SBAS499A JULY 2012REVISED AUGUST 2012
Reset (RESET)
There are two methods to reset the ADS1299: pull the RESET pin low, or send the RESET opcode command.
When using the RESET pin, take the pin low to force a reset. Make sure to follow the minimum pulse width
timing specifications before taking the RESET pin back high. The RESET command takes effect on the eighth
SCLK falling edge of the opcode command. On reset, 18 t
CLK
cycles are required to complete initialization of the
configuration registers to default states and start the conversion cycle. Note that an internal RESET is
automatically issued to the digital filter whenever the CONFIG1 register is set to a new value with a WREG
command.
START
The START pin must be set high or the START command sent to begin conversions. When START is low or if
the START command has not been sent, the device does not issue a DRDY signal (conversions are halted).
When using the START opcode to control conversions, hold the START pin low. The ADS1299 features two
modes to control conversions: continuous mode and single-shot mode. The mode is selected by SINGLE_SHOT
(bit 3 of the CONFIG4 register). In multiple device configurations, the START pin is used to synchronize devices
(see the Multiple Device Configuration subsection of the SPI Interface section for more details).
Settling Time
The settling time (t
SETTLE
) is the time required for the converter to output fully-settled data when the START
signal is pulled high. When START is pulled high, DRDY is also pulled high. The next DRDY falling edge
indicates that data are ready. Figure 35 shows the timing diagram and Table 8 shows the settling time for
different data rates. The settling time depends on f
CLK
and the decimation ratio (controlled by the DR[2:0] bits in
the CONFIG1 register). Table 7 shows the settling time as a function of t
CLK
. Note that when START is held high
and there is a step change in the input signal, 3 × t
DR
is required for the filter to settle to the new value. Settled
data are available on the fourth DRDY pulse.
Figure 35. Settling Time
Table 8. Settling Time for Different Data Rates
DR[2:0] NORMAL MODE UNIT
000 521 t
CLK
001 1033 t
CLK
010 2057 t
CLK
011 4105 t
CLK
100 8201 t
CLK
101 16393 t
CLK
110 32777 t
CLK
Copyright © 2012, Texas Instruments Incorporated Submit Documentation Feedback 29
Product Folder Link(s): ADS1299