User Guide

English
4422
Connect the 4 weighing machines (one for each wheel of the hospi-
tal bed) by the plugs to the distributor box. Check that they are
arranged correctly according to the coloured marking.
Each cable is 6 m long.
When getting started care must be taken to ensure that the cable
is not obstructed.
Connect the display unit to the mains.
Caution:
Before powering on the scale, position the weighing machines
under the 4 wheels of the hospital bed. Position the weighing
machines under the wheels right up to the limit stop.
The weighing machines are designed to take wheels with a diame-
ter of 150 mm!
1 2 3 4
Power-on
The Soehnle Professional logo is displayed during the power-on routine which
includes the offset and reference measurements.
The measuring points are briefly listed as they are detected.
The scale is zeroed after the power-on routine
At power-on the last used application program is reactivated.
Zero range limits at power-on
Calibratable: zero range at power-on is -10% to +60% of the weighing range.
Non calibratable: zero range at power-on is -99% to +99% of the weighing range.
If power-on is outside zero range, the scale shows an error message.
Once the cause of the underload or overload has been eliminated, the scale goes
to zero without the need to power-on again and is operative.