Guidance and Manufacturer’s Declaration: Electromagnetic Immunity
The Crossre™ System is intended for use in the electromagnetic environment specied below. The
customer or the user of Crossre™ System should ensure that it is used in such an environment
Immunity Test IEC 60601 Test
Level
Compliance Level Electromagnetic
Environment: Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6kV contact
±8kV air
±2,4,6kV contact
±2,4,8kV air
Floors should be wood,
concrete, or ceramic tile.
If oors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2kV for power
supply lines
±1kV for input/output
lines
±2kV for power supply
lines
±1kV for input/output
lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
±1kV differential
mode
±2kV common mode
±0.5, 1kV differential
mode
±1, 2kV common mode
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% Ut (>95% dip in
Ut) for 0.5 cycle
40% Ut (60% dip in
Ut) for 5 cycles
70% Ut (30% dip in
Ut) for 25 cycles
<5% Ut (>95% dip in
Ut) for 5 sec.
<5% Ut (>95% dip in Ut)
for 0.5 cycle
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles
<5% Ut (>95% dip in Ut)
for 5 sec.
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
Crossre™ System requires
continued operation during
power mains interruptions,
it is recommended that
Crossre™ System
be powered from an
uninterruptible power supply
or a battery.
Power frequency
(50/60Hz)
magnetic eld
IEC 61000-4-8
3 A/m N/A Power-frequency magnetic
elds should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE: Ut is the a.c. mains voltage prior to application of the test level.