SHEL LAB SI6/SI6R Shaking Incubator Operational Manual Table of Contents Using the Unit Safely 4 Introduction ..................................................................................................................... 4 General Safety Considerations ........................................................................................ 4 Precautions for Your Unit ............................................................................................... 4 Meanings of Symbols ..............
SHEL LAB SI6/SI6R Shaking Incubator Operational Manual Table of Contents Getting Assistance ........................................................................................................ 20 Obtaining Nameplate Information ................................................................................ 20 Returning Your Unit ..................................................................................................... 20 SHEL LAB Contact Information .........................................
SHEL LAB SI6/SI6R Shaking Incubator Operational Using the Unit Safely Introduction Thank you for choosing a SHEL LAB shaking orbital incubator. SHEL LAB sets the standard for quality and reliability. Your unit is backed by over 30 years of design and manufacturing excellence in the scientific, research, and medical equipment industries.
SHEL LAB SI6/SI6R Shaking Incubator Operational Meanings of Symbols About this Manual In this manual and on labels attached to the product, graphic symbols have the following meanings: Throughout this manual, the words WARNING and CAUTION have the following meanings: Symbol Indcation You should consult this manual for a description or discussion of a control or item Temperature WARNING A potentially hazardous situation that, if not avoided, could result in serious injury or death.
SHEL LAB SI6/SI6R Shaking Incubator Operational Manual Features of Your Unit Product description Key Features Your shaking orbital incubator provides: Controlled environment For continuous growth of biological organisms. Vibration-free operation A unique adjustable counterbalance system provides vibration-free operation regardless of load. Large chamber A large six cubic foot chamber facilitates throughput. Shelves in the top provide space for static incubation during shaking sessions.
SHEL LAB SI6/SI6R Shaking Incubator Operational Manual Receiving Your Unit Unpacking Your Unit and 6 counter weights (some of which are located in the metal pocket at the back of the unit) 4 leveling feet Inspecting Before leaving our factory, all units are packaged in high quality shipping materials designed to provide protection from transportation related damage. If the carton is not damaged, open the crate and remove its contents.
SHEL LAB SI6/SI6R Shaking Incubator Operational Manual Installing the Unit that the door can swing open at least 90 degrees. Do not block access to the Installation Overview power cord, circuit breaker or fuses. To install your unit, you need to: 1. Select a suitable operating location for the unit. 2. Level the unit. 3. Sterilize the unit. 4. Install the sample tray. 5. Plug the unit into a power source.
SHEL LAB SI6/SI6R Shaking Incubator Operational Manual sterilizing your unit, see “Disinfecting Your Unit“ on page 14. Installing Sample Tray Your unit comes with a sample tray as standard equipment. To install the sample tray 1. Enclose all corners of the shaking mechanism within the lips of the sample tray. This can be done easily by positioning the front two corners and then setting the rest of the tray down. 2. Shake the tray by its handles to confirm that it is firmly in place.
SHEL LAB SI6/SI6R Shaking Incubator Operational Manual Operating the Unit Getting the Unit Ready for Use Control Panel Overview Before turning the incubator on for the first time, take a moment to familiarize yourself with its controls and features. Following is an overview of the control panel. WARNING This equipment is NOT intended for the processing of Flammable materials. 1. Main temperature control 2. Turn the shaking mechanism on and adjust the speed of the shaking mechanism.
SHEL LAB SI6/SI6R Shaking Incubator Operational Manual 3. 4. When you turn the unit on for the first time, use a screwdriver or coin to turn the Safety Temp knob fully clockwise to its maximum position. This deactivates the OverTemperature Protection (OTP) feature. For more information on the OTP, see on page 12. If they match, you do not need to calibrate your unit for that temperature. If they do not match, you need to calibrate your unit. To calibrate your unit 1.
SHEL LAB SI6/SI6R Shaking Incubator Operational Manual To set the safety temperature thermostat Using the Timer 1. Using the incubator’s timer, you can run the shaker platform at a preset RPM for a preset time. The timer can be set at intervals of one (1) minute up to a maximum of a 999 (16 hours 39 minutes). See “Calibrating the Main Temperature Control” on page 11. Upon completion of the timing cycle, the TIMER alarm LIGHT will turn ON and an alarm will sound. 2.
SHEL LAB SI6/SI6R Shaking Incubator Operational Manual 4. Rotate the counterweight platform until the counterweights are in full view. Remove the wingnuts and adjust the counterweights according to the diagrams shown below. 5. Replace the counterweight wingnuts. Adjusting the Shaker Movement You can adjust the shaker movement to gentle, moderate, or vigorous shaking. Which shaking movement you use depends on the oxygenation needs and cell strength of your samples.
SHEL LAB SI6/SI6R Shaking Incubator Operational Manual Interior Accessory Outlet This unit features an accessory outlet to provide power for equipment such as magnetic stirrers, rockers, etc. The weight of this equipment should not exceed 22 pounds (10 kg) per shelf. This equipment may provide additional heat that could affect the temperature range of this incubator. We recommend testing the incubator and any accessory equipment to ensure that the desired operating conditions can be met.
SHEL LAB SI6/SI6R Shaking Incubator Operational Manual Maintaining the Unit The only regular maintenance required for your unit is to keep it clean and free from contamination. Use the guidelines and instructions in this section to maximize the life of your incubator and help prevent contamination of your samples. Disinfect the parts with a 70% alcohol solution. Rinse with distilled water and wipe dry with a soft cloth.
SHEL LAB SI6/SI6R Shaking Incubator Operational Manual Troubleshooting Solving Problems WARNING Should the proper function of your unit come into question, use this section to help you determine what the problem is and how to fix it. Replacing fuses with wrong type and value can result in serious damage to the equipment and property. ONLY Replace fuses with the same type and amperage indicated on the replacement fuse labels. Check if your question is similar to those listed below.
SHEL LAB SI6/SI6R Shaking Incubator Operational Manual What is the problem? Possible Causes Over Temperature Protection (OTP) is set too low. - Or - To solve the problem... 3. Heating element failure. 4. If this doesn’t solve the problem... Be sure your reference thermometer is certified. If this is not the problem... Turn the OTP fully clockwise. If this doesn’t solve the problem... Call customer service. See page 21 The Main Temperature Control displays “LO”.
SHEL LAB SI6/SI6R Shaking Incubator Operational Manual What is the problem? Possible Causes To solve the problem... Refrigeration (SSI5R models only) (SI6R) The unit will not cool. The evaperator has too much ice built up on it. The unit is not calibrated correctly. There is not enough space between the unit and adjacent walls or partitions. 1. - Or - - Or - - Or - The door seal does not work properly. If this does not solve the problem... 2. The door gasket leaks.
SHEL LAB SI6/SI6R Shaking Incubator Operational Manual What is the problem? Possible Causes To solve the problem... The wall fuse/circuit break blows often. Too many things may be plugged in. 1. Check to see what other loads are on the same circuit as the unit. We recommend that you plug your incubator into a circuit separate from other equipment. The front panel displays fail to turn on but the rest of the unit receives power. Controller failure 1. Call customer service.
SHEL LAB SI6/SI6R Shaking Incubator Operational Manual Getting Your Unit Serviced Getting Assistance Returning Your Unit Your incubator will provide years of trouble-free operation. Should you require assistance, however, SHEL LAB’s customer service and support is available to assist you. If you need to return your unit for any reason, first contact your customer representative for return authorization number (RA#). Be sure to print the RA# clearly on the package in which you ship your unit.
SHEL LAB SI6/SI6R Shaking Incubator Operational Manua Replacement Parts and Accessories Replacement Parts Part 115V 220V Adjustable feet 2700500 2700500 Alarm Display Vertical 1750681 1750681 Counterweight, Single 5460662 Door Gasket 115V 220V Motor, Circulation 4880527 4880528 Outlet, Interior 4200518 6100531 5460662 Platform (Sample Tray) 9750758 9750758 3450562 3450562 Power Cord 1800510 1800500 Drive Belt, Oscillator 0500512 0500512 7010543 9570703 9570738 Refrigerati
SHEL LAB SI6/SI6R Shaking Incubator Operational Manual Specifications SSI5 (SI6) SSI5R (SI6R) Unit Range 10C above Ambient to 60 C 10C above Ambient to 60C Uniformity +0.8 C at 37 C +0.8 C at 37 C Temperature Accuracy Alarms +0.1 C +0.1 C Visual Safety Lamps Visual Safety Lamps Capacity 156 m3 (5.5 cu. ft.) 156 m3 (5.5 cu. ft.) Shelves Supplied 2 stainless steel 2 stainless steel Shelf Dimensions 47 × 47 cm (18.5 × 18.5 in.) 47 × 47 cm (18.5 × 18.5 in.
SHEL LAB SI6/SI6R Shaking Incubator Operational Manual WIRE DIAGRAMS SSI5 (SI6) 110V (9851457) 2800502 EMI FILTER BLUE INLET 4200505 BROWN BROWN BLUE POWER SWITCH 7850533 BROWN BLACK/VIOLET BLACK/WHITE BLACK BLACK/YELLOW 120V / 240V TO 90V STEP DOWN TRANFORMER 8350509 TIMER SWITCH 7850553 YELLOW YELLOW J11 MOTOR RED BROWN ORANGE MOTOR N HEATER MOTOR HALLS RPM SWITCH 7850553 J16 FERRITE BEAD 7830500 RED BODINE MOTOR CABLE 0860503 BROWN ORANGE TIMER DISPLAY 1750679 DOOR SWITCH 78505
SHEL LAB SI6/SI6R Shaking Incubator Operational Manual SSI5-2 (SI6-2) 220V (9851458) 2800502 EMI FILTER BLUE BLUE INLET 4200505 BROWN BROWN BLUE BLACK 120V / 240V TO 90V STEP DOWN TRANFORMER 8350509 BLACK/VIOLET BLACK/WHITE BROWN POWER SWITCH 7850533 BLACK/YELLOW TIMER SWITCH 7850553 YELLOW YELLOW J11 MOTOR RED BROWN ORANGE MOTOR N HEATER MOTOR HALLS RPM SWITCH 7850553 J16 FERRITE BEAD 7830500 RED TEMPERATURE SENSOR BODINE MOTOR CABLE 0860503 BROWN ORANGE TIMER DISPLAY 1750679 D
SHEL LAB SI6/SI6R Shaking Incubator Operational Manual SSI5R (SI6R) 110V (9851459) 2800502 EMI FILTER BLUE INLET 4200505 BROWN BROWN BLUE POWER SWITCH 7850533 BROWN BLACK/VIOLET BLACK/WHITE BLACK BLACK/YELLOW 120V / 240V TO 90V STEP DOWN TRANFORMER 8350509 TIMER SWITCH 7850553 YELLOW YELLOW J11 MOTOR RED BROWN ORANGE MOTOR N ORANGE TEMP DISPLAY 1750791 J12 MAIN CONTROL BOARD 1750786 RPM DISPLAY 1750788 DOOR SWITCH 7850578 LOAD HOT VALVE HEATER J10 MOTOR HALLS RPM SWITCH 7850553 J
SHEL LAB SI6/SI6R Shaking Incubator Operational Manual SSI5R-2 (SI6R-2) 220V (9851460) 2800502 EMI FILTER BLUE BLUE INLET 4200505 BROWN BROWN BLUE BLACK 120V / 240V TO 90V STEP DOWN TRANFORMER 8350509 BLACK/VIOLET BLACK/WHITE BROWN POWER SWITCH 7850533 BLACK/YELLOW TIMER SWITCH 7850553 YELLOW YELLOW J11 MOTOR RED BROWN ORANGE MOTOR N ORANGE TEMP DISPLAY 1750791 J12 MAIN CONTROL BOARD 1750786 RPM DISPLAY 1750788 DOOR SWITCH 7850578 LOAD HOT VALVE HEATER J10 MOTOR HALLS RPM SWITCH