User's Manual
Table Of Contents
- 1. General description
- 2. Indications
- 3. Contraindications
- 4. Warnings and precautions
- 5. Adverse events
- 6. Clinical studies
- 7. Patient selection and treatment
- 8. Patient counselling information
- 9. Declaration of conformity
- 10. Physician guidelines
- 11. Patient information
- 12. How supplied
- 13. Device description
- 14. Implant procedure
- 14.1. Necessary equipment
- 14.2. Packaging
- 14.3. Optional equipment
- 14.4. Before opening the package
- 14.5. Prior to implantation
- 14.6. Device placement
- 14.7. Choosing the type of lead
- 14.8. Shock configuration (+ -> -)
- 14.9. Measurement of thresholds at implant
- 14.10. Leads connection
- 14.11. Device implantation
- 14.12. Tests and programming
- 15. Special modes
- 16. Main functions
- 17. Patient follow-up
- 18. Supplemental Information
- 19. Physical characteristics
- 20. Electrical characteristics
- 21. Programmable parameters
- 22. Non programmable parameters
- 23. Limited warranty
- 24. Patents
- 25. Explanation of symbols
Clinical status Values Nominal value "As shipped"
value
V oversensing
ON-OFF OFF OFF
High AT/AF burden
ON-OFF OFF OFF
AT/AF limit (on 24h)
(h)
0.5-1-3-6-12-24 6 6
Fast V Rate during
AT/AF
ON-OFF OFF OFF
Fast V Rate limit
(ppm)
80-90-100-110-120 100 100
Fast V Duration limit
(h)
0.5-1-3-6-12-24 1 1
Limited % of V pac-
ing in CRT
ON-OFF OFF OFF
Limited % of V pac-
ing (%)
50-70-80-85-90-95 80 80
Therapy informa-
tion
Values Nominal value "As shipped"
value
Shock disabled
ON-OFF ON ON
Shocks delivered
OFF-All shocks-Inefficient shock-
Inefficient max shock
All shocks All shocks
ATP delivered
ON-OFF OFF OFF
21. PROGRAMMABLE PARAMETERS
68 SORIN – PLATINIUM 4LV SonR CRT-D 1844 – U906A