User's Manual

LabQuest
®
3 – User Manual
17
what is recommended for the connected sensor, or that the number of samples could lead
to performance issues. You can still collect data using these settings; however, you may
have undesirable results.
The red-text warning indicates one of the following, and you cannot exit the dialog box
until the warning condition has been resolved:
The rate has been set to a value slower or faster than what a connected device
and/or sensor configuration can support
The number of samples exceeds the storage available
The number of samples exceeds 2000 (for rates faster than 80,000 samples/s)
Triggering Triggering is only available in the
Time-Based mode. When enabled, LabQuest
waits for a trigger condition to be met before
recording data collection. To set the trigger,
choose the sensor and set the threshold
conditions. You can also set the number of
points to be saved before the trigger condition is
met.
Advanced – The Advanced field has three options that can be independently enabled by
selecting the associated check box.
Oversampling can be used with data-
collection rates less than ten samples per
second. When enabled, the sensor will
sample at a higher rate than the number of
samples per second that you set, and then
LabQuest averages those readings and
records the averages in the data table. This
setting can be used to reduce measurement
noise by combining a burst of readings into
one value. As an example, oversampling
can sometimes reduce the influence of
unseen but real variations, such as as those
from a flickering light source.
Repeat can be used with data-collection rates less than 250 samples per second.
When enabled, a new data
-
collection run is started as soon as the current run has