Technical information
50
West Mountain Radio Operating Manual
1. Start a test at the intended current for the battery you are planning to
test with the wiring you are planning to use for your test set up.
2. Record the voltage that the software reads “V_CBA:”
3. Record the voltage that your voltmeter reads “V_Battery:” at the
calibration point
4. Record the average “Current” “I:” shown by the software while the
voltages are being measured.
5. Stop the test and go to “Tools --> Lab Calibrate --> Voltage
Calibrate”.
In the calibrate voltage window, click “Reset” to clear any previous
calibration and then enter readings you have recorded. Press
“Calculate” and the calculated resistance will be valid for all future
tests with this CBA until it is reset. The voltage readings graphed
and displayed will be corrected accordingly. Press "OK" to save the
calibration settings for future tests with this CBA. If you change the
test setup in any way you should re-calibrate the setup or reset the
calibration.