Datasheet

Trigger system
Trigger modes Auto (supports Roll Mode for 40 ms/div and slower), Normal, Single Sequence
B trigger Trigger after time or events
Trigger after time range 13.2 ns to 50 s
Trigger after events range 1 to 9,999,999 events
Trigger types
Edge Conventional level-driven trigger. Positive or negative slope on any channel. Coupling selections: AC, DC, Noise Reject, HF
Reject, LF Reject
Video Trigger on all lines or individual lines, odd/even or all fields on NTSC, PAL, SECAM
Extended video Trigger on specific lines in broadcast and non-broadcast (custom) standards and on analog HDTV formats (1080i, 1080p, 720p,
480p). Requires TDS3VID or TDS3SDI application module
Pulse width (or glitch) Trigger on a pulse width <, >, =, ≠ to a selectable time limit ranging from 39.6 ns to 50 s
Runt Trigger on a pulse that crosses one threshold but fails to cross a second threshold before crossing the first again
Rise/fall time Trigger on pulse edge rates that are either faster or slower than a set rate. Edges can be rising, falling, or either
Pattern Specifies AND, OR, NAND, NOR when true or false for a specific time
State Any logic state. Triggerable on rising or falling edge of a clock. Logic triggers can be used on combinations of 2 inputs (not 4)
Comm Provides isolated pulse triggering required to perform DS1/DS3 telecommunications mask testing per ANSI T1.102 standard.
Requires TDS3TMT application module
Alternate Sequentially uses each active channel as a trigger source
Acquisition system
DPO Captures and displays complex waveforms, random events and subtle patterns in actual signal behavior. DPOs provide
3 dimensions of signal information in real time: Amplitude, time, and the distribution of amplitude over time
Sample Sample data only
Average Waveform averaged, selectable from 2 to 512
Envelope Min-max values acquired over one or more acquisitions
Peak detect High-frequency and random glitch capture. Captures glitches as narrow as 1 ns (typical) using acquisition hardware at all time
base settings
WaveAlert
®
Monitors the incoming signals on all channels and alerts the user to any waveform that deviates from the normal waveform being
acquired
Single sequence Use the Single Sequence button to capture a single triggered acquisition sequence at a time
Datasheet