000 Series Precision Multi Product Calibrator Operation Manual TRANSMILLE LTD. Version 1.
1000 SERIES OPERATION MANUAL IMPORTANT NOTICE THIS CALIBRATOR WILL REQUIRE AN UNLOCK CODE AFTER THE EVALUATION PERIOD HAS EXPIRED. (60 Days after invoice date) AFTER THE EVALUATION PERIOD HAS EXPIRED THE OPERATION OF THE CALIBRATOR IS LOCKED AND THE DISPLAY SHOWS A NUMBER WHICH MUST BE QUOTED TO TRANSMILLE TO RECEIVE THE UNLOCK CODE THE UNLOCK CODE IS AVAILALBLE FROM TRANSMILLE ONLY AFTER PAYMENT HAS BEEN RECEIVED. This code only needs to be entered once in the life of the instrument.
1000 SERIES OPERATION MANUAL DECLARATION OF CONFORMITY Manufacturer’s Name: Manufacturer’s Address: Declares, that the product Product Name: Model Number: Product Options: Transmille Ltd.
1000 SERIES OPERATION MANUAL TABLE OF CONTENTS 1000 SERIES CALIBRATION INTRODUCTION .......................................................................................... 6 MAIN FEATURES ................................................................................................................................................ 6 ACCURACY AND FUNCTIONALITY .....................................................................................................................
1000 SERIES OPERATION MANUAL SETTING PASSIVE RESISTANCE OUTPUT .......................................................................................................... 48 SETTING CAPACITANCE OUTPUT ..................................................................................................................... 50 SETTING FREQUENCY OUTPUT. ....................................................................................................................... 52 THERMOCOUPLE SIMULATION .................
1000 SERIES OPERATION MANUAL 1000 Series Calibration Introduction The 1000 series of calibrators offer the smallest and by far the most portable multi product multi-function calibrator in the world.
1000 SERIES OPERATION MANUAL Accuracy and Functionality The 1000 Series calibrators are available in a rugged portable case or a bench case. A rack mounting option is also available. True Multiproduct Calibration Designed to provide an accurate cost effective portable instrument for the calibration of multimeters, clamp meters, frequency meters and temperature meters. Designed for use in the laboratory or portable onsite calibration.
1000 SERIES OPERATION MANUAL Preparing The Calibrator For Use. Initial inspection. After shipment the calibrator should be inspected for any signs of external damage. Should external damage be found contact the carrier immediately. Do not connect a damaged instrument to the line power as this may result in internal damage. Please retain the original packaging; this should be used when returning the calibrator for service and recalibration.
1000 SERIES OPERATION MANUAL Positioning the Calibrator The 1000 series calibrator is available in a variety of different casings for use on site, on a bench or in a rack. Care should be taken when positioning the calibrator to ensure that items are not placed against the cooling vents.
00 SERIES OPERATION MANUAL Power and Interface Connections Connections on the front panel are for Line Power via a 3 Pin IEC connector incorporating the Line fuse and on-off switch, and a Female Type B USB connector for interfacing with a computer.
1000 SERIES OPERATION MANUAL Checking the Line Voltage Warning: The line power cord must have an earth conductor to avoid risk of shock. This instrument must be correctly earthed. The calibrator must be ordered with 100-120 Volt line supply or 220 - 240 Volt line supply option. Check that the line voltage matches the configured voltage before connecting the power to the instrument. The input voltage is indicated on the front panel underneath the power switch.
1000 SERIES OPERATION MANUAL Warning: Ensure Line Power has been disconnected from the instrument before performing this procedure Using a Flat Head screwdriver, insert the end of the screw driver into the recess on the fuse holder. The screwdriver must be parallel to the front panel of the calibrator Once the screwdriver is in the recess of the fuse holder, gently hinge the fuse holder up (away from the front panel). Be careful not to damage the fuse holder by applying too much force.
1000 SERIES OPERATION MANUAL Connecting to a computer A USB cable (supplied) should be used to connect the calibrator to a USB port on a computer. Connection Details Connection from calibrator to a computer : The USB connection on the 1000 series calibrator is a Female type B connection. Also supplied is a USB driver on CD : For details on installing USB driver see appendix A. TRANSMILLE LTD. Version 1.
1000 SERIES OPERATION MANUAL Powering up the calibrator After connecting line power, the calibrator can be switched on with the line power switch above the mains inlet socket on the front panel. The fan will start and the front panel display will illuminate indicating power. The display will show a firmware version number and after a short delay, during which time the processor performs a self-test of the instrument, the display will show an output of 0.0000mV DC.
1000 SERIES OPERATION MANUAL Output Connections Warning: Risk of shock. High voltages may be present on the output sockets. Output sockets are 4mm safety type, the voltage pair contacts are low thermal gold plated for minimum thermal EMF. The 1000 series calibrator’s outputs have been designed to allow most multimeters to be calibrated without changing connections. There are 4 sets of outputs: 1) Voltage, Resistance, Capacitance, Frequency. 2) Current to 1A. 3) High 10A Current.
1000 SERIES OPERATION MANUAL It is recommended that the voltage and low current leads be high quality screened cable with gold plated 4mm plugs fitted. The cable must be able to withstand 1025 volts AC and have an insulation resistance greater than 1T to avoid introducing any shunting effect on the high resistance ranges.
1000 SERIES OPERATION MANUAL Operation Safety Warnings WARNING: The information in this section is intended only for qualified personnel. The user must at all times be adequately protected from electric shock. Qualified personnel must ensure that operators of the equipment are adequately insulated from connection points. WARNING: This instrument is capable of generating both DC and AC high voltages.
Front Panel Controls and Indicators LCD Display with integral Backlight Menu buttons (Soft Keys) Numeric Keypad Multiplier Keys Function Keys Thermocouple Output Active Terminal Indication LEDs Safety Output Terminals Digital Control Enter Key Shift Key Adapter Interface TRANSMILLE LTD. Standby Button Output On Button Ref Key and Directional Arrows Version 1.
Graphic LCD Display The graphic backlit LCD display shows the present output, instrument status, % or ppm change from the entered value, and also the new value being entered. The bottom portion of the display is used to assign the function of the four ‘soft keys’. Output Value & Range Dynamic Soft key Menu Low to Ground Output On / Standby Indicators Deviation (ppm) Function Specific Configuration Display TRANSMILLE LTD. Version 1.
1000 SERIES OPERATION MANUAL Front Panel Keyboard The front panel of the 1000 Series Calibrator utilises a high quality custom rubber keyboard with tactile feel buttons and integral display window. The front panel is therefore sealed against the ingress of moisture and dirt enabling the calibrator to be used in most working environments without risk of early failure of the operating buttons. The front panel can easily be wiped clean with a soft cloth. Care should be taken not scratch the display window.
1000 SERIES OPERATION MANUAL The Keyboard is divided into sections to allow easy operation.
1000 SERIES OPERATION MANUAL Digital Control and Cursor Keys A digital potentiometer allows the ‘highlighted digit’ on the display to be incremented (turning clockwise) or decrement (turning anti-clockwise). As an output is changed, the deviation from the original value entered on the keyboard is shown in either % or ppm depending upon the magnitude of the change.
1000 SERIES OPERATION MANUAL Terminal status LED’s LED’s above the terminals indicate which pair is active. When terminals are not active they are electrically isolated from each other All 4mm safety sockets share a single common terminal. Active terminals indicated by illuminated LED Voltage Output Terminal Pair (Black & White) Low thermal 4mm safety terminals Used for all voltage outputs up to 1025V, for resistance, capacitance and frequency.
1000 SERIES OPERATION MANUAL Low Current Output Terminal Pair (Black & Blue) 4mm safety terminals Used for current outputs up to 1 Amp High Curent Output Terminals (Black and Yellow) 4mm Safety terminals Used for all currents above 1 Amp. Thermocouple Output Terminals (White) Mini thermocouple socket used for generation of thermocouple output. Ensure the correct thermocouple wiring is used to match the type (i.e. Type K, Type J) TRANSMILLE LTD. Version 1.
1000 SERIES OPERATION MANUAL 9 Pin Adapter Interface Connector. To expand the 1000 Series calibrators, an adapter interface is provided on the front panel. This allows for connection to external adapters used for extending calibration capability, e.g. Pressure measurement etc. Incorporates a yellow LED to indicate when the adapter interface is active.
1000 SERIES OPERATION MANUAL Soft Key Menus Menu Structure To access advanced functions of the calibrator, the 4 soft keys below the screen can be used to select the menu functions that appear at the bottom of the screen. There are 3 pages of menus, each featuring 3 separate functions and a next key. Key Name +/- Function In DC functions, pressing this key will invert the polarity, e.g.
1000 SERIES OPERATION MANUAL Key Name Function Pressing this key will enter the Calibration function, further described Pressing this key will enter the setup menu, further described Pressing this key will display the info screen, further described CALIBRATE SETUP INFO Procedure Menu The 1000 series calibrators has the ability to store procedures in memory for on site use. After pressing the ‘PROCEDURE’ key the calibrator will display a list of procedures that have been loaded.
1000 SERIES OPERATION MANUAL Setup Menu To configure the 1000 series calibrator, a setup menu is provided. This allows users to configure options, i.e. the Calibration Password. Using the arrows keys (as described on page 22), move the cursor to the desired function and press the ‘Select’ soft key or press the enter button. Function Name Beeper Password Adapters Range Hold TRANSMILLE LTD. Description This menu item allows the beeper to be turned on or off.
1000 SERIES OPERATION MANUAL Connection Diagrams Provided in this section are example connection diagrams for typical pieces of test equipment, indicating the terminals to use DC / AC Voltage DC / AC Current – Outputs below 1A TRANSMILLE LTD. Version 1.
1000 SERIES OPERATION MANUAL DC/AC Current – Outputs above 1A Resistance TRANSMILLE LTD. Version 1.
1000 SERIES OPERATION MANUAL Capacitance Frequency TRANSMILLE LTD. Version 1.
1000 SERIES OPERATION MANUAL Thermocouple PRT / RTD TRANSMILLE LTD. Version 1.
1000 SERIES OPERATION MANUAL Insulation Resistance / Test Voltage Measurement Continuity Resistance / Current Measurement TRANSMILLE LTD. Version 1.
1000 SERIES OPERATION MANUAL Setting An Output Using the Keyboard Setting the output of the calibrator is similar to entering values on a calculator. Simply press the keys to enter the value required, select a multiplier (i.e. m for milli), and then select the units, i.e. volts. The new value will appear under the current set value on the calibrator display.
1000 SERIES OPERATION MANUAL Adjusting the output using the digital control After the output has been set, any digit of the output display can be incremented or decremented using either the digital control or the up and down arrow keys. This function makes calibration of analogue meters easy, where deviating the output of the calibrator rather than interpreting the indicated value from the UUT provides more accurate results.
1000 SERIES OPERATION MANUAL Setting a DC Voltage Output Complete the following procedure to set a dc voltage output. The and keys can be used to edit the entry in the event of an incorrect key press. Warning ENSURE THE OUTPUT DOES NOT EXCEED RATING FOR UUT INPUT 1) Ensure the calibrator’s output has been set to Standby. This can be verified by ensuring that standby is indicated on the display, and the output standby status LED is lit.
1000 SERIES OPERATION MANUAL 10) Press . The new value will replace the existing output in the middle of the display. 11) Press to activate the calibrator output. The LED next to the key will light up, as well as the terminal indicator; indicating that the output is active. The calibrator will now produce 56.789 mV DC at the voltage terminals. Once on a range, any new output within that range can be set without the calibrator returning to standby.
1000 SERIES OPERATION MANUAL Setting a DC Current Output Complete the following procedure to set a DC current output. The and keys can be used to edit the entry in the event of an incorrect key press. Warning: ENSURE CONNECTIONS TO UUT ARE CORRECT To avoid damaging UUT protection fuses, ensure the correct terminals on the UUT are used before sourcing currents 1) Ensure the calibrator’s output has been set to Standby.
1000 SERIES OPERATION MANUAL 10) Press . The new value will replace the existing output in the middle of the display. 11) Press to activate the calibrator output. The LED next to the key will light up, as well as the terminal indicator; indicating that the output is active. The calibrator will now source 29 mA DC at the low current terminals. Once on a range, any new output within that range can be set without the calibrator returning to standby.
1000 SERIES OPERATION MANUAL Setting an AC Voltage Output. Complete the following procedure to set a AC voltage output. The and keys can be used to edit the entry in the event of an incorrect key press. WARNING ENSURE THE OUTPUT DOES NOT EXCEED RATING FOR UUT INPUT 1) Ensure the calibrator’s output has been set to Standby. This can be verified by ensuring that standby is indicated on the display, and the output standby status LED is lit.
1000 SERIES OPERATION MANUAL 9) Press . The new voltage will replace the existing output in the middle of the display 10) Press the numeric and decimal point keys to enter the required frequency, e.g. 1.234 11) Press the multiplier key (if required) e.g. 12) Press 13) The display will now indicate the value that has been entered below the currently set output 14) Press 15) Press The new value will replace the currently set frequency to activate the calibrator output.
1000 SERIES OPERATION MANUAL The display will show the frequency in the bottom right hand corner of the display. The frequency can be adjusted using the digital control or the directional keys when the cursor is placed over the frequency. To move the cursor from voltage setting to frequency setting, press the key, followed by For protection of UUT’s, the 1000 Series calibrators high voltage outputs are ramped, so that voltage is increased gradually.
1000 SERIES OPERATION MANUAL Setting an AC Current Output Complete the following procedure to set an AC current output. The and keys can be used to edit the entry in the event of an incorrect key press. Warning: ENSURE CONNECTIONS TO UUT ARE CORRECT To avoid damaging UUT protection fuses, ensure the correct terminals on the UUT are used before sourcing currents 1) Ensure the calibrator’s output has been set to Standby.
1000 SERIES OPERATION MANUAL 9) Press . The new value will replace the existing output in the middle of the display. 10) Press the numeric and decimal point keys to enter in the desired frequency, e.g. 80 11) Press the multiplier key (if required) e.g. 12) Press 13) The display will show the new entry below the previous set value. 14) Press The new value will replace the existing frequency on the right of the display. 15) Press to activate the calibrator output.
1000 SERIES OPERATION MANUAL The 1000 Series calibrator is fitted with a temperature controlled high current output. This will automatically turn the high current output off once the internal temperature has reached a pre determined limit. The 1000A will indicate ‘Over Temperature’ on the Screen for approximately 5 seconds, and then return to the normal screen, with ‘Temp !’ displayed in the status section. Further information regarding the temperature output cut-out is available on Page 70.
1000 SERIES OPERATION MANUAL Setting the Simulated Resistance Output The calibrator defaults to simulated resistance output. This provides a fully variable resistance output from 0 Ohms to 10M Ohms. Complete the following example to set 10 Ohms output using the simulated resistance output. 1) Ensure the calibrator’s output has been set to Standby. This can be verified by ensuring that standby is indicated on the display, and the output standby status LED is lit.
1000 SERIES OPERATION MANUAL 9) Press to activate the calibrator output. The LED next to the key will light up, as well as the terminal indicator; indicating that the output is active. To adjust the value, either type in a new value using the keyboard, for example 110 Ohms, or deviate the output using the cursor keys or the digital control. Nulling the UUT The value displayed/set on the calibrator is the value at the terminals.
1000 SERIES OPERATION MANUAL Setting Passive Resistance Output Note: The calibrator uses standard resistors of fixed decade values. The nearest available resistance to the entered value will be automatically selected. The 1000 series calibrator features passive resistance output as an alternative to simulated resistance. This method does not allow the output to be varied, as the values are fixed passive standards. Complete the following example to set an output using the passive resistance output.
1000 SERIES OPERATION MANUAL 10) Press 11) The display will now indicate the value that has been entered below the currently set output 12) Press to activate the calibrator output. The LED next to the key will light up, as well as the terminal indicator; indicating that the output is active The output can be changed by using the and Using these keys will switch to the next decade value. Nulling the UUT The value displayed/set on the calibrator is the value at the terminals.
1000 SERIES OPERATION MANUAL Setting Capacitance Output Note: The calibrator uses standard capacitors of fixed values. The nearest available capacitance to the entered value will be automatically selected. Complete the following procedure to select 100nF. Complete the following example to set an output using the simulated resistance output. 1) Ensure the calibrator’s output has been set to Standby.
1000 SERIES OPERATION MANUAL 9) Press . The new value will replace the existing output in the middle of the display 10) Press to activate the calibrator output. The LED next to the key will light up, as well as the terminal indicator; indicating that the output is active. The output can now be changed by using the and . Using these keys will switch to the next decade value. Capacitance is available at the voltage terminals as indicated by the terminal indicator LED.
1000 SERIES OPERATION MANUAL Setting Frequency Output. Note: The calibrator uses a precision crystal oscillator for frequency output. Frequencies from 1Hz to 100kHz in 1Hz steps are available. Complete the following example to set an output using the frequency output. 1) Ensure the calibrator’s output has been set to Standby. This can be verified by ensuring that standby is indicated on the display, and the output standby status LED is lit.
1000 SERIES OPERATION MANUAL 9) Press to activate the calibrator output. The LED next to the key will light up, as well as the terminal indicator; indicating that the output is active. To adjust the value, either type in a new value using the keyboard, for example 10 Hz, or deviate the output using the cursor keys or the digital control as explained on page 35 The output can also be changed by using the and Using these keys will either multiply or divide the currently set value by a factor of 10.
1000 SERIES OPERATION MANUAL Thermocouple Simulation Thermocouple Adapter Connection Connect the temperature meter under test to the thermocouple output of the 1000A/B using compensation cable that matches the thermocouple type. 1) Ensure the calibrator’s output has been set to Standby. This can be verified by ensuring that standby is indicated on the display, and the output standby status LED is lit.
1000 SERIES OPERATION MANUAL 8) If required, press the AUTO CJ soft key to automatically take a measurement of the cold junction at the thermocouple socket. Once in Auto Cold Junction mode the temperature will continue to update automatically. 9) Press to activate the calibrator output. The LED next to the key will light up, as well as the terminal indicator; indicating that the output is active To change thermocouple type, press the Type soft key, this will bring up the Thermocouple Type menu.
1000 SERIES OPERATION MANUAL PRT Output The calibrator can simulate PRT temperature values with R0’s (the nominal resistance at 0ºC) of 25100, 250500 and 1000in the range -200C to 800C 1) Ensure the calibrator’s output has been set to Standby. This can be verified by ensuring that standby is indicated on the display, and the output standby status LED is lit.
1000 SERIES OPERATION MANUAL 9) The display will now indicate the value that has been entered below the currently set output 10) Press . The new value will replace the existing output in the middle of the display 11) Press to activate the calibrator output.
1000 SERIES OPERATION MANUAL EA002 – 2/10/50 Turn Coil Adapter (Option) The 1000 Calibrators can be used with the optional current coil EA002 (2, 10 and 50 turn coils). This enables currents both DC and AC of up to 500A to be produced for the calibration of current clamp meters. The current from the calibrator is connected to the appropriate coil connections.
1000 SERIES OPERATION MANUAL 1) Ensure the calibrator’s output has been set to Standby. This can be verified by ensuring that standby is indicated on the display, and the output standby status LED is lit. 2) Connect the EA002 to the appropriate terminals 3) Ensure the coil platform is attached to the EA002 and that the clamp coil is correctly aligned. 4) Press 5) Press 6) Press 7) Press the COIL soft key 8) The following menu will appear 9) Select the number of turns (e.g.
1000 SERIES OPERATION MANUAL 14) The display will now indicate the value that has been entered below the currently set output . The new value will replace the existing output in the middle 15) Press of the display 16) Press to activate the calibrator output. The LED next to the key will light up, as well as the terminal indicator; indicating that the output is active. Note : When setting AC Frequencies, at high frequencies ( >300Hz) the coil will emit a high frequency tone.
1000 SERIES OPERATION MANUAL Insulation Resistance Calibration (Option) The insulation resistance simulation function provides resistance calibration for insulation testers. 1) Ensure the calibrator’s output has been set to Standby. This can be verified by ensuring that standby is indicated on the display, and the output standby status LED is lit. 2) Set the insulation tester to the required Insulation voltage range and connect it to the voltage terminals of the calibrator. 3) Press the INS TEST soft key.
1000 SERIES OPERATION MANUAL 9) Press and hold the Insulation tester TEST button and ensure it is kept depressed for the duration of the test 10) Press to activate the calibrator output.
1000 SERIES OPERATION MANUAL Insulation Test Voltage Measurement (Option) The calibrator can measure insulation test voltage measurement from an insulation/continuity or installation tester up to 1000V 1) Ensure the calibrator’s output has been set to Standby. This can be verified by ensuring that standby is indicated on the display, and the output standby status LED is lit. 2) Set the insulation tester to the required insulation voltage range and connect it to the voltage terminals of the calibrator.
1000 SERIES OPERATION MANUAL 7) The 1000 Series will display the test voltage measured from the tester (UUT) Note : If the polarity of the input from the insulation tester is reversed, the 1000 series will display INCORRECT INPUT. In this case, simply reverse the input from the insulation tester to resolve this. TRANSMILLE LTD. Version 1.
1000 SERIES OPERATION MANUAL Continuity Resistance Calibration (Option) The 1000 Series provides a continuity resistance function for calibrating the continuity ranges of insulation testers. 1) Ensure the calibrator’s output has been set to Standby. This can be verified by ensuring that standby is indicated on the display, and the output standby status LED is lit. 2) Set the tester to the required continuity range and connect it to the voltage terminals of the calibrator. 3) Press the CONT. soft key.
1000 SERIES OPERATION MANUAL 10) The 1000 Series will show the resistance value which can be read from the tester (UUT) display TRANSMILLE LTD. Version 1.
1000 SERIES OPERATION MANUAL Continuity Current Measurement [1 Ohm] (Option) The calibrator can measure continuity current from a tester using a 1 Ohm load. 1) Ensure the calibrator’s output has been set to Standby. This can be verified by ensuring that standby is indicated on the display, and the output standby status LED is lit. 2) Set the insulation tester to the required continuity range and connect it to the voltage terminals of the calibrator. 3) Press the CONT. soft key. If the CONT.
1000 SERIES OPERATION MANUAL 7) The 1000 Series will display the continuity current measured from the tester (UUT) TRANSMILLE LTD. Version 1.
1000 SERIES OPERATION MANUAL Output Protection and Safety Features Warning and Output Overload Indications The self-test function of the 1000 series calibrator continuously monitors the output of the calibrator for overload or fault conditions. In the event of the calibrator not being able to drive the load, it will automatically trip into standby and the display will show Standby.
1000 SERIES OPERATION MANUAL High Voltage Current Limit The 1000 Series is fitted with the same high voltage current limiting circuitry as the 3000A Series Calibrator. This circuit consists of completely independent circuit that monitors the current flowing in the high voltage section of the instrument. A fast acting Triac is used so that in the event of too much current being drawn from the high voltage output, the output is cut off immediately.
1000 SERIES OPERATION MANUAL Remote Programming WARNING The 1000 series calibrators can produce high voltages up to 1025V and must be programmed with due caution to prevent dangerous voltages from being present at the output terminals without warning to the operator.
1000 SERIES OPERATION MANUAL Returning to Local Control When the calibrator has been controlled from the interface, the front panel controls are disabled. To regain front panel control use the LOCAL soft key Programming Commands Overview The 1000 Series calibrators are controlled by a set of simple high level commands which can be used either individually or as part of a command sequence. The commands can be joined together using the / (forward slash) character.
1000 SERIES OPERATION MANUAL Response Codes The 1000 Series calibrators will respond to any command with a fixed code beginning with a star (*) - the codes are listed below Response Code Description *0 *1 *2 *3 *4 *5 *6 *7 COMMAND RECEIVED OK ERROR IN COMMAND LINE ERROR IN RANGE COMMAND ERROR IN FREQUENCY COMMAND ERROR IN O/P COMMAND ERROR IN CAL FACTOR SENT ERROR IN CAL FACTOR COMPARE COMMAND OUT OF RANGE ( A1,A2 ETC ) OR PASSWORD NOT SET 10A/HV TIMEOUT or OVER TEMPERATURE OUTPUT ERROR *8 *9 TRANSMI
1000 SERIES OPERATION MANUAL DC Voltage Commands Function Range DC Voltage 100mV 1V 10V 100V 1000V Standby Mode Standby ON Standby OFF Output Set Output Command R1 R2 R3 R4 R5 S1 S0 O (not zero) The DC voltage section consists of a set of range commands which are used in conjunction with the standby and output command. To enable a DC Voltage to be set and an output assigned, the following command sequence should be used: /
1000 SERIES OPERATION MANUAL Additional examples 90mV DC R1/O90/S0 (sets 150mV output on the 100mV range) 22V DC (sets 22V output on the 100V range) R3/O22/S0 If a command includes a value which cannot be set due to, for example, the value being higher than the range maximum, the calibrator will reject the command and stay set as it is (the calibrator will also beep to signify a rejected command) The calibrator will respond to the commands sent with the response codes as detailed at the beginn
1000 SERIES OPERATION MANUAL AC Voltage Commands Function Range AC Voltage 100mV 1V 10V 100V 1000V Standby Mode Standby ON Standby OFF Set Output Fxxxxx R12 R13 R14 R15 R16 S1 S0 Output AC Frequency Command O (not zero) E.G. 10kHz = F10000 The AC voltage section consists of a set of range commands which are used in conjunction with the standby and output command.
1000 SERIES OPERATION MANUAL Additional examples 35mV @ 1kHz AC R12/O35/F1000/S0 (sets 35mV @ 1kHz output on the 100mV range) 255V @ 200Hz AC R16/O255/F200/S0 (sets 255V @ 200Hz output on the 1000V range) If a command includes a value which cannot be set due to, for example, the value being higher than the range maximum, the calibrator will reject the command and stay set as it is (the calibrator will also beep to signify a rejected command) The calibrator will respond to the commands sent with t
1000 SERIES OPERATION MANUAL DC Current Commands Function Range DC Current 100uA 1mA 10mA 100mA 1A 10A Standby Mode Standby ON Standby OFF Output Set Output Command R6 R7 R8 R9 R10 R11 S1 S0 O (not zero) The DC current section consists of a set of range commands which are used in conjunction with the standby and output command.
1000 SERIES OPERATION MANUAL Additional examples 25mA DC R9/O25/S0 (sets 25mA output on the 100mA range) 5A DC (sets 5A output on the 10A range) R11/O5/S0 If a command includes a value which cannot be set due to, for example, the value being higher than the range maximum, the calibrator will reject the command and stay set as it is (the calibrator will also beep to signify a rejected command) The calibrator will respond to the commands sent with the response codes as detailed at the beginning
1000 SERIES OPERATION MANUAL AC Current Commands Function Range AC Current 100uA 1mA 10mA 100mA 1A 10A Standby Mode Standby ON Standby OFF Set Output Fxxxxx R17 R18 R19 R20 R21 R22 S1 S0 Output AC Frequency Command O (not zero) E.G. 10kHz = F10000 The AC current section consists of a set of range commands which are used in conjunction with the standby and output command.
1000 SERIES OPERATION MANUAL Additional examples 25mA @ 500Hz AC R20/O25/F500/S0 (sets 25mA @ 500Hz output on the 100mA range) 7A AC @ 300 Hz R22/O7/F300/S0 (Sets 7A @ 300Hz output on the 10A range) If a command includes a value which cannot be set due to, for example, the value being higher than the range maximum, the calibrator will reject the command and stay set as it is (the calibrator will also beep to signify a rejected command) The calibrator will respond to the commands sent with the res
1000 SERIES OPERATION MANUAL Passive Resistance Commands Function Passive Resistance Range 10 Ohms 100 Ohms 1 kOhm 10 kOhms 100 kOhms 1 MOhms 10 MOhms 100 MOhms Standby Mode Standby ON Standby OFF Resistance Passive output Simulated output Command R26 R27 R28 R29 R30 R31 R32 R33 S1 S0 I0 I2 The resistance section consists of a set of range commands which are used in conjunction with the passive/simulated mode and standby commands.
1000 SERIES OPERATION MANUAL For example, to set the 1kOhm range passive resistance output with the output switched on, the command is: R28/I0/S0 R28 = 1KOhm Range (as detailed in the table above) I0 = 2 Wire passive mode S0 = Standby OFF (i.e. output switched ON) = Carriage Return (ASCII character 13) Additional examples 10 MOhm Passive R32/I0/S0 The calibrator will respond to the commands sent with the response codes as detailed at the beginning of this section.
1000 SERIES OPERATION MANUAL Capacitance Commands Function Range Capacitance Command 10nF 100nF 1uF Standby Mode Standby ON Standby OFF R35 R38 R39 S1 S0 The capacitance section consists of a set of range commands which are used in conjunction with the standby command. To enable a capacitance to be set, the following command sequence should be used: / The use of the output (O) command is not necessary as the capacitance ranges are fixed ranges at decade points.
1000 SERIES OPERATION MANUAL Simulated Resistance Commands Function Mode Activation Command I2 Simulated Resistance Range 0 Ohms – 10.0 Ohms 10 Ohms – 50 Ohms 50 Ohms – 100 Ohms 100 Ohms – 1k Ohms 1kOhms – 9.99kOhms 10kOhms – 99.9kOhms 100kOhms – 999kOhms 1MOhms – 9.99MOhms Standby Mode Standby ON Standby OFF Output Set Output Command R26 R53 R27 R28 R29 R30 R31 R32 Output Ohms Ohms Ohms kOhms kOhms kOhms MOhms MOhms S1 S0 x.
1000 SERIES OPERATION MANUAL Frequency Commands Function Frequency Mode Activation Command 1Hz 10Hz 100Hz 1kHz 10kHz 20kHz 50kHz 100kHz Standby Mode Standby ON Standby OFF r96 H0 H1 H2 H3 H4 H5 H6 H7 S1 S0 The frequency section consists of a mode activation command followed by a set of range commands which are used in conjunction with the standby command.
1000 SERIES OPERATION MANUAL For example, to set the 10 kHz output with the output switched on, the command would be: R58/H4/S0 R58 = Frequency Mode Activation (as detailed in the table above) H4 = 10 kHz output (as detailed in the table above) S0 = Standby OFF (i.e.
1000 SERIES OPERATION MANUAL Thermocouple Simulation Commands This function requires the use of the optional Thermocouple simulation adapter. This is used in conjunction with the adapter interface on the 1000 Series to provide the thermocouple simulation.
1000 SERIES OPERATION MANUAL To enable thermocouple simulation to be set up, the following command sequence should be used: //// For example, to set the following configuration: 6) AUTOMATIC COLD JUNCTION COMPENSATION 7) TYPE R 8) 250°C 9) Output ON Send the following command sequence: R60/K1/L4/O250/S0 R60 = Thermocouple simulation mode activation K1 = Automatic cold junction compensation (as detailed in the table above) L4 = Type R the
1000 SERIES OPERATION MANUAL Additional examples Type K: Auto CJC: 500°C = R60/K1/L1/O500/S0 Type K: Auto CJC: 1500°C = R60/K1/L1/O1500/S0 Type E: Manual CJC (0°C): 400°C = R60/K0/L6/O400/S0 Type N: Auto CJC: -100°C = R60/K1/L7/O-100/S0 Type K: Manual CJC (24.5°C): 200°C = R60/KT24.
1000 SERIES OPERATION MANUAL Technical Description General The 1000 Series calibrators use the latest in reference, resistor and processor technology designed to minimise cost and size yet maximise performance. The microprocessor controls and monitors all functions of the calibrator. Calibration constants are held in a non-volatile memory allowing the calibration to be performed without removing the covers. There are no internal adjustments required in normal service. Warning risk of shock.
1000 SERIES OPERATION MANUAL Internal Fuses In normal operation these fuses should never need to be replaced. Only under fault conditions will they require changing. NOTE: To access these fuses it is necessary to dismantle the case, this should only be carried out by a qualified person. Warning risk of shock.
1000 SERIES OPERATION MANUAL Opening The Case Warning risk of shock. The line power cord must be disconnected before removing the instrument from it chassis 6 screws are used to attach the 1000 series to the chassis. The locations of these screws are highlighted in Red on the image below TRANSMILLE LTD. Version 1.
1000 SERIES OPERATION MANUAL Access to Internal Fuses After removing the instrument from the case, the fuses as accessible as shown in the diagram below. TRANSMILLE LTD. Version 1.
1000 SERIES OPERATION MANUAL Getting the best out of the calibrator. The 1000 series are very accurate calibrators producing a very wide range of output signals. To make the best possible use of the range of outputs and to eliminate errors this section details some common sources of errors and offers some techniques to reduce them. Thermally generated EMF voltage errors. At every connection in a measuring system different metals come into contact with each other, each junction forms a thermocouple.
1000 SERIES OPERATION MANUAL Power line and low frequency Pick up and noise These effects are most noticeable when using high resistance (100kohms and above) and low current. All constant current sources have a very high output impedance which will pick up noise just like the high value resistance. To reduce pickup, use screened leads and try earthing the low side of the calibrator output. For high value resistance it is essential that the cables insulation resistance will not affect the accuracy.
1000 SERIES OPERATION MANUAL Calibration and Maintenance WARNING The information in this section is intended only for qualified personnel. The user must at all times be adequately protected from electric shock. General The 1000 Series calibrators’ maintenance requirements are listed below. Please note that the calibrator does not require any regular internal servicing or adjustment.
1000 SERIES OPERATION MANUAL Cleaning of the Fan Vents WARNING : Risk of Shock Ensure calibrator is disconnected from line power before proceeding. Fan ducts may be cleaned with brush and vacuum cleaner Cleaning the external case Use a damp cloth with a mild water based cleaner for the outside case and front panel. Do not use alcohol based cleaners or solvents and do not spill or allow liquid to enter the case. TRANSMILLE LTD. Version 1.
1000 SERIES OPERATION MANUAL Calibration To adjust the 1000 Series calibrator the calibrator can either be connected to a computer via the USB interface and adjusted using ProCal software or adjusted directly from the front panel. Adjustment can be completed without disassembly of the calibrator. REFER TO THE SERVICE MANUAL FOR CALIBRATION PROCEDURE. THE CALIBRATION OF THE INSTRUMENT SHOULD ONLY BE CARRIED OUT BY QUALIFIED PERSONNAL TRANSMILLE LTD. Version 1.
0 SERIES OPERATION MANUAL Guarantee and service Transmille Ltd. guarantees this instrument to be free from defects under normal use and service for a period of 1 year from purchase. This guarantee applies only to the original purchaser and does not cover fuses, or any instrument which, in Transmille’s opinion, has been modified, misused or subjected to abnormal handling or operating conditions.
1000 SERIES OPERATION MANUAL Transmille Ltd. Unit 4, Select Business Centre Lodge Road Staplehurst Kent. TN12 0QW United Kingdom. Tel Fax : +44 0 1580 890700 : +44 0 1580 890711 Email : sales@transmille.com Web : www.transmille.com 1000 Series Fax Back Form Your 1000 Series Multi-Product Calibrator is fitted with a security system which requires a security code to be entered to allow continued operation of the unit beyond the 60 Day evaluation period.
1000 SERIES OPERATION MANUAL Appendix A Installing the USB Interface Driver (Windows XP) Insert the supplied USB lead driver CD into the computer CD drive Click on menu to install driver – follow on screen prompts.
1000 SERIES OPERATION MANUAL Installing the USB Interface Driver (Windows Vista / 7) Insert the supplied USB lead driver CD into the computer CD drive Click on menu to install driver – follow on screen prompts.
1000 SERIES OPERATION MANUAL Checking the COM Port setting for the USB Interface Once the USB interface driver is installed, it will have assigned a ‘virtual’ COM port number which is needed for setting up the instrument for computer control (via optional ProCal Calibration software).