Programming Manual

Algovita Spinal Cord Stimulation System
Clinician Programming Manual 57
14. System Specications
iPad Specications
For non-therapy related precautions and Regulatory Statements regarding the use of the iPad, on which the Clinician
Programmer application resides, refer to the Apple iPad resources available at support.apple.com.
Bridge Communicator Specications
e Bridge Communicator complies with Safety/Compatibility standards:
IEC 60601-1, Medical Electrical Equipment Safety
IEC 60601-1-2, Electromagnetic Compatibility
is section of the manual provides technical information about the Bridge Communicator.
Bridge Communicator Specications
Operating temperature 0°C to 40°C (32°F to 104°F)
Storage temperature -20°C to 60°C (-4°F to 140°F)
Operating/storage humidity 10% to 85%
Operating/storage atmospheric pressure 70 kPa to 106 kPa (20.7 inHg to 31.3 inHg)
Size (approximate) 3.6 cm x 6.5 cm x 1.8 cm (1.45 in x 2.55 in x 0.72 in)
Weight including battery (approximate) 37.5 grams (1.3 oz)
Battery Chargeable lithium-ion battery
Mode of operation Continuous
Bridge Communicator Electromagnetic Compatibility Declaration Tables
is section lists the EMC Declaration tables. e Bridge Communicator is intended for use in the electromagnetic
environment specied below. e customer or the user of the Bridge Communicator should assure that it is used in such an
environment. e Bridge Communicator contains RF transmission and receiving capabilities; consequently, it is possible
that other portable and mobile RF communications equipment may interfere with the Bridge Communicator.
e power supply cable (maximum length 1500 cm [59 in]) is included in the system testing to demonstrate compliance
with the requirements of IEC 60601-1-2 2014. Use of accessories and cables other than those specically listed may result
in increased emissions or decreased immunity of the Bridge Communicator.
e Bridge Communicator should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is
necessary, the Bridge Communicator should be observed to verify normal operation in the conguration in which it will be
used.
0300-000176-001 Rev.2 SCS CP_Mobile.indb 57 7/12/19 9:20 AM