Datasheet

HVP120
High Voltage Probes
Typical Voltage Derating HVP120
Measurement Category I
0
200
400
600
800
1000
1200
100K 1M 10M100M1G
Frequency [Hz]
Amplitude AC rms [V]sinus
Electrical Characteristics
Bandwidth 400 MHz
Risetime (10% - 90%) 900 ps (typical)
Maximum Input Voltage*
Measurement Category II 1000 Vrms
Measurement Category I 4000V Transient Overvoltage at 1000 Vrms
6000V Transient Overvoltage at 0 Vrms
Pollution Degree* 2
Input Capacitance 7.5 pF (typical)
Compensation Range 10 pF - 50 pF (typical)
Attenuation Ratio 100:1 ± 2%
Environmental
Temperature (Operating) 0°C to 50°C
Temperature (Non-Operating) -40°C to 71°C
Humidity (Operating) 80% RH (Non-Condensing) up to 31°C,
decreasing linearly to 40% RH at 50°C
Altitude (Operating) up to 2,000 m
Altitude (Non-Operating) up to 15,000 m
General Characteristics
Weight (probe) 67 g (0.15 lbs)
Cable Length 2 m (6.56 ft)
Probe Tip Diameter 5 mm (0.20 inches)
* As dened in IEC 61010-031
Specications
The HVP120 is a high voltage passive probe designed for probing up to
1,000 Vrms and capable of handling up to 6,000 V peak transients. Its fast
rise time and excellent frequency response make it suitable for a wide variety
of high voltage measurement applications. The HVP120 features a spring
loaded probe tip and a variety of standard accessories to make probing high
voltages safe and easy.
HVP120
RMS vs. Peak Pulse Voltage
Measurement Category I
0
1000
2000
3000
4000
5000
6000
7000
8000
0200 400600 8001000
AC rms or DC [V]
Peak Pulse [V]
Key Features
400 MHz probe bandwidth
900 ps rise time
1000 Vrms maximum input
Up to 6 kV transient overvoltage

Summary of content (2 pages)