User manual

Bonding Procedures of BCM Strain Gauges
BCM SENSOR TECHNOLOGIES BVBA
Brechtsebaan 2
B-2900 Schoten - Antwerpen, BELGIUM
Industriepark Zone 4,
Tel.: +32-3-238 6469
Fax: +32-3-238 4171
website: www.bcmsensor.com
email: sales@bcmsensor.com
1. Introduction
This document describes the working steps required for bonding and soldering of BCM strain gauges on to
the load-cell element or on to the surface of the object for stress analysis. The described processes are
tailored and proven to be helpful for manufacturing of precision load cells or for precision strain
measurement. The described procedures can be applied to both foil strain gauges and semiconductor strain
gauges.
It is assumed that the strain gauges are mainly got bonded with heat curing adhesives, like BCM B-610
adhesive. So the application notes described below will mainly focus on the gauge installation with the heat
curing adhesives, especially on the use of BCM B-610 adhesive.
2. Surface Preparation
It is important to be sure that the load-cell element had no any contact with silicon-containing cutting oil
during machining. Otherwise the silicon residuals have to be removed completely by means of suitable
solvents before starting the working steps described below. This is very necessary because the smallest
amount of silicon contamination can make a bond useless. Therefore, we recommend very much to keep
the working area free from any silicon containing product.
The staff in charge of strain gauge application should know that cosmetics such as hand cream frequently
contain silicon. Therefore, it is recommended using the hand cream that is free of silicon.
2.1. Clean load-cell element first with degreasing solvent such as methyl-ethyl-ketone (MEK), acetone, or
isopropyl alcohol (IPA). After the cleaning, do not touch application area by hand for all further working
steps.
2.2. Make it ready in hand of sandpaper adhesion surface area with grit size 320 - 400 or sandblast with
aluminum oxide grit 120 - 400. When using a sandblast system, operate it with oil-free compressed air and
use it for surface preparation only. By using the system for other than surface preparation there is the
danger of contaminating the system with silicon or other chemicals that may cause a bad bond.
2.3. Remove grinding dust with a non-fluffy throw-away cloth and a solvent, e.g. IPA, or by brushing with
IPA.
2.4. If required, the position of strain gauge to be bond can now be marked by a brass needle or a hard 4H-
pencil.
3. Handling and Cleaning of Strain Gauges
Strain gauges are handled only with rounded tweezers or vacuum pen, but never touched with the fingers.
Hold the strain gauge at the backing support, not at the grid sensing area.
To manufacture high precision load cells, it is recommended to clean BCM strain gauges before the bonding
process. One can clean the gauges with ethyl alcohol, MEK, acetone, or IPA. After the cleaning, never
touch the gauges until bonding. Should the strain gauges have been touched accidentally by hand, clean it
immediately with MEK or IPA and cotton tipped applicators (do not use cotton buds with plastic grip). Put
the cleaned gauge onto a chemically cleaned glass plate with the gauge bonding side upper.