Year 2000 Conversion Instructions and Test Procedure
OLYMPUS PK7200 Automated Microplate System
Year 2000 Conversion Instructions:
14. Power on the instrument using the Power On switch on the front of the DPR.
***OPERATION MENU***
# SELECT No.>
_____________________________________________________________
1 START CONDITION 8 DATA…… 21……
2 TEST REQUISITION 9 QC 22……
3 TEST DATA DISPLAY
4 ANALYZER STATUS
5.SAMPLE LIST
6 DATA MANAGEMENT
7 REPORT
1: 2: 3:WARM 4: 5:DIAG 6: 7:END 8:
15. From the Operation Menu select F5: DIAG then press the Yes key at the DIAG ? Yes or No
prompt.
***OPERATION MENU***
# SELECT No.>
_____________________________________________________________
1 DIAGNOSTIC (DPR)
2 DIAGNOSTIC (DPR)
3 DIAGNOSTIC (DPR)
4 DIAGNOSTIC (DPR)
5 MAINTENANCE (DPR)
6 MAINTENANCE (ANL)
7 MANUAL MEASURE
1: 2: 3:WARM 4: 5: 6: 7:END 8:
16. Select 5, Maintenance (DPR) and press enter .
*** 5 MAINTENANCE (DPR) ***
# SELECT No.>