Owner manual

SHEL LAB SI6/SI6R Shaking Incubator Operational Manual
14
Interior Accessory Outlet
This unit features an accessory outlet to provide power
for equipment such as magnetic stirrers, rockers, etc.
The weight of this equipment should not exceed 22
pounds (10 kg) per shelf. This equipment may provide
additional heat that could affect the temperature range
of this incubator. We recommend testing the incubator
and any accessory equipment to ensure that the
desired operating conditions can be met.
Caution: This incubator operates at conditions that
might damage accessory equipment.
Verify that your accessory equipment is
capable of operating under the same
conditions as the incubator.
The outlet is located inside the incubator in the upper
right rear of the chamber. The voltage available at the
accessory outlet is the SAME as the voltage supplied to
the incubator. For example, a 120-vac incubator will
have 120 vac at the accessory outlet, and a 240-vac
incubator will have 240 vac at the accessory outlet. DO
NOT exceed 500 va rating of the accessory outlet.