Datasheet

www.ti.com
INTERNAL TEMPERATURE SENSOR
Converter
AGND
SNSVDD
TEMP1
TEMP2
+IN
-IN
DV +
kT
q
@ ln(N)
(3)
T +
q @ DV
k @ ln(N)
(4)
TSC2004
SBAS408E JUNE 2007 REVISED MARCH 2008
In some applications, such as battery recharging, an ambient temperature measurement is required. The
temperature measurement technique used in the TSC2004 relies on the characteristics of a semiconductor
junction operating at a fixed current level. The forward diode voltage (V
BE
) has a well-defined characteristic
versus temperature. The ambient temperature can be predicted in applications by knowing the +25 ° C value of
the V
BE
voltage and then monitoring the delta of that voltage as the temperature changes.
The TSC2004 offers two modes of temperature measurement. The first mode requires calibration at a known
temperature, but only requires a single reading to predict the ambient temperature. The TEMP1 diode, shown in
Figure 27 , is used during this measurement cycle. This voltage is typically 580mV at +25 ° C with a 10 µ A current.
The absolute value of this diode voltage can vary by a few millivolts; the temperature coefficient (T
C
) of this
voltage is very consistent at 2.1mV/ ° C. During the final test of the end product, the diode voltage is stored at a
known room temperature, in system memory, for calibration purposes by the user. The result is an equivalent
temperature measurement resolution of 0.3 ° C/LSB (1LSB = 610 µ V with V
REF
= 2.5V).
Figure 27. Functional Block Diagram of Temperature Measurement Mode
The second mode does not require a test temperature calibration, but uses a two-measurement (differential)
method to eliminate the need for absolute temperature calibration and for achieving 2 ° C/LSB accuracy. This
mode requires a second conversion of the voltage across the TEMP2 diode with a resistance 91 times larger
than the TEMP1 diode. The voltage difference between the first (TEMP1) and second (TEMP2) conversion is
represented by:
Where:
N = the resistance ratio = 91.
k = Boltzmann's constant = 1.3807 × 10
-23
J/K (joules/kelvins).
q = the electron charge = 1.6022 × 10
-19
C (coulombs).
T = the temperature in kelvins (K).
This method can provide much improved absolute temperature measurement, but a lower resolution of
1.6 ° C/LSB. The resulting equation to solve for T is:
Where:
Δ V = V
BE
(TEMP2) V
BE
(TEMP1) (in mV).
T = 2.573 Δ V (in K),
or T = 2.573 Δ V 273 (in ° C).
Temperature 1 and/or temperature 2 measurements have the same timing as Figure 46 .
Copyright © 2007 2008, Texas Instruments Incorporated Submit Documentation Feedback 17
Product Folder Link(s): TSC2004