Contents Dear Customers ................................................................... 2 Preliminary Remarks............................................................. 2 Precautions .......................................................................... 2 Parts Identification ............................................................... 3 Symbols ............................................................................... 4 Features ............................................................
Dear Customers Congratulations on purchasing a state-of-the-art A&D ear thermometer, one of the most advanced thermometers available today. We recommend that you read through this manual carefully before using the device for the first time. Preliminary Remarks This device conforms to the European Directive 93/42 EEC for Medical mark of conformity. Products. This is made evident by the (0366: The reference number to the involved notified body) Environment for use The device is for use indoors.
Parts Identification Probe cover Detection probe LCD On/Scan button with mark Battery cover Battery cover latch Part of Display MEMORY mark Scanning in progress Low battery mark Unit of temperature Select before measurement Temperature reading English 3
Symbols Symbols SN Function / Meaning Recommended Action Standby and Power on. Direct current Serial number Date of manufacture Type B: The device is designed to provide special protection against electrical shocks. Last reading stored in MEMORY. The thermometer will automatically turn off after 30 seconds of no operation When appears: The battery is low. Still able to measure. When flashes: The battery is low. Unable to measure. The ambient temperature is over 40 °C.
Features The UT-601 Ear Thermometer is a multi-use thermometer. It detects body temperature from the ear canal. It can also detect the temperature of solids, liquids or air in a fast and non-contact manner. The thermometer measures the infrared heat generated by the eardrum or other objects. Ear temperature accurately reflects core body temperature since the eardrum shares blood supply with the temperature control center in the brain (the hypothalamus).
Using the Thermometer Installing / Changing the Batteries 1. Use a small pointed object such as a ballpoint pen to press the battery cover latch and release the latch. 2. Remove the battery cover by pushing the sides of the cover upward while pressing the battery cover latch. 3. Insert two LR44 button batteries. Ensure that the plus poles are positioned upward as indicated by the marking on the battery compartment. 4. Replace the battery cover.
Using the Thermometer Measuring Ear Temperature 1. Press the On/Scan button to turn the thermometer on. All the display segments appear. Note: The thermometer will turn off automatically after 30 seconds of no operation. When turned on 2. Select the unit of measure as necessary. While all the display segments appear, press and hold the On/Scan button to toggle between °C and °F. Release the button when the unit to be used appears.
Using the Thermometer Measuring the Temperature of Liquid or Solid Objects Note: The display temperature range is 0 °C to 100 °C. If the temperature of the object is out of the range, an error, HI or Lo, appears. 1. Press the On/Scan button to turn the thermometer on. All the display segments appear. 2. The last reading stored in MEMORY appears and a short beep sounds to indicate that the thermometer is ready to scan. 5 mm 3.
Troubleshooting Problem Nothing appears on the display, even when the On/Scan button is pressed. Possible Reason Recommended Action Replace both batteries with new Batteries are drained. ones. Reinstall the batteries with Battery terminals are negative and positive terminals not in the correct matching those indicated on the position. battery compartment. Errors Display Meaning The thermometer will turn off automatically after 30 seconds of no operation. The ambient temperature is over 40 °C.
Maintenance Do not open the device. It uses delicate electrical component that could be damaged. If you cannot fix the problem using the troubleshooting instructions, request service from your dealer or from the A&D service group. The A&D service group will provide technical information to authorized dealers. The device was calibrated at the time of manufacture. If at any time you question the accuracy of measurement, please contact your dealer or the A&D service group.
Calibration Checks-Testing Instructions For trained specialist staff only Calibration checks must be performed by the following institutioins or persons: Manufacturer Trade measurements authorities Persons who fulfil the requirements of the legislation governing the operation of medical devices. (A) Function testing Function testing of the thermometer can only be carried out on a person or with a suitable simulator.
1-243 Asahi, Kitamoto-shi, Saitama 364-8585 JAPAN Telephone: [81] (48) 593-1111 Fax: [81] (48) 593-1119 A&D INSTRUMENTS LTD. Unit 24/26 Blacklands Way, Abingdon Business Park, Abingdon, Oxfordshire OX14 1DY United Kingdom Telephone: [44] (1235) 550420 Fax: [44] (1235) 550485 A&D ENGINEERING, INC. 1756 Automation Parkway, San Jose, California 95131 U.S.A. Telephone: [1] (408) 263-5333 Fax: [1] (408)263-0119 A&D Australasia Pty Ltd.