8. When the firmware download is complete, return the device to the
protected mode.
9. Click the Disconnect button to disconnect the PC serial port.
10. Remove the short-block (jumper) at JP3 pins 1 and 3.
11. Recycle power into the radio board.
Figure 5–3.
This completes the firmware download procedure.
Before returning an instrument, you must contact Thermo for a Return
Material Authorization number. You can contact Thermo at:
Thermo Electron Corporation
Process Instruments Division
1410 Gillingham Lane
Sugar Land, TX 77478 USA
Phone: 713-272-0404
Fax: 713-272-2272
Web: www.thermo.com