KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com Tel: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com Operating instructions Precision/ compact balances and platform scales KERN 572/573/KB/DS/FKB/FCB/KBJ Version 7.
GB KERN 572/573/KB/DS/FKB/FCB/KBJ Version 7.0 07/2013 Operating instruction Precision/compact balances and platform scales Table of contents: 1 2 MODE - MENUE ....................................................................................................... 3 Technical data ......................................................................................................... 5 2.1 2.2 2.3 2.4 2.5 2.6 KERN 572 ......................................................................................
1 MODE - MENUE Models 572/573/KB/DS/FKB/FCB: Important! The modified settings - as well as the adjustment - must be stored when switching-off, via the ON/OFF button.
Models KBJ: Important! The modified settings - as well as the adjustment - must be stored when switching-off, via the ON/OFF button.
2 Technical data 2.1 KERN 572 KERN 572-30 572-31 572-32 572-33 Readability (d) 0.001 g 0.001 g 0.001 g 0.01 g Weighing range (max) 240 g 300 g 420 g 1600 g Taring range (subtractive) 240 g 300 g 420 g 1600 g 0.001 g 0.002 g 0.002 g 0.01g ± 0.003 g ± 0.005 g ± 0.005 g ± 0.03 g Smallest piece weight 0.001 g 0.001 g 0.001 g 0.01 g Adjustment points 50/100/ 50/100/ 100/200/ 0.5/1.0/ 200/240 g 200/300 g 300/400 g 1.5/1.
KERN 572-35 572-37 572-39 572-43 Readability (d) 0.01 g 0.01 g 0.01 g 0.1 g Weighing range (max) 2400 g 3000 g 4200 g 10000 g Taring range (subtractive) 2400 g 3000 g 4200 g 10000 g Reproducibility 0.01 g 0.02 g 0.02 g 0.1g ± 0.03 g ± 0.05 g ± 0.05 g ± 0.3 g Smallest piece weight 0.01 g 0.01 g 0.01 g 0.1 g Adjustment points 0.5/1.0/ 1.0/1.5/ 1.0/2.0/ 2/5/10 kg 2.0/2.4 kg 2.0/3.0 kg 3.0/4.
KERN 572-45 572-49 572-55 572-57 0.05 g 0.1 g 0.05 g 0.1 g Weighing range (max) 12000 g 16000 g 20000 g 24000 g Taring range (subtractive) 12000 g 16000 g 20000 g 24000 g 0.05 g 0.1 g 0.1 g 0.1g ± 0.15 g ± 0.3 g ± 0.25 g ± 0.3 g 0.05 g 0.1 g 0.05 g 0.
2.2 KERN 573 KERN Accuracy class 573-34NM II 573-46NM II Readability (d) 0.01 g 0.1 g Verification value (e) 0.1 g 1g Weighing range (max) 650 g 6500 g Minimum load (Min) 0.5 g 5g Taring range (subtractive) 650 g 6500 g Reproducibility 0.01 g 0.1 g ± 0.03 g ± 0.3 g 0.01 g 0.1 g 200g/500g/600g 2/5/6 kg 500 g + 100 g 5 kg + 1 kg Linearity Smallest piece weight Adjustment points Recommended adjusting weight F1 (not supplied) Humidity of air max. 80% rel.
2.3 KERN KB/KBJ KERN KB 120-3N KB 240-3N KB 360-3N 0.001 g 0.001 g 0.001 g Weighing range (max) 120 g 240 g 360 g Taring range (subtractive) 120 g 240 g 360 g 0.001 g 0.001 g 0.002 g ± 0.003 g ± 0.003 g ± 0.005 g 0.001 g 0.001 g 0.001 g 20/50/100/120 g 100/150/ 100/200/ 200/240 g 300/360 g 200 g 200 g + 100 g Readability (d) Reproducibility Linearity Smallest piece weight Adjustment points Recommended adjusting weight F1 (not supplied) 100 g Humidity of air max. 80% rel.
KERN KB 1200-2N KB 2000-2N KB 2400-2N KB 3600-2N Readability (d) 0.01 g 0.01 g 0.01 g 0.01 g Weighing range (max) 1 200 g 2 000 g 2 400 g 3 600 g Taring range (subtractive) 1 200 g 2 000 g 2 400 g 3 600 g Reproducibility 0.01 g 0.01 g 0.01 g 0.02 g ± 0.03 g ± 0.03 g ± 0.03 g ± 0.05 g 0.01 g 0.01 g 0.01 g 0.01 g Linearity Smallest piece weight Adjustment points Recommended adjusting weight F1 (not supplied) 0.2/0.5/1.0/1.2 kg 0.5/1.0/1.5/2.0 kg 1.0/1.5/2.0/2.4 kg 1.0/2.0/3.
KB 10000-1N KB 10k0.05N 0.1 g 0.05 g Weighing range (max) 10 000 g 10 000 g Taring range (subtractive) 10 000 g 10 000 g Reproducibility 0.1 g 0.05 g Linearity 0.3 g 0.15 g Smallest piece weight 0.1 g 0.05 g 2/5/10 kg 2/5/10 kg Recommended adjusting weight F1 (not supplied) 10 kg 10 kg Stabilization time (typical) 3 sec. 3 sec. +10 °C ... + 40 °C +10 °C ...
KB 650-2NM KB 6500-1NM KBJ 650-2NM Accuracy class II II II Readability (d) 0.01 g 0.1 g 0.01 g Verification value (e) 0.1 g 1g 0.1 g Weighing range (max) 650 g 6 500 g 650 g Minimum load (Min) 0.5 g 5g 0.5 g Taring range (subtractive) 650 g 6 500 g 650 g Reproducibility 0.01 g 0.1 g 0.01 g ± 0.03 g ± 0.3 g ±0.03 g 0.01 g 0.1 g 0.01 g 200/500/600 g 2/5/6 kg - 600 g 6.0 kg Int.
2.4 KERN DS DS 3K0.01S DS 5K0.05S DS 8K0.05 DS 10K0.1S Readability (d) 0.01 g 0.05 g 0.05 g 0,1 g Weighing range (max) 3 000 g 5 000 g 8 000 g 10.000 g Taring range (subtractive) 3 000 g 5 000 g 8 000 g 10.000 g Reproducibility 0.02 g 0.05 g 0.05 g 0.1 g ± 0.05 g ± 0.15 g ± 0.15 g ±0.3 g 0.01 g 0.05 g 0.05 g 0.
DS 16K0.1 DS 20K0.1 DS 30K0.1 0.1 g 0.1 g 0.1 g Weighing range (max) 16 000 g 20 000 g 30 000 g Taring range (subtractive) 16 000 g 20 000 g 30 000 g 0.1 g 0.1 g 0.2 g ± 0.3 g ± 0.3 g ± 0.5 g 0.1 g 0.1 g 0.1 g Adjustment points 5/10/15/16 kg 5/10/15/20 kg 10/15/20/30 kg Recommended adjusting weight F1 (not supplied) 10 kg + 5 kg 20kg 20 kg + 10 kg KERN Readability (d) Reproducibility Linearity Smallest piece weight Humidity of air max. 80% rel.
DS 36K0.2 DS 30K0.1L DS 36K0.2L DS 60K0,2 0.2 g 0.1 g 0.2 g 0.2 g Weighing range (max) 36 000 g 30 000 g 36 000 g 60 000 g Taring range (subtractive) 36 000 g 30 000 g 36 000 g 60 000 g 0.2 g 0.2 g 0.2 g 0.4 g ± 0.6 g ± 0.5 g ± 0.6 g ± 1.0 g 0.2 g 0.1 g 0.2 g 0.
DS 65K0.5 DS 100K0.5 DS 150K1 0,5 g 0,5 g 1g Weighing range (max) 65 000 g 100 000 g 150 000 g Taring range (subtractive) 65 000 g 100 000 g 150 000 g 0,0 g 0,5 g 1g ± 1,5 g ± 1,5 g ±3g 0,5 g 0,5 g 1g 20/30/50/60 kg 20/50/100 kg 50/100/150 kg 50 kg 50 kg + 50 kg KERN Readability (d) Reproducibility Linearity Smallest piece weight Adjustment points Recommended adjusting weight F1 (not supplied) Humidity of air 3 sec. Permitted environmental temperature +10 °C ...
Accuracy class DS 65K1M II Readability (d) 1g Verification value (e) 10 g KERN Weighing range (max) 65 000 g Minimum load (Min) 50 g Taring range (subtractive) 65 000 g Reproducibility 1g Linearity ±2g Smallest piece weight 1g Adjustment points 20/30/50/ 60 kg Recommended adjusting weight F1 (not supplied) Humidity of air 50 kg max. 80% rel. (non-condensing) Stabilization time (typical) 3 sec. Permitted environmental temperature +10 °C ...
2.5 KERN FCB FCB 6K0.02B FCB 12K0.05B FCB 12K0.1B FCB 24K0.1B FCB 24K0.2B Readability (d) 0.02 g 0.05 g 0.1 g 0.1 g 0.2 g Weighing range (max) Taring range (subtractive) Reproducibility Linearity ± Smallest piece weight 6 000 g 12 000 g 12 000 g 24 000 g 24 000 g 6 000 g 12 000 g 12 000 g 24 000 g 24 000 g 0.04 g 0.1 g 0.02 g 0.05 g 0.15 g 0.05 g 0.1 g 0.3 g 0.1 g Adjustment points 2/5/6 kg 2/5/10/12 kg 2/5/10/12 kg 0.1 g 0.3 g 0.1 g 5/10/15/ 20/24 kg 0.2 g 0.6 g 0.
2.6 KERN FKB FKB 6K0.02 FKB 8K0.05 FKB 16K0.05 FKB 16K0.1 Readability (d) 0.02 g 0.05 g 0.05 g 0.1 g Weighing range (max) 6 000 g 8 000 g 16 000 g 16 000 g Taring range (subtractive) 6 000 g 8 000 g 16 000 g 16 000 g Reproducibility 0.04 g 0.05 g 0.1 g 0.1g Linearity ± 0.1 g ± 0.15 g ± 0.25 g ± 0.3 g Smallest piece weight 0.02 g 0.05 g 0.05 g 0.
FKB 36K0.1 FKB 36K0.2 FKB 65K0.2 FKB 65K0.5 0.1 g 0.2 g 0.2 g 0.5 g Weighing range (max) 36 000 g 36 000 g 65 000 g 65 000 g Taring range (subtractive) 36 000 g 36 000 g 65 000 g 65 000 g 0.2 g 0.2 g 0.4 g 0. g ± 0.5 g ± 0.6 g ± 1.0 g ± 1.5 g 0.1 g 0.2 g 0.2 g 0.
Accuracy class FKB 8K0.1M II FKB 65K1M II Readability (d) 0.1 g 1g Verification value (e) 1.0 g 10 g 8 000 g 65 000 g 5.0 g 50 g 8 000 g 65 000 g 0.1 g 1g ± 0.3 g ±3g 0.1 g 1g Adjustment points 2/5/7/8 kg 20/30/50/60 kg Recommended adjusting weight F1 (not supplied) 5 kg + 2 kg 50 kg KERN Weighing range (max) Minimum load (Min) Taring range (subtractive) Reproducibility Linearity Smallest piece weight Humidity of air max. 80% rel.
3 Fundamental information (general) It is essential to read through and observe the complete operating instructions before installing and commissioning. 3.1 Intended use The balance you have acquired serves to determine the weighing value of the material to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing plate.
4 4.1 Fundamental safety information Observe the information in the operating instructions Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances. Versions in other languages are non-binding translations. The only binding version is the original document in German. 4.2 Staff training The device may only be operated and looked after by trained members of staff. 5 Transport and storage 5.
6 Unpacking, installation and commissioning 6.1 Place of installation, place of use The balance is constructed in such a way that reliable weighing results can be achieved under normal application conditions. By selecting the correct location for your balance, you will be able to work quickly and precisely.
6.4 Battery operation FKB To insert the batteries (6 x 1.5 V) remove the battery compartment cover. Remove it with the help of a coin. In the each battery tube insert three batteries in the same polarity sense. Screw down again the battery cover. To save the battery, the background illumination can be switched off (see chap. 7.3). Moreover the AUTO-OFF function can be activated (see chap. 7.2.10).
6.8 Adjusting (go to 7.2.6) Using a precision weight, the accuracy of the balance can be checked at any time and adjusted. Warning: With verified balances, this precision adjustment is not possible Adjustment procedure: Check that the surrounding conditions are stable. A short warm-up time of about 15 minutes is recommended for stabilisation. 6.8.1 Adjusting for verification (KERN 573; FKB-M; KB-NM; FCB-M; DS-M) General Information: Prior to the procedure of verification the balance has to be adjusted.
6.9 Verification General introduction: According to EU directive 90/384/EEC balances must be verified if they are used as follows (legally controlled area): a) For commercial transactions if the price of goods is determined by weighing. b) For the production of medicines in pharmacies as well as for analyses in the medical and pharmaceutical laboratory. c) For official purposes. d) For manufacturing final packages. In cases of doubt, please contact your local trade in standard.
7 7.1 Operation Display control panel Keyboard ON / OFF Print weighing result in MODE: No / descending in % and counting mode: Form reference in MODE: Yes / ascending MODE configuration (see mode structure diagram page 2) Formula weighing in MODE: to left Change g-pieces Unit changeover in MODE: to right Tare in MODE: back in weighing operation Display-Symbol Meaning ==OVERLOAD== Overload: Weighing range has been exceeded = ============ Underload: Weighing range has been fallen short of .
7.
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1370
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1370 31
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1370
7.2.13 Animal weighing: Activate via mode/animal weigh. Y/N (4xMode) Put container without measured object on the weighing surface and tare Place the measured object (animal) in the container and start with REF/YES key The value averaging in shown in the display with AVG as well as by extinguishing of the character blocks and the subsequently fixed display value. Printing and deleting of the average value can be triggered by the PRINT key Simple deleting is possible by pressing the TARE key.
7.2.15 Contents of the form printout: MODE / PRINTER / PRINTERFORM Printerform REF / YES MODE Var 1 Type xx = Gross Var 2 Type xx = Net Var 3 Type xx = Tare Var 4 Type xx = Pretare MODE Var 1 Row xx XXxx Gross value active MODE Var 1 Col xx Var 1 Type 02 = Gross Line 2 Var 1 Row 02 Column 14 Var 1 Col 14 Var 5 Type xx = Pcs / % Printout “Gross“ line 2 / column 14 Var 6 Type xx = * / Mw Var 7 Type xx = Numerator 34 xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx 1. line xxxxxxxxxxxxxG 15000.0g xxxxxx 2.
7.3 Display background illumination After having switched on your balance and the zero display, press the „MODE“ button to select the menu point „Backlight“. Acknowledge by pressing the „YES“ button in order to switch on the permanent background illumination. By pressing the „NO“ button, the background illumination is switched off.
7.5 RS 232 C Data output via interface RS 232 C RS 232 C Data output via interface RS 232 C General information As a condition for the data transfer between the balance and a peripheral device (for instance printer, PC ...) both devise have to be set on the same interface parameter (for instance baud rate, parity ...). There are 5 methods for the data output via RS 232 C Data output via PRINT-Key The printing process can be released by the PRINT-key.
7.5.1 Description of the data transfer Structure of each data transfer: Bit-Nr. N B*: . B, 0, , g: E CR: LF: 1 B* 2 N = = = = = = 3 N 4 N 5 B 6 B 7 B 8 B 9 B 10 0 11 . 12 0 13 0 14 E 15 E 16 E 17 CR 18 LF Numerator Blank or in the range of zero point. Blank or weighing value with unit, depending on the load on the weighing plate. Unit Carriage Return Line Feed 7.5.2 Numerator The Numerator is found under the menu point “Printer” and can be activated and deactivated.
7.7 Underfloor weighing Objects which, because of their size or shape, cannot be put on the scale, can be weighed by means of underfloor weighing. Proceed as follows: Switch off the balance. Turn the balance over, without loading the balance plate. Open the cover plate on the base of the balance. Hang on the hook for underfloor weighing . Place the balance over an opening. Hang the item to be weighed on the hook and carry out weighing.
8 Maintenance, upkeep, disposal 8.1 Cleaning Please disconnect the device from the operating voltage before cleaning. Only use a cloth dampened with mild suds and not aggressive cleaning agents (solvents or similar). Please ensure that fluids are not able to get into the device and rub off using a clean, soft cloth. Loose sample residue/powder can be removed carefully using a brush or hand vacuum cleaner. Remove any spilt material to be weighed immediately. 8.
9 Troubleshooting The balance should be switched off for a short time following an interruption in the programme sequence and disconnected from the mains supply. It is then necessary to repeat the weighing process from the beginning. Help: Interruption Possible cause Weight display is not illuminated. The balance is not switched on. The mains supply connection has been interrupted (mains cable not plugged in/faulty). Power supply interrupted. .
10 Declaration of conformity KERN & Sohn GmbH D-72322 Balingen-Frommern Postfach 4052 E-Mail: info@kern-sohn.de Tel: 0049-[0]7433- 9933-0 Fax: 0049-[0]7433-9933-149 Internet: www.kern-sohn.
KERN & Sohn GmbH D-72322 Balingen-Frommern Postfach 4052 E-Mail: info@kern-sohn.de Tel: 0049-[0]7433- 9933-0 Fax: 0049-[0]7433-9933-149 Internet: www.kern-sohn.