Datasheet

Trigger Setup
Trigger type Trigger conditions
Setup and Hold Trigger when a logic data input changes state inside of
the setup or ho
ld time relative to a clock edge.
Setup is the amount of time that data should be stable
and not change before a clock edge occurs. Hold is the
time that data
should be stable and not change after
a clock edge occurs.
MSO2000B series oscilloscopes are capable of multiple
channel Setu
p and Hold triggering, and can monitor
the state of an entire bus for setup and hold violations.
You can use up to 20 channels for a Setup and Hold
trigger (4 a
nalog and 16 digital) with MSO2000B series
oscilloscopes.
Use the Clock side-bezel button to select the clock
channel. U
se the Select control, Data,andNot u sed
buttons to select one or more channels you want to
monitor for setup and hold violations.
Rise/Fall Time Trigger on rise and fall times. Trigger on pulse edges
that trav
erse between two thresholds at faster or slower
rates than the specied time. Specify pulse edges as
positive or negative or either.
Video
Trigger on specied elds or lines of a c omposite video
signal.
Only composite signal formats are supported.
Trigger on NTSC, PAL, or SECAM. Works with
Macrovision signals.
Bus Trigger on various bus conditions.
I
2
Cre
quires a DPO2EMBD module.
SPI requires a DPO2EMBD module.
CAN requires a DPO2AUTO module.
LIN r
equires a DPO2AUTO module.
RS-232, RS-422, RS-485, and UART require a
DPO2COMP module.
Para
llel requires a n MS O2000B series oscilloscope.
(See page 13, Application Module Free Trial.)
MSO2000B and DPO2000B Series Oscilloscopes User Manual 69