User's Guide

32
Warning/Error
Screen
Cause
Solution
20-15
20-15
Battery critical
The battery must be
charged. Connect the
analyzer to power before
starting a new test.
WARNING
(This row also applies
to 20-16 and
20-20)
The battery level has become
critical (less than 10%) while
a test is in progress.
If a test has already been
started prior to this warning
message appearing, the test
can be completed.
Before a new test can be
started you must connect the
analyzer to an external power
source, which will also charge
the battery.
20-21
20-21
WARNING
Power not connected
Connect the analyzer to
Power before starting
the software update.
When a software upgrade is
being performed the analyzer
must be connected to
external power. This is to
ensure that the upgrade will
not be interrupted.
Connect the analyzer to
external power and start the
software upgrade again.
20-22
20-22
WARNING
Software update failed
Power down and
try again.
If this recurs,
contact Customer Support.
Something stopped the
software upgrade from being
performed.
Start the software upgrade
again. If it continues to fail,
contact Customer Support
(see page 38).
In the meantime, you can
continue to use your analyzer
using the existing software
version.
30-01
30-01
Partial fill error
Not enough sample was
applied to the strip.
Restart the test with
a new strip.
ERROR
Insufficient sample was
applied to the strip and the
test could not be completed.
You must not apply additional
sample to the test strip after
the test has begun.
Discard the strip and start the
test again. Apply the sample
as instructed on page 14.