Exit calibration mode
TM-XDOP-01XN B–15
Exit calibration mode
When you have completed calibration, hit the EXIT key.
The SCPI command is:
CAL:STAT OFF, "0000"
Restore Factory Calibration
To restore the unit to the calibration constants set at the factory:
The SCPI command is:
CALibration:RESTore
Step # Do This You Will See
1 Output V Cal
2
or
Scroll to select Factory Cal.
Factory Cal
3 Select Y to Restore the factory
calibration. Selecting N will exit.
Restore? Y


Important:
This procedure should not be used in place of regular calibration, but may
be useful to restore the unit to an operational state in case of failure.