Integra Series Specifications
SWITCH/MEASURE SYSTEMS
22
Integra Series Multimeter/Switch Systems
Mainframes
AC SPECIFICATIONS
1
Accuracy: ±(% of reading + % of range), 23°C ± 5°C
Function Range Resolution Calibration Cycle 3 Hz–10 Hz 10 Hz–20 kHz 20 kHz–50 kHz 50 kHz–100 kHz 100 kHz–300 kHz
Voltage
2
100.0000 mV 0.1 µV 90 Days 0.35 + 0.03 0.05 + 0.03 0.11 + 0.05 0.6 + 0.08 4.0 + 0.5
1.000000 V 1.0 µV (all ranges)
10.00000 V 10 µV 1 Year 0.35 + 0.03 0.06 + 0.03 0.12 + 0.05 0.6 + 0.08 4.0 + 0.5
100.0000 V 100 µV (all ranges)
750.000 V 1.0 µV
(Temp. Coeff.)/°C
3
0.035 + .003 0.005 + .003 0.006 + .005 0.01 + .006 0.03 + .01
3 Hz–10 Hz 10 Hz–3 kHz 3 kHz–5 kHz
Current
2
1.000000 A 1.0 µA 90 Day/1 Year 0.30 + 0.04 0.10 + 0.04 0.14 + 0.04
3.00000 A
14
10 µA 0.35 + 0.06 0.16 + 0.06 0.18 + 0.06
(Temp. Coeff.)/°C
3
0.035 + 0.006 0.015 + 0.006
(3 Hz–500 kHz) (333 ms–2 µs)
Frequency
4
100 mV 0.333 ppm 90 Day/1 Year 100 ppm + 0.333 ppm (SLOW, 1s gate)
and Period to 3.33 ppm 100 ppm + 3.33 ppm (MED, 100ms gate)
750 V 33.3 ppm 100 ppm + 33.3 ppm (FAST, 10ms gate)
ADDITIONAL UNCERTAINTY ±(% OF READING)
Low Frequency Uncertainty MED FAST
20 Hz – 30 Hz 0.3
30 Hz – 50 Hz 0
50 Hz – 100 Hz 0 1.0
100 Hz – 200 Hz 0 0.18
200 Hz – 300 Hz 0 0.10
>300 Hz 0 0
CREST FACTOR:
5
1–2 2–3 3–4 4–5
Additional Uncertainty: 0.05 0.15 0.30 0.40
Max. Fundamental Freq.: 50kHz 50kHz 3kHz 1kHz
Maximum Crest Factor: 5 at full-scale.
AC MEASUREMENT CHARACTERISTICS
AC VOLTS
MEASUREMENT METHOD: AC-coupled, True RMS.
INPUT IMPEDANCE: 1M ±2% // by <100pF.
INPUT PROTECTION: 1000Vp or 400VDC. 300Vrms with plug in modules.
AC CURRENT
MEASUREMENT METHOD: AC-coupled, True RMS.
SHUNT RESISTANCE: 0.1.
BURDEN VOLTAGE: 1A <0.5Vrms, 3A <1.5Vrms. Add 1.5Vrms when used with plug in modules.
INPUT PROTECTION: 3A, 250V fuse.
FREQUENCY AND PERIOD
MEASUREMENT METHOD: Reciprocal counting technique.
GATE TIME: SLOW 1s, MED 100ms, and FAST 10ms.
AC GENERAL
AC CMRR
6
: 70dB.
VOLT HERTZ PRODUCT: <= 8 × 10
7
.
AC MEASUREMENT SPEEDS
7, 13
Single Channel, 60Hz (50Hz) Operation
Function Digits Readings/s Rate Bandwidth
ACV, ACI 6.5 2s/Reading SLOW 3 Hz–300 kHz
6.5 4.8 (4) MED 30 Hz–300 kHz
6.5
9
40 (32) FAST 300 Hz–300 kHz
Frequency, 6.5 1 (1) SLOW 3 Hz–300 kHz
Period 5.5 9 (9) MED 30 Hz–300 kHz
4.5 35 (35) FAST 300 Hz–300 kHz
4.5
10
65 (65) FAST 300 Hz–300 kHz
Multiple Channel
7710 SCANNING ACV
10, 11
: 500/s.
7710 SCANNING ACV WITH AUTO DELAY ON : 2s/reading.
AC SYSTEM SPEEDS
7, 9, 11
2700/2750 2701
AC System Speed: (19.2k) (115.2K)
Range Changes:
12
4/s (3/s) 4/s (3/s)
Function Changes:
12
4/s (3/s) 4/s (3/s)
Autorange Time: < 3s < 3s
ASCII Readings to RS-232 (19.2k baud): 50/s 300/s
Max. External Trigger Rate: 250/s 2000/s
AC NOTES
1. 20 % overrange except on 750V and 3A.
2. Specification are for SLOW mode and sine wave inputs >5% of range. SLOW and MED are multi-sample A/D
conversions. FAST is DETector:BANDwidth 300 with nPLC = 1.0.
3. Applies to 0°–18°C and 28°–50°C.
4. For square wave inputs >10% of ACV range, except 100mV range. 100mV range frequency must be >10Hz
if input is <20mV.
5. Applies to non-sine waves >5Hz.
6. For 1k unbalance in LO lead.
7. Speeds are for 60Hz (50Hz) operation using factory defaults operating conditions (*RST). Autorange off,
Display off, Limits off, Trigger delay=0..
8. For ACV inputs at frequencies of 50 or 60Hz (±10%), add the following to “% of Range” uncertainty: 100mV
0.25%, 1V 0.05%, 10V 0.13%, 100V 0.03%, 750V 0.015 (Model 2701 only)..
9. Auto Zero off.
10. Sample count = 1024.
11. DETector:BANDwidth 300 with nPLC = 0.006 (2701 only).
12. Maximum useful limit with trigger delay = 175ms.
13. Includes measurement and binary data transfer out GPIB or ASCII data transfer for Ethernet and RS-232
(Reading Element only).