When the bladder wall thickness measurement is selected, the thickness value xxmm, 3D images and
projection images are displayed on the left side of the screen. Scan completion icon appears at
the bottom left of the screen. After scanning, click this key to enter patient information input, editing, stor-
age and printing. The hospital name, department name, date and time, battery power are displayed at the
top right end of the screen, followed by setting Icon , probe frequency , gallery view icon
, grid view icon , add new patient icon , gender selection , patient history information
review . On the right side of the screen, gender information appears below the patient history
information review icon. Slide the figure up and down to the left or right side of the bladder projection cir-
cle, the screen brightness adjustment icon will appear. Along with the finger movement up and down, the
brightness of the screen also changes.
Remark: The icons are active when the color is white. When it turns grey, it means they system is not
responding. Only after scan is completed, scan completion icon becomes active.
6.3 Gender Selection
Select gender icon on the main page, gender selection menu appears (Figure 5-3) including options
as Male, Female, FemaleH, Child and Phantom. The selected gender appears on the up right corner of
the screen after the choice is made. By default gender selection is Male.
6.4 Bladder Volume Measurement Scan
On the main interface (Figure 5-2) press the "D" position to choose the measurement object “Bladder Vol-
ume to enter scan interface.
6.4.1 Bladder Pre-scan
Apply the coupling agent to the surface of the probe which is the main contact area with the patients.
Place the probe on the skin of the lower abdomen where bladder is located. Press SCAN button once to
start pre-scan. Under expert mode, real-time B-Mode image of the bladder (Figure 5-4) appears in the
screen. Cross-sectional image of the bladder (Figure 5-8) appears under easy mode and bladder
projection images are displayed under intelligence mode and meanwhile B-mode images appear on the
left part of the screen.
Under pre-scan of the expert mode, observe the B-Mode images and move the probe till the biggest
image is obtained. While achieving the biggest B-Mode image, make sure the centre of the ultrasonic