User Manual Part 1
INFORMATION FOR USE
X-RAY IDENTIFIER
1-23
Table 1-3. Characteristics as shipped (continued)
Parameter
Setting
Pacing Therapy available
DDDR (DR models) VVIR (VR models)
Sensor Accelerometer (MV for respiratory rate trend)
Pace/Sense Configuration RA: BI/BI (DR models)
Pace/Sense Configuration RV: BI/BI
The pulse generator is shipped in a power-saving Storage mode to extend its
shelf life. In Storage mode, all features are inactive except:
• Telemetry support, which allows interrogation and programming
• Real-time clock
• Commanded capacitor re-formation
• STAT SHOCK and STAT PACE commands
The device leaves Storage mode when one of the following actions occurs;
however, programming other parameters will not affect the Storage mode:
• STAT SHOCK or STAT PACE is commanded
• Tachy Mode is programmed to:
–Off
– Monitor Only
– Monitor + Therapy
Once you have programmed the pulse generator out of Storage mode, the
device cannot be reprogrammed to that mode.
X-RAY IDENTIFIER
The pulse generator has an identifier that is visible on x-ray film or under
fluoroscopy. This identifier provides noninvasive confirmation of the
manufacturer and consists of the following:
• The letters, BOS, to identify Boston Scientificasthemanufacturer
• The number, 112, to identify the Model 2868 PRM software application
needed to communicate with the pulse generator
- DRAFT -