9 Diagnosis, asset management and service
VEGAPULS 64 • 4 … 20 mA/HART - two-wire
51141-EN-160616
Code
Text message
Cause Rectication DevSpec State in CMD
48
S601
Overlling
Danger of vessel
overlling
Make sure that there is
nofurtherlling
Check level in the vessel
Bit 9 of Byte 14…24
S603
Impermissible operating
voltage
Operating voltage below
speciedrange
Check electrical con-
nection
If necessary, increase
operating voltage
Bit 11 of Byte 14…24
The following table shows the error codes and text messages in the
status message "Maintenance" and provides information on causes
as well as corrective measures.
Code
Text message
Cause Rectication DevSpec State in CMD
48
M500
Error with the reset deliv-
ery status
With the reset to delivery
status, the data could
not be restored
Repeat reset
LoadXMLlewithsen-
sor data into the sensor
Bit 0 of Byte 14…24
M501
Error in the non-active lin-
earizationtable
Hardware error
EEPROM
Exchanging the elec-
tronics
Send instrument for
repair
Bit 1 of Byte 14…24
M502
Error in the event memory
Hardware error
EEPROM
Exchanging the elec-
tronics
Send instrument for
repair
Bit 2 of Byte 14…24
M503
Meas. reliability too low
The echo/noise ratio
is too small for reliable
measurement
Antenna dirty or defec-
tive
Check installation and
process conditions
Change polarisation
direction
Use instrument with
higher sensitivity
Clean the antenna
Bit 3 of Byte 14…24
M504
Error on an device inter-
face
Hardware defect Check connections
Exchanging the elec-
tronics
Send instrument for
repair
Bit 4 of Byte 14…24
M505
No echo available
Sensor does not detect
an echo during opera-
tion
Antenna dirty or defec-
tive
Clean the antenna
Use a more suitable
antenna/sensor
Remove possible false
echoes
Optimizesensorposition
and orientation
Bit 5 of Byte 14…24
M506
Installation/Setup error
Error during setup Check or correct instal-
lation and/or parameter
adjustment
Bit 6 of Byte 14…24
Maintenance