Owner manual

■■■■■■■■■■■■■■■■■■■■■■
4 Installation
940 Professional IC Vario ONE/SeS/Prep 1 (2.940.1410)
■■■■■■■■
13
Removing the handle
4.3 Removing transport locking screws
To avoid damage to the drives for the high-pressure pump and the vac-
uum pump during transport, the pumps are secured with transport lock-
ing screws. These are located at the rear of the instrument and labeled
with Transport security screws.
Remove these transport locking screws before the initial start-up.
Accessories
For this step you need:
Hex key 4 mm (6.2621.030)