A-2
T
ELEMETRY TRANSMITTER (96281)
A
PPENDIX A — GUIDANCE AND
M ANUFACTURER S DECLARATION
Electromagnetic Immunity
Table A-2 Electromagnetic Immunity
Immunity Test IEC 60601 Test Level
Compliance
Level
Electromagnetic
Environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
N/A N/A
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
N/A N/A
Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
for 0.5 cycle)
40% U
T
(60% dip in U
T
for 5 cycles)
70% U
T
(30% dip in U
T
for 25 cycles)
<5% U
T
(>95% dip in U
T
for 5 seconds)
N/A N/A
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency
magnetic fields should be
at levels characteristic of
a typical location in a
commercial or hospital
environment.
Note:
U
T
is the AC mains voltage prior to application of the test level.
Conducted RF
IEC 61000-4-6
N/A N/A
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the monitoring
device, including
cables, than the
recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Radiated RF
IEC 61000-4-3
3V/m
80 MHz to 2.5 GHz
3V/m