TROUBLESHOOTING
TEMPERATURE
Temperature too high
1/ controller set too high
2/ controller failed on call Customer Service
3/ wiring error call Customer Service
Display reads "HI" or "400"+
Probe is unplugged, is broken or wire to sensor is broken trace wire
from display to probe; move wire and watch display to see intermittent
problems
Chamber temp spikes over set point and
then settles to set point
Recalibrate
Temperature too low
1/ high limit set too low see section 4.5
2/ controller set too low
3/ unit not recovered from door opening wait for display to stop changing
4/ unit not recovered from power failure or being turned off incubators
will need 24 hours to warm up and stabilize
5/ element failure call Customer Service
6/ controller failure confirm with front panel lights that controller is calling
for heat
7/ high limit failure confirm with front panel lights that High Limit is
operating correctly
8/ wiring problem check all functions and compare wiring to owners
manual - especially around any areas recently worked on
9/ loose connection check shadow box for loose connections
Display reads "LO"
1/ sensor is plugged in backwards call Customer Service
2/ if ambient room temperature is lower than range of unit compare set
points and ambient temperature to rated specifications in Section 10.0
Unit will not heat over a temperature that
is below set point
1/ confirm that blower motor is moving and that amperage and voltage
match data plate check blower motor motion in shadow box and feel for
air movement in chamber
2/ confirm that set point is set high enough turn High Limit clockwise and
see if High Limit light comes on
3/ check connections to sensor
4/ check calibration using independent reference thermometer.
Unit will heat up at all
1/ verify that controller is asking for heat by looking for controller light if
controller light is not on, there is a problem with the controller
2/check amperage amperage should be virtually at maximum rated
(data plate) amperage
3/ do all controller functions work?
4/ is the High Limit set high enough? for diagnostics, should be fully
clockwise with the pilot light never on
5/ HAS THE FUSE/CIRCUIT BREAKER BLOWN?
Indicated chamber temperature unstable
1/ ±0.1 may be normal
2/ is blower motor working? remove top panel and verify movement of