Manual

Page 25
5.2.6 If code parameters are to be measured, select the desired
parameter using the Code Function Switch. Refer to Section 2.1 for
a description of these functions. Peak levels are measured using
the same scales as used in the NORM mode. Rate is read using the
black CODE RATE scale. On Time is read using the black percent-
age (%) scale. Code parameters may be read on any range provided
that the peak level is approximately ten percent of full scale.
5.2.7 Disconnect the test leads from the circuit under test and
return the Range Switch to the OFF position.