 Troubleshooting - Error messages Sensor error Refer to manual Press any key Figure 32. 32 . This error message is displayed if one of the sensors, which read the black dots on the cassettes, causes error. If the error message is displayed analyses can still be made but the accuracy of the results cannot be guaranteed since the volume of the measurement chamber may not be read correctly. 8.
8 Troubleshooting - Error messages This page was intentionally left blank 36
9 Technical specifications 9 Technical specifications 9.1 Technical specifications for NucleoCounter YC-100 Specificity Cells stained with the DNA specific fluorescent dye, PI. Sample types Suspension of dispersed yeast cells. Sample consumption A minimum of 100µl is recommended for the NucleoCassette loading. Approximately 50 µl of sample is loaded into the cassette. Analysis volume Approximately 1 µl of the sample is being analyzed in the NucleoCounter YC-100.
9 Technical specifications Other environmental operating conditions for the NucleoCounter Indoor use only. Altitude Up to 2000 m. INSTALLATION CATEGORY II (Refer to UL 61010A-1) OVERVOLTAGE CATEGORY II (Refer to UL 61010A-1) POLLUTION DEGREE II (Refer to UL 61010A-1) 9.2 The NucleoCassette™ Reagent Each NucleoCassette contains approximately 2,8 µg propidium iodide. Storage Store the NucleoCassettes in a sealed foil bag at max. 30°C.
Table Tab le 4 . Equipment and Accessories for NucleoCounter system Item NucleoCounter YC-100 Part no. 900-0300 Description Instrument 4 NucleoCassette, 100 pcs. 941-0002 A disposable device used for performing the cellcount Reagent Y100, 500 ml 910-0300 5 A reagent used for lysis of cells prior to measurement CD with NucleoView v. 2.
Item Part no. Description using ink ribbon, for the NucleoCounter Thermo Printer PCNEOS-S2BN 939-0006 Optional thermo printer, using thermo paper, for the NucleoCounter Paper Impact Printer, 5 RL 939-0004 5 roles of paper for impact type printer Ribbon Impact Print, 1pcs 939-0005 One printer ribbon for impact type printer Paper Thermo Printer, 5 RL 939-0007 5 roles of thermo paper Applic. Note 300-GB 994-0300 Applic.
UL listed, type SVT, rated min. 60C, 18 AWG, 3 conductors. Provided with molded-on grounding-type (NEMA 5-15P) attachment plug, rated min. 125 Vac, min. 2.5A. Opposite end terminates in molded on, IEC320 style connector, rated min. 125 Vac, min. 2.5 A. For Europe 250Vac Cord type min. H05RR-F or min. H03VV-F or min. H03VVH2-F, rated min. 60C, 3x0.75 mm2. Provided with molded-on grounding-type attachment Plug, rated min. 250 Vac, min. 2.5 A.
10.5 External Printer (optional) An external impact or thermo printer may optionally be connected to the NucleoCounter for the printing of test results. For connection of a printer (Impact type CM no. 939-0003; thermo type CM type no. 939-0006) please refer to Technical Note 001 (Impact type CM no. 994-0013; thermo type CM no. 994-0014). The NucleoCounter is equipped with a printer output port.
2 3 1 Metal Enclosure Figure 36 Printer Output connector seen from the cable entry side.
This page was intentionally left blank 44
11 Appendix 1: WEEE directive information in more EU languages United Kingdom Correct Disposal of T his Product (Waste Electrical & Electro nic Eq uip ment) - Europe onl y This marking sho wn on the product or its literature, indicates that it should not be disposed together with other household wastes at the end of its working life.
Norwa y Korrekt a vhendi ng a v dette produkt (Avfall elektrisk og elektronisk utstyr) - Kun Europa Denne merkingen so m vises på produk tet eller dens doku mentas jon, indikerer at den ikke skal kastes sa mmen med annet husholdningsavfall ved slutten av sin levetid. For å hindre mulig skade på miljøet eller menneske lig helse fra ukontrollert avfallsavhending, vennligst atskill dette fra andre typer avfall og resirkuler det ansvarlig for å fremme b ærekraftig gje nbruk av materielle ressurser.
Finland T ämän tuotteen turvalli nen hä vittämi nen (elektroniikka ja sä hkölaitteet) - Vain Eurooppa Oheinen merkintä tuotteessa tai tuotteen oheismateriaalissa merkitsee, että t ätä tuotetta ei tule hävittää kotitalousjätteen mukana sen elinkaaren päätyttyä. Hallitsematto mas ta jätteen käsittelystä y mpäristölle ja kanssaihmisten terveydelle aiheutuvien vahinkojen välttä miseksi tuote tulee käsitellä muista jätteistä erillään. Jäte on hyvä kierrättää raaka-aineiksi kestävän ympäristökehityksen ta kia.
The information contained herein is to the best of our knowledge accurate and complete. However cell species and cell environments may vary in property. Therefore systematic and/or NucleoCounter random deviation between estimates obtained by the and other cell counting methods may occur. As such, nothing contained or stated herein including results obtained from use of the NucleoCounter or NucleoCassette™ shall be construed to imply any warranty or guarantee.