9 Diagnostics and service
VEGAFLEX 81 • 4 … 20 mA/HART - four-wire
41825-EN-130314
Code
Text mes-
sage
Cause Rectication
S601
Overlling
Level echo in the close
range not available
Reduce level
100 % adjustment: Increase
value
Check mounting socket
Remove possible interfering
signals in the close range
Use coaxial probe
S602
Level with-
in the search
range, com-
pensation
echo
Compensation echo super-
imposed by medium
100 % adjustment: Increase
value
S603
Impermissi-
ble operating
voltage
Operating voltage below
specied range
Check electrical connection
if necessary, increase
operating voltage
The following table shows the error codes and text messages in the
status message "Maintenance" and provides information on causes
as well as corrective measures.
Code
Text mes-
sage
Cause Rectication
M500
Error in the
delivery sta-
tus
With the reset to delivery
status, the data could not
be restored
Repeat reset
Load XML le with sensor
data into the sensor
M501
Error in the
non-active
linearization
table
Hardware error EEPROM Exchanging the electronics
Send instrument for repair
M502
Error in the
event mem-
ory
Hardware error EEPROM Exchanging the electronics
Send instrument for repair
M503
Reliability too
low
Measurement certainty is
too low for a reliable meas-
urement
Process component or
probe contaminated or
defective
Check installation and
process conditions
Clean or exchange process
component or probe
M504
Error on an
device inter-
face
Hardware defect Exchanging the electronics
Send instrument for repair
Maintenance