User's Manual

74 6390-009-001 REV B www.stryker.com
Return To Table of Contents
Recommended Separations Distances for Model 6390 Power-LOAD System
Model 6390 Power-LOAD System is intended for use in the electromagnetic environment in which radiated disturbances
are controlled. The customer or user of Model 6390 Power-LOAD System can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF Communications Equipment
and Model 6390 Power-LOAD System as recommended below, according to the maximum output power of the
communications equipment.
Max Output
Power
(Watts)
Separation (m)
150kHz to 80MHz
D=(3.5/V1)(√P)
Separation (m)
80 to 800MHz
D=(3.5/E1)(√P)
Separation (m)
800MHz to 2.5GHz
D=(7/E1)(√P)
0.01 0.1166 0.1166 0.2333
0.1 0.3689 0.3689 0.7378
1 1.1666 1.1666 2.3333
10 3.6893 3.6893 7.3786
100 11.6666 11.6666 23.3333
POWER-LOAD (CONTINUED)