Owner's manual

73
Maintenance
Eppendorf BioSpectrometer
®
basic
English (EN)
8.3.2 Device self-test
The frequency of the automatic self-test (approx. 1 minute) can be set using the Device settings function
(see Device settings on p. 64). The factory setting for Self-test interval is "Weekly".
The self-test checks the following:
Verification of the detector
Determination of the random error across the entire available spectrum
Verification of the light source
Verification of the maximum available energy of the light source and quality of the light transmission
through the device
Determination of the signal random error at the reference sensor
Determination of the signal level at the reference sensor
Separate determination of light intensity in UV range
Determination of the systematic and random error of the wavelength
Position of an intensity peak in the UV range of the spectrum
Precision of the position of an intensity peak in the UV range of the spectrum
Select the Perform self-test function in the Device calibration group and confirm with enter.
At the end of the self-test, the display shows the message PASSED.
If the display shows the message FAILED, the self-test has failed. If this error cannot be corrected (see
Error messages on p. 76), contact Eppendorf Service.
8.4 Replacing fuses
The fuse holder is located between the mains connection socket and the mains power switch.
1. Disconnect the power plug.
2. Press the upper and lower end of the plastic springs 1 together and pull the fuse holder 2 fully out.
3. Replace faulty fuses and reinsert the fuse holder. Make sure that the guiding rail 3 is positioned
correctly.
DANGER! Electric shock.
Switch off the device and disconnect the power plug before starting maintenance or
cleaning work.
1
2
3