Programming Manual

Algovita Spinal Cord Stimulation System
Clinician Programming Manual
39
Deleting a Program or Subprogram
Deleting a program deletes all its subprograms. To delete a program or subprogram:
1. Tap Delete in the bottom right of the Summary screen.
e Delete Program popup displays.
2. Tap a program or subprogram dot.
3. Tap Delete in the popup.
Managing a Disabled Program
If a program has an orange triangle dispalyed next to its number on the Summary screen, the stimulator has disabled the
program, and it can no longer run. e stimulator disables programs when background impedance is on and the stimulator
detects high impedance. If this happens, run an impedance check and review the program. If the impedance check does
not detect errors and the program remains disabled, call Nuvectra Customer Service toll-free at 1-844-727-7897 within the
United States or 1-214-618-4980.
0300-000176-001 Rev.2 SCS CP_Mobile.indb 39 7/12/19 9:20 AM