Owner manual

■■■■■■■■■■■■■■■■■■■■■■
4 Installation
940 Professional IC Vario ONE/SeS/Prep 1 (2.940.1410)
■■■■■■■■
37
Rinse the separation column with eluent for approximately 10
minutes.
Use manual control in MagIC Net to stop the high-pressure pump
again: Off.
4.22 Conditioning
After installation, after each time the instrument is switched on and after
each time the eluent is changed, the system has to be conditioned with
eluent long enough that a stable baseline is attained.
NOTE
The conditioning time can lengthen considerably after changing the
eluent.
Conditioning the system
1
Preparing the software
CAUTION
Ensure that the configured flow is not higher than the flow permit-
ted for the corresponding column (refer to the column leaflet and
the chip data record).
Start the MagIC Net computer program.
Open the Equilibration tab in MagIC Net: Work-
place Run Equilibration.
Select (or create) a suitable method.
Also see: MagIC Net tutorial and online help.
2
Preparing the instrument
Ensure that the column is inserted correctly in relation to the flow
direction marked on the sticker (arrow has to point in the direc-
tion of flow).
Ensure that the eluent aspiration tubing is immersed in the eluent
and there is enough eluent in the eluent bottle.
3
Starting equilibration
Start the equilibration in MagIC Net: Workplace Run Equi-
libration Start HW.