Chapter 4: High Performance Features & Safety Precautions
Product Service 1-800-522-7658
25
12. Speed Control (Integral Blower Models Only) regulates the speed of the
motorized impeller and is used by the certifier to validate and adjust the
inflow velocity. It is located behind the upper light assembly. See Figures
4-2 and Figure 3-4.
13. Space-Saving Design increases effective laboratory workspace, because
the impeller and carbon filter are self-contained on integral blower models.
On remote blower models, the interior space is increased to 23" interior
depth vs. 17.19" for integral blower models. See Figure 4-1.
14. Sash. The sash is constructed of 1/4 inch tempered safety glass. It pivots
up for loading and cleaning. See Figures 4-1 and 4-2.
15. Control Panel. The control panel, which is located above the sash,
contains the control switches and the electronics. See Figures 4-1 and 4-2.
16. Lamp Ballast (Not Shown) for the fluorescent lamp is located behind the
control panel.
17. Accessory Work Surface is dished and contoured to fit the dimensions of
the enclosure to contain spills. For histology/pathology applications, a wet
chemistry work surface is available with a sink and valves. See Chapter 7.
18. Carbon Filters (for Integral Blower Models Only) are necessary on
integral blower models. Carbon filters are designed to remove small
amounts of noxious fumes and chemicals. The carbon filters are supported
behind the perforated baffle. Two carbon filters are required. Carbon
filters are listed in Chapter 7. See Figure 4-2 and Figure 6-4.
19. Accessory Exhaust Connection Kits (For Remote Blower Models Only)
are available to duct to the outside. The exhaust connection allows the
operator additional protection from hazardous fumes or vapors and should
always be used if carbon filters capacity are not appropriate for your
application. See figures in Chapter 3 and Chapter 7 to order.
20. Utility Shelves (not shown) allow the user to hold utensils and printers on
shelves inside the enclosure. See Chapter 7.
Safety Precautions
1. Although the enclosure has been engineered to maintain optimum operator
safety, caution should always be used while working. Prior to using the
enclosure, check to make sure that the exhaust blower is operating and that
air is entering the enclosure at its specified face velocity. The use of an
airflow monitor is recommended to alert the user to a problem with airflow.
2. Use good housekeeping in the enclosure at all times. Clean up spills
immediately. Periodically clean enclosure interior.
3. Do not overload the work surface with apparatus or work material. The safe
operation of the enclosure is based upon having proper airflow through the
structure. Do not place large objects directly on the work surface. Instead,
elevate the object 3/4" on blocks to allow a flow of air under the object and
into the rear baffle exhaust slots. Ensure blocks are level and secured in place.