H-5223 Features
! Programming and data retrieval is through the serial port to a PC or a PDA using off the shelf
software. No special software required for normal operation.
! Same type of serial user interface as the XL series data logger allows current XL users a
quick learning curve.
! Compact menus built around PDA systems.
! Four general purpose analog input channels. Very high resolution with differential
measurements on channels 1 and 2.
! A precision 5.000 volt excitation that turns on automatically for each scan and a 12.0 volt
unregulated excitation programable to be turned on for each scan or left on all the time.
! Two digital I/O ports for general purpose I/O, pulsed or continuous outputs, counter inputs,
or a single quadrature input.
! Complete serial interface provides both a command mode for automated use and a menu
mode for human interaction. Both modes provide complete system programming and data
retrieval.
! Built-in data logging functions which store the data in non-volatile internal memory.
! Built in GOES radio supports 100, 300, and 1200 Baud GOES transmissions.
! Built in GPS time keeping for the GOES radio.
! Easily upgrades as the GOES radio in an XL series data logger system.
! Designed to be compatible with all SDI-12 sensors.
DOES NOT HAVE
! No frequency input options
! No 4 -20 mA output options
! No spare serial ports for applications requiring other serial devices.
! No built in functions or BASIC programming ability.
! No sensor mode for multiplexer operation.
! No built in display or key pad.
! Simplified data file system allows only 4 files all built in.
! Only a 5.0 volt input on the analog inputs.