User Manual
Appendix
36
Recommended separation distances between portable and mobile
RF communications equipment and the CardioMessenger
Transmission
Frequency
150 kHz to 80 MHz 80 MHz up to 800 MHz 800MHz to 2.5GHz
Equation
D = 1.17 D = 1.17 D = 2.34
Rated power of
transmitter (W)
Safe distance (m) Safe distance (m) Safe distance (m)
0,01 0,12 0,12 0,24
0,1 0,37 0,37 0,74
1 1,17 1,17 2,34
10 3,7 3,7 7,4
100 11,7 11,7 23,4
For transmitters whose maximum nominal output is not indicated in the above table, the distance can be calculated using the equation in the
column, where P is the maximum nominal output of the transmitter in Watts (W) according to the transmitter’s manufacturer.
Note: The higher frequency range applies at 80 MHz and at 800 MHz.
Note: These guidelines may not be applicable in all cases. The propagation of electromagnetic values is affected by absorption and reflection
by structures, objects and people.
P P P
CardioM essenger_II-S_en.book Page 36 Tuesday, May 6, 2008 1:11 PM